开云体育

ctrl + shift + ? for shortcuts
© 2025 开云体育
Date

Re: S11 and S22 comparisons on 8753

David Kirkby
 

On 23 December 2012 18:43, David McQuate <mcquate@...> wrote:
On "missing parts": some opens(mostly early, and I'm specifically
thinking of APC 3.5 connectors) consisted of a body having no center
conductor attached, plus an insertable center conductor on a plastic
rod. Without the inserted center conductor, the characteristics of the
"open" would not match the values assumed in the calibration calculations.
Was sort of like this



though for 3.5 mm, not N as in this diagram?

My own 3.5 mm kit does not have anything like this - the standards are
self contained - no extra bits needed, and nowhere to fit anything.


On "open / short combo": Neither standard has much length, so the two
can easily be packaged into one assembly, strictly for convenience (less
likely to be misplaced because it is larger). There's no coupling
whatsoever between the two ends.
True - they are convenient. There are 3 types of these

1) With just the open and short.
2) With open, short and load
3) With open, short, load and thru.

Eiether way, one needs to have the data on the bits.

If you read the 8753 manual, you will find the calibration kit it is
deisgned to work with is the HP 85032B (as in the photo I showed).
Any other cal kit has to be entered as a user calibration.

Dave


Re: S11 and S22 comparisons on 8753

 

On "missing parts": some opens(mostly early, and I'm specifically
thinking of APC 3.5 connectors) consisted of a body having no center
conductor attached, plus an insertable center conductor on a plastic
rod. Without the inserted center conductor, the characteristics of the
"open" would not match the values assumed in the calibration calculations.

On "open / short combo": Neither standard has much length, so the two
can easily be packaged into one assembly, strictly for convenience (less
likely to be misplaced because it is larger). There's no coupling
whatsoever between the two ends.

Dave

On 12/23/2012 7:11 AM, Peter Bunge wrote:

Merry Christmas all!
My friends and I have been comparing what we see from the S Parameter
Test Set on the two ports. These tests are with the VNA uncalibrated
and with proper opens or shorts (even from APC7 cal kits).
Typically we get completely different shaped curves ranging from -35
to -55 dB.
I get very similar results using the same S Parameter Test Set on an
HP8753B and an HP8753C so the curve shape is caused by the test set
not the VNA.
Typically we see one port 10 dB better and flatter than the other.
Any comments?

p.s. the cal kit discussion was very informative and timely. I still
don't understand the comment (that was later corrected) about the open
usually missing a part. My friend's Type N cal kit does not have extra
parts for the open. Also I have seen an open/short combo advertised
and wonder if this comes apart to use the open??? I have not used one
and find the concept of one end a short (OK) and the other an open
(how?) confusing.

--
Clear Stream Technologies


Re: HP 3456a was Re: HP 3478A general questions

 

Thanks again, I've downloaded the manuals on most of the HP DMM's from that era. I'm now looking at the 3456 or 3457. I've got some prices on a cal and yes around $140 should do it for a NIST traceable cal plus shipping. Of the two meters, 3456 or 3457 which would be the smarter purchase ? The issues with the batteries I'm well aware of. My Tek 2465BDM lost it's Dallas chip internal battery. I've had to recal it after replacing the chip. A process that takes a while. It's accurate within about 0.1%. Good enough for a scope. The DMM is not anywhere near it's rated cal of 0.03% due to the data loss.

In the interest of saving space I'm also looking at the HP 3468a or 3479a. I'm not seeing a major difference between the two except for perhaps some lower ranges on the 3478a. Any helpful info regarding those devices.

If I can score a 3456a or 3457a for a reasonable price then I'll be heading that direction. At the moment I'm stuck doing projects with 4,000 count B&K meters and a 20,000 DMM on the Tek 2465 which is not in cal to it's spec. So my purchases will be pretty soon. I have a Fluke 8010 on the way which I'm sure will also need cal.

Thanks,

Jeff

On 12/23/2012 7:08 AM, marvgozum wrote:



You're welcome, and all, enjoy the holidays. It may pay to download the 1993 HP Catalog, that was the last year all in the 345x line were available for sale, so you can compare and contrast the specs, also against the 3478.



The HP manuals, Keithley handbook and the volt-nuts archives are free, just download and search as needed, so you can get all the basic cal and operations info without cost to your misses before you actually buy something ;)

If you intend to stack a lot of gear ontop the 3456a and not sure about weight, put a 2x4 plank so weight get distributed to the edges of the chassis, this is were the most strength lies, while the top and bottom chassis covers are the weakest as there is also no cross bar for structural support. The plank allows a gap for added cooling, even if vents on the 3456a are on the sides and rear, the plank method works for all rack mounted sized gear.

As always too, heavies are best at the bottom, light at the top, as often the boat anchors are also made with stronger chassis to take not just its own weight but others too.

Lastly, I'm not a fan of designs that use battery backed up SRAM for system data, I prefer EEPROM at the least. Since all your purchases are used, the state of the battery is unknown and should be replaced, ASAP. If it snafus, or worst just corrupted, such as in buying "CANNOT TEST" "AS IS" type eBay gear, you'll need budget the possibility of adding $100-140 to your purchase cost to get it formally recalibrated.

With the 3456a, the all pots method puts the user in full control of its cal state, you can cal any range, any time [ keeping metrology guidelines in mind], so if you get one working or not, its also much easier to repair to working order entirely by yourself.

--- In hp_agilent_equipment@... <mailto:hp_agilent_equipment%40yahoogroups.com>, Jeff Machesky <jeff@...> wrote:

I have to say this response has really given me the most food for
thought. It provided exact details about the models and the pro's and
con's of each. It also made me realize something important. There is no
reason I have to buy one of these devices and run out and get it
calibrated. I can make sure it's readings are within reason and take it
out later on to get it calibrated. My main goal is resolution, so I can
have that now and have the accuracy and the resolution later. It will
also give me time to clean up and gain confidence in the device before
spending more money on it.

Very good feedback, Thanks.

Jeff


Re: HP-608A Repair

 

----- Original Message -----
From: "Patrick Wong" <patwong3@...>
To: <hp_agilent_equipment@...>
Sent: Sunday, December 23, 2012 9:35 AM
Subject: [hp_agilent_equipment] Re: HP-608A Repair


Hi Richard,

Thank you for your reply.

As an example, suppose I am on band C, set to the highest frequency. Peak the amp tuning knob. Set the power output knob so the RF output meter reads +4 dBm. Set the attenuator knob to 0 dBm. Measure the actual output (ideally using an RF power meter, but in my case I use my spectrum analyzer and RF probe on DMM.)

Then, move to band D. Turn frequency knob to the low end, same frequency as previously used on band C. Peak the amp tuning knob. Set power output so the RF output meter reads +4 dBm. Leave the attenuator knob at 0 dBm. Measure the actual power output.

In my case, I find that the actual output will vary even though the RF output meter reading remains constant.

Maybe there is some weird capacitance coupling effect between the attenuator probe antenna and the final amp output coil, as the bandswitch moves from one band to the next?

Patrick Wong AK6C
It is very odd. I need to give this some thought. The only think that comes to mind immediately is that the monitor may be affected by harmonics that do not affect the other instruments. I would also check the output at lower levels to see if the difference is affected by the position of the attenuator probe.


--
Richard Knoppow
Los Angeles
WB6KBL
dickburk@...


Re: S11 and S22 comparisons on 8753

David Kirkby
 

On 23 December 2012 17:09, David Kirkby <david.kirkby@...> wrote:



The small bit of gold played meter you see, is designed to increase
the diamteer of the last few mm of the pin, so it is 50 Ohms, and not
higher.
I meant "gold plated metal", not "gold played meter"

Dave


Re: HP 3456a was Re: HP 3478A general questions

 

I treat buying gear much like a trip to Vegas, I save up a few dollars and enjoy myself knowing full well that I'm wasting money. In the end it's not what you pay, it's how much you enjoy the experience. So if I pay too much for a piece of gear and it turns out to meet all my expectations..then money well spent. If on the other hand I get a lemon and have to sink more cash in to it..then it was a bad trip. It's often why I don't mind paying more for auctions that show proof of life. Better still proof of accuracy. I paid too much for my Tek 2465BDM, however it's in mint condition, works perfectly and I use it just about everyday. Money well spent.

Thanks,

Jeff

On 12/23/2012 10:20 AM, Bob Albert wrote:

Thanks, I guess I will have to just keep looking. For $50 I would love to have one of these, or one of its sister units. And being in Los Angeles, there is always a likelihood I can find one that doesn't have to be shipped.

Bob

--- On Sun, 12/23/12, marvgozum <mxg003@... <mailto:mxg003%40jefferson.edu>> wrote:

From: marvgozum <mxg003@... <mailto:mxg003%40jefferson.edu>>
Subject: [hp_agilent_equipment] HP 3456a was Re: HP 3478A general
questions
To: hp_agilent_equipment@...
<mailto:hp_agilent_equipment%40yahoogroups.com>
Date: Sunday, December 23, 2012, 5:20 AM


eBay, if you're in no rush, patience is rewarded by cheap
buys. I've tracked 3456a prices on eBay for 4 years,
and the price I posted are ranges I've found. The
killer is often S&H, as its a wide and heavy beast, but
luckily its short and flat. Its about the perimeter of
a small coffee table, which makes it easy to stack
instruments ontop of it.

Long ago, this website tracked prices for a lot of good
gear, but its seems to be orphaned. But it still gives
what hi-lo used to be before 2008, as well as Tektronix.





Many links are broken but you can drill down without a
problem.


--- In hp_agilent_equipment@...
<mailto:hp_agilent_equipment%40yahoogroups.com>,
Bob Albert <bob91343@...> wrote:

Where can you buy one of these for $50?

Bob



------------------------------------

Yahoo! Groups Links


hp_agilent_equipment-fullfeatured@...
<mailto:hp_agilent_equipment-fullfeatured%40yahoogroups.com>


Re: HP 3456a was Re: HP 3478A general questions

Bob Albert
 

Thanks, I guess I will have to just keep looking. For $50 I would love to have one of these, or one of its sister units. And being in Los Angeles, there is always a likelihood I can find one that doesn't have to be shipped.

Bob

--- On Sun, 12/23/12, marvgozum <mxg003@...> wrote:

From: marvgozum <mxg003@...>
Subject: [hp_agilent_equipment] HP 3456a was Re: HP 3478A general questions
To: hp_agilent_equipment@...
Date: Sunday, December 23, 2012, 5:20 AM


eBay, if you're in no rush, patience is rewarded by cheap
buys.? I've tracked 3456a prices on eBay for 4 years,
and the price I posted are ranges I've found.? The
killer is often S&H, as its a wide and heavy beast, but
luckily its short and flat.? Its about the perimeter of
a small coffee table, which makes it easy to stack
instruments ontop of it.

Long ago, this website tracked prices for a lot of good
gear, but its seems to be orphaned.? But it still gives
what hi-lo used to be before 2008, as well as Tektronix.





Many links are broken but you can drill down without a
problem.


--- In hp_agilent_equipment@...,
Bob Albert <bob91343@...> wrote:

Where can you buy one of these for $50?

Bob



------------------------------------

Yahoo! Groups Links


? ? hp_agilent_equipment-fullfeatured@...


Re: S11 and S22 comparisons on 8753

David Kirkby
 

On 23 December 2012 15:11, Peter Bunge <bunge@...> wrote:
Merry Christmas all!
Merry Christmas

My friends and I have been comparing what we see from the S Parameter Test Set on the two ports. These tests are with the VNA uncalibrated and with proper opens or shorts (even from APC7 cal kits).
Typically we get completely different shaped curves ranging from -35 to -55 dB.
If you have not calibrated the VNA, then you are just seeing the
uncorrect performance, and the fact that varies is no surprise. In
fact, I'm a bit surprised you managed to get 35 dB uncorrected. There
are a large number of systematic errors in a VNA. It does not matter,
as the whole idea of the error correction is to remove those.

I get very similar results using the same S Parameter Test Set on an HP8753B and an HP8753C so the curve shape is caused by the test set not the VNA.
Don't worry about it.

Typically we see one port 10 dB better and flatter than the other.
Any comments?
It is only important once calibrated. The calibration removes the
systematic errors.

p.s. the cal kit discussion was very informative and timely. I still don't understand the comment (that was later corrected) about the open usually missing a part. My friend's Type N cal kit does not have extra parts for the open.
It depends on what cal kit he has, but on for example and 85032B you
will often see something like this:



The small bit of gold played meter you see, is designed to increase
the diamteer of the last few mm of the pin, so it is 50 Ohms, and not
higher.

If you think about it, to get as near as possible to an open, you want
to have a 50 Ohm line that is cut off sharp at the end. Now consider
what an open N male looks like. The male pin is thicker at the bottom.
That bits makes 50 Ohm transmission line. Then the diameter is reduced
for about 5 mm, so it can go in the female. That will not make a 50
Ohm transmission line, but a higher impedance line, due to the formula

Zo=60 log(d_outer/d_inner)

So an open N plug is not as close as possible to an open. Does that
make sense? Hence the female part has an extender.

I would add it is possible to calibrate a VNA with an open N connector
without that. Accuracy suffers a bit, but that technique is used in
Agilent's portable VNA's. A better bet is an shielded N. But again you
need to know the paramters.

Also I have seen an open/short combo advertised and wonder if this comes apart to use the open??? I have not used one and find the concept of one end a short (OK) and the other an open (how?) confusing.
If you see an open/short advertised you MUST make sure you have the
paramters on it. It will have an offset in ps, and three capacitance
coefficents C0, C1, C2 and C3. You should then enter them in your VNA.
If you don't have that information, the kit is effectivly useless. The
short will have an offset too. The short might also have inductance
paramters L0, L1, L2, and L3, but you can't enter them in the 8753
series, so don't worry about them. They are not important for the the
frequency range of an 8753.

I hope that makes a bit more sence.

Dave


Re: S11 and S22 comparisons on 8753

J. Forster
 

A 30 dB return loss is nothing to sneeze at.

-John

=====================

Correction! We are seeing this with a good load attached, no calibration.
Same load gives different curve for each port.
I would expect the reflected power to be much lower than -30 dBm.
Of course with the open or short the reflected is close to 100%.
Too much egg nog.

--- In hp_agilent_equipment@..., "Peter Bunge" <bunge@...>
wrote:

Merry Christmas all!
My friends and I have been comparing what we see from the S Parameter
Test Set on the two ports. These tests are with the VNA uncalibrated and
with proper opens or shorts (even from APC7 cal kits).
Typically we get completely different shaped curves ranging from -35 to
-55 dB.
I get very similar results using the same S Parameter Test Set on an
HP8753B and an HP8753C so the curve shape is caused by the test set not
the VNA.
Typically we see one port 10 dB better and flatter than the other.
Any comments?

p.s. the cal kit discussion was very informative and timely. I still
don't understand the comment (that was later corrected) about the open
usually missing a part. My friend's Type N cal kit does not have extra
parts for the open. Also I have seen an open/short combo advertised and
wonder if this comes apart to use the open??? I have not used one and
find the concept of one end a short (OK) and the other an open (how?)
confusing.


Re: S11 and S22 comparisons on 8753

 

Correction! We are seeing this with a good load attached, no calibration. Same load gives different curve for each port.
I would expect the reflected power to be much lower than -30 dBm.
Of course with the open or short the reflected is close to 100%.
Too much egg nog.

--- In hp_agilent_equipment@..., "Peter Bunge" <bunge@...> wrote:

Merry Christmas all!
My friends and I have been comparing what we see from the S Parameter Test Set on the two ports. These tests are with the VNA uncalibrated and with proper opens or shorts (even from APC7 cal kits).
Typically we get completely different shaped curves ranging from -35 to -55 dB.
I get very similar results using the same S Parameter Test Set on an HP8753B and an HP8753C so the curve shape is caused by the test set not the VNA.
Typically we see one port 10 dB better and flatter than the other.
Any comments?

p.s. the cal kit discussion was very informative and timely. I still don't understand the comment (that was later corrected) about the open usually missing a part. My friend's Type N cal kit does not have extra parts for the open. Also I have seen an open/short combo advertised and wonder if this comes apart to use the open??? I have not used one and find the concept of one end a short (OK) and the other an open (how?) confusing.


Re: HP-608A Repair

 

I am not sure I understand this. Is the output of the generator varying or is the indication on the power monitor varying?
I think its possible that weak tubes, especially the buffer and amplifier may be responsible for both large variation in power at the monitor and poor modulation. If the power out is not consistent despite the monitor showing zero the monitor crystal could be bad or the little antenna wires at the end of the probe tube could be distorted.
I seem to remember letting 608's warm up for hours before doing final calibration on them. There is lots of mass and no compensation. I think this was the reason for the later version that could be stabilized with a synchronizer.

--
Richard Knoppow
Los Angeles
WB6KBL
dickburk@...

----- Original Message -----
From: "Patrick Wong" <patwong3@...>
To: <hp_agilent_equipment@...>
Sent: Sunday, December 23, 2012 7:28 AM
Subject: [hp_agilent_equipment] Re: HP-608A Repair


It appears that there are several members with 608s responding to this string.

With regards to the HP 608D, there are at least two attenuator circuit designs and therefore two schematics that can be found on line via Google search.

One design, the newer one, has two capacitors, and therefore resistance measurement at the RF output connector will normally show infinity.

The other has one capacitor and therefore a normal resistance measurement at the RF output connector will show ~53 ohms.

It is relatively easy to remove the attenuator probe for inspection, no special tools are needed, and it is not very hard to reinsert the probe although obviously care is needed to avoid damage to the grounding fingers surrounding the probe end.

I've never been able to figure out why the RF power meter on my 608D is not very accurate, at least based upon comparisons with my 8642B. The error is +/- 4 dB or more.

What is really weird is that if I produce a signal at the top frequency of one band, supposedly producing 0 dBm as measured per the meter, then switch to the next higher band, turn the tuning capacitor all the way to the low end of the range to produce the same frequency, and set output to 0 dBm per the meter, I find the actual output as measured by either my 8591E SA or Fluke 87 with RF probe is much different. The probe shouldn't care what frequency range the 608D is set to.

I tried cleaning the bandswitch contacts using electrical contact cleaner. The attenuator resistors measure correctly and show no signs of physical damage. The attenuator had been replaced at some point by HP service as a sticker on the chassis shows a notice to that effect.

My 608D has way too much frequency drift for my liking as well as tremendous incidental FM when using AM modulation. I posted about this a few months ago along with photos showing the incidental FM.

I'd be interested if other members owning 608-vintage equipment as well as spectrum analyzers can look at the quality of their AM signals with regards to incidental FM, and post photos showing good or bad results.

I recognize that if you are serious about using that generator, you have to leave it on 24 hr/day but I don't need to use a signal generator enough to have it running continuously, consuming 400W or whatever its power requirement is. So it sits around taking up space and serves mostly as a cabinet to put other stuff on.

Good luck,

Patrick Wong AK6C


Re: HP 3456a was Re: HP 3478A general questions

 

It also fits nicely in a suitcase... :)

--
Anders Gustafsson
Engineer, CNE6, ASE
Pedago, The Aaland Islands (N60 E20)
www.pedago.fi
phone +358 18 12060
mobile +358 40506 7099
fax +358 18 14060


"marvgozum" <mxg003@...> 2012-12-23 15:20 >>>

eBay, if you're in no rush, patience is rewarded by cheap buys. I've tracked 3456a prices on eBay for 4 years, and the price I posted are ranges I've found. The killer is often S&H, as its a wide and heavy beast, but luckily its short and flat. Its about the perimeter of a small coffee table, which makes it easy to stack instruments ontop of it.

Long ago, this website tracked prices for a lot of good gear, but its seems to be orphaned. But it still gives what hi-lo used to be before 2008, as well as Tektronix.





Many links are broken but you can drill down without a problem.


--- In hp_agilent_equipment@..., Bob Albert <bob91343@...> wrote:

Where can you buy one of these for $50?

Bob


S11 and S22 comparisons on 8753

 

Merry Christmas all!
My friends and I have been comparing what we see from the S Parameter Test Set on the two ports. These tests are with the VNA uncalibrated and with proper opens or shorts (even from APC7 cal kits).
Typically we get completely different shaped curves ranging from -35 to -55 dB.
I get very similar results using the same S Parameter Test Set on an HP8753B and an HP8753C so the curve shape is caused by the test set not the VNA.
Typically we see one port 10 dB better and flatter than the other.
Any comments?

p.s. the cal kit discussion was very informative and timely. I still don't understand the comment (that was later corrected) about the open usually missing a part. My friend's Type N cal kit does not have extra parts for the open. Also I have seen an open/short combo advertised and wonder if this comes apart to use the open??? I have not used one and find the concept of one end a short (OK) and the other an open (how?) confusing.


HP 3456a was Re: HP 3478A general questions

marvgozum
 

You're welcome, and all, enjoy the holidays. It may pay to download the 1993 HP Catalog, that was the last year all in the 345x line were available for sale, so you can compare and contrast the specs, also against the 3478.



The HP manuals, Keithley handbook and the volt-nuts archives are free, just download and search as needed, so you can get all the basic cal and operations info without cost to your misses before you actually buy something ;)

If you intend to stack a lot of gear ontop the 3456a and not sure about weight, put a 2x4 plank so weight get distributed to the edges of the chassis, this is were the most strength lies, while the top and bottom chassis covers are the weakest as there is also no cross bar for structural support. The plank allows a gap for added cooling, even if vents on the 3456a are on the sides and rear, the plank method works for all rack mounted sized gear.

As always too, heavies are best at the bottom, light at the top, as often the boat anchors are also made with stronger chassis to take not just its own weight but others too.

Lastly, I'm not a fan of designs that use battery backed up SRAM for system data, I prefer EEPROM at the least. Since all your purchases are used, the state of the battery is unknown and should be replaced, ASAP. If it snafus, or worst just corrupted, such as in buying "CANNOT TEST" "AS IS" type eBay gear, you'll need budget the possibility of adding $100-140 to your purchase cost to get it formally recalibrated.

With the 3456a, the all pots method puts the user in full control of its cal state, you can cal any range, any time [ keeping metrology guidelines in mind], so if you get one working or not, its also much easier to repair to working order entirely by yourself.

--- In hp_agilent_equipment@..., Jeff Machesky <jeff@...> wrote:

I have to say this response has really given me the most food for
thought. It provided exact details about the models and the pro's and
con's of each. It also made me realize something important. There is no
reason I have to buy one of these devices and run out and get it
calibrated. I can make sure it's readings are within reason and take it
out later on to get it calibrated. My main goal is resolution, so I can
have that now and have the accuracy and the resolution later. It will
also give me time to clean up and gain confidence in the device before
spending more money on it.

Very good feedback, Thanks.

Jeff


HP 3456a was Re: HP 3478A general questions

marvgozum
 

eBay, if you're in no rush, patience is rewarded by cheap buys. I've tracked 3456a prices on eBay for 4 years, and the price I posted are ranges I've found. The killer is often S&H, as its a wide and heavy beast, but luckily its short and flat. Its about the perimeter of a small coffee table, which makes it easy to stack instruments ontop of it.

Long ago, this website tracked prices for a lot of good gear, but its seems to be orphaned. But it still gives what hi-lo used to be before 2008, as well as Tektronix.





Many links are broken but you can drill down without a problem.

--- In hp_agilent_equipment@..., Bob Albert <bob91343@...> wrote:

Where can you buy one of these for $50?

Bob


FS: Oval power cords

 

Purchased some extra NOS Belden power cords for the older HP equipment a while back, offering some for sale as I will probably never need this many.
$12/ea including shipping within the conUS, or 5 for $55.
Still can't believe how much they go for on ePay! (The reason I snagged these at the time).

Regards,
Chris kc9ieq

Sent from my outdated iPhone wireless thingy


Re: HP-608A Repair

 

A DC measurement across the output connector should show infinity, which is a good sign, as it shows that neither the 2 caps. nor the coax-cable are shorted. The caps. are in series with the pick-up loop, the resistors and the centre of the coax-cable.

Merchison

On 2012-12-23 12:34 AM, Max Robinson wrote:
A DC measurement across the output connector should show
53 ohms. I know this is not a standard value but that is what the diagram
says. My 608 shows infinity. That's not a good sign.

Regards.

Max. K 4 O DS.


Re: HP-608A Repair

Jeremy Nichols
 

This is a common failure when someone using the 608 to test a transceiver accidentally transmits into the 608. Use of the HP-11509A Fuseholder will prevent such failures.

Jeremy
N6WFO

On 12/22/2012 9:34 PM, Max Robinson wrote:

It's really very simple. The center conductor of the connector goes through
a 53 ohm resistor to one side of the pickup coil, the other side of the
pickup coil goes to the sleeve. The center conductor also goes through the
series combination of a 53 ohm resistor and a variable capacitor to the
sleeve. The capacitor is designated as c37. The two resistors are
designated as R58 and R59. I can't find these listed in the parts list I
suspect that HP probably sold the entire attenuator assembly as a single
replacement part. A DC measurement across the output connector should show
53 ohms. I know this is not a standard value but that is what the diagram
says. My 608 shows infinity. That's not a good sign.

Regards.

Max. K 4 O DS.


Re: HP-608A Repair

 

On 12/23/2012 12:34 AM, Max Robinson wrote:

It's really very simple. The center conductor of the connector goes
through
a 53 ohm resistor to one side of the pickup coil, the other side of the
pickup coil goes to the sleeve. The center conductor also goes through
the
series combination of a 53 ohm resistor and a variable capacitor to the
sleeve. The capacitor is designated as c37. The two resistors are
designated as R58 and R59. I can't find these listed in the parts list I
suspect that HP probably sold the entire attenuator assembly as a single
replacement part. A DC measurement across the output connector should
show
53 ohms. I know this is not a standard value but that is what the diagram
says. My 608 shows infinity. That's not a good sign.

Regards.

Max. K 4 O DS.

Email: max@... <mailto:max%40maxsmusicplace.com>
/snip/

I was contacted by one of the readers of this thread who pointed me to a
pdf manual for the 608D. It's at


The description and part number of the resistors is as follows:
53.3 ohms ±1% 1/8 Watt carbon film. HP p/n 0721-0006. They list the mfr
as 19701, but I can't find a code-to-name list anywhere in that manual
to say
who the manufacturer actually is.
I'd guess that if you can somehow find a source of 1/8 W 1% carbon film
resistors
made by anybody, they would work. (I believe these are old-style
resistors with
axial leads, not surface mount or anything weird.) I think the warning
about not
overheating them is to keep them from changing value.
The capacitor, C69, however, from the description, is definitely weird!

--doug, WA2SAY



[Non-text portions of this message have been removed]


Re: HP 3456a was Re: HP 3478A general questions

 

I have to say this response has really given me the most food for thought. It provided exact details about the models and the pro's and con's of each. It also made me realize something important. There is no reason I have to buy one of these devices and run out and get it calibrated. I can make sure it's readings are within reason and take it out later on to get it calibrated. My main goal is resolution, so I can have that now and have the accuracy and the resolution later. It will also give me time to clean up and gain confidence in the device before spending more money on it.

Very good feedback, Thanks.

Jeff

On 12/22/2012 10:17 AM, marvgozum wrote:



The 3456a was the most accurate in the HP fleet when introduced in 1981. The model line was 345x, so in its heritage was the 3455, 3457, and finally the ersatz standard DMM today, introduced in ~1989, the 3458. HP introduced a high accuracy ADC method, multislope II, with 3456a, you'll see marked change in posted accuracy versus the 3455a. The 3458a further improves that technique. In summary, if you're looking for accuracy these series of DMMs are one to shoot for.

There are pros and cons to the discontinued models, 56a and 57a. For example, The 56a is as good as the 57a in DCV at 1 year ppm, while the 57a is 10x more stable in 1yr as a ohmmeter. The 57a uses all close case calibration and has more unobtanium in its parts list, versus the 56a, which was the last to use all pots based calibration. The basic 57a reads only to 300Vrms maximum, but also has a current mode.

As for prices, the low end is similar on both, about $50 without shipping, but on the high end with calibration, the 56a runs to $300 while 57a about $700.

The 3478a was a general purpose working meter, ~10x less accurate in DCV than the 3456a.

As mentioned earlier you can use a resistor network; dekaviders or KV bridges in DC, to generate other DC voltages needed to check ranges, but maintaining stability in the reference enough to check the meter's accuracy, is the challenge. Its enough a whole new forum discusses this issue alone, volt-nuts@... <mailto:volt-nuts%40febo.com>. A similar problem exists for AC.

To calibrate a meter to realize its best accuracy requires grasping key concepts in metrology. A simplest approach is to take your DMM and compare its readings against a known good DMM. The reference DMM should ideally be 4x to 10x more accurate than the DUT. This is the concept of TUR, test uncertainty ratio. When making comparisons, say voltages, the test volt must be stable and precise enough during the transfer, for example if you are checking the 1V range between DMMs, you need to generate a 1V voltage stable to 1uV resolution or less, 1.000 000 VDC. Without a high level of thoroughness and stability, your 6.5 digit DMM will have increasing uncertainty as you get to its LSD.

On the low end of the range, special conditions apply that are fully covered in Keithley's low level measurement handbook, and these need to be controlled once dealing with uV, mohms, and uA adjustments.

Lastly, one has to decide if the effort described in volt nuts is worth it, or should one just take your DUT to the calibrator periodically for $100-$150 to insure your 6.5 digit DMM is as good as you expect it to be.

--- In hp_agilent_equipment@... <mailto:hp_agilent_equipment%40yahoogroups.com>, Jeff Machesky <jeff@...> wrote:

So it's about the time of year where I'm allowed to buy one piece of
test gear and I'm thinking I want a bench meter with higher
resolution. The best I've got so far is the Tek 2465BDM which has a
20,000 count 0.03% accuracy DMM. It's however a little out of cal and
I just don't want to spend the cash to get it back in perfect spec.
It's also a bit crazy to have a scope running just to use the DMM.
This brings me to the 3478A. While I could technically afford to buy a
new 120,000 count DMM I just can't justify the cost for the hobby type
work that I do. Plus I love working with the older gear and saving
money.

Couple quick questions about the 3478A:

Can one range be calibrated at a time? This is not highly clear in the
service manual. Can I calibrate just the 3V range for example?
Are they reliable, or will I have to fix up most of what I would buy
on say ebay?

The problem I have with the meter on my Tek is that you have to run the
entire cal for each measurement type, If you start with DC you have to
go all the way up the range from millivolts up to several hundred
volts. As we all know it's easy to get a precision calibrated low
voltage references such as a DMM check board, it's another thing to get
500 volts or higher reference at the spec of the meter.

So is this really a meter to buy if your willing to send it off to be
calibrated or have a calibration transfer standard meter that's above
the specs of the 3478A? In other words if all you have is just a few
standards are you wasting your time?

Has anyone had any luck using a precision reference, some precision
resistors and an op-amp to generate other reference voltages with any
degree of accuracy? Also has anyone found a good source of say 500
volts 60 Hz AC? The right audio amp and a sig gen can get you some of
the other odd cal standards. High voltage DC is just so much easier to
obtain then a nice high voltage sine wave.

Sorry for the random questions, and thanks for any replies. The
54111D is still running great. My thanks again for everyone who helped
me fix the ROM's.

Jeff