Re: HP PN 04194-81803 What is it?
It is an analog to digital converter from the 4195A.
Sent: Wednesday, March 5, 2025 9:32 AM
To: [email protected]
Subject: Re: [HP-Agilent-Keysight-equipment] HP PN
By
Francesco Ledda
·
#150690
·
|
Shipping overseas?
HP list,
I have attempted to assist a current HP list member in Germany with boards
from a 8660C.
The Post Office is telling me they cannot ship electronics to Germany.
(From the US). Parts,
By
Gary Hitchner
·
#150689
·
|
Re: HP PN 04194-81803 What is it?
Francesco,
From the Keysight parts finder, the picture shows that part number has several discrete components on a circuit board. Three ICs, and numerous R and C. No commercial equivalent but might be
By
Steve - Home
·
#150688
·
|
Re: HP 3403C True RMS Voltmeter display fault
50+ years ago, 400 Hz power supply plugged into 60 Hz source. Runs for a few minutes and then throws its guts all over the bench. Back before the days of power supplies that work from 48 - 440 Hz, 100
By
Steve - Home
·
#150687
·
|
Re: HP 3403C True RMS Voltmeter display fault
Yep, I have seen the fireworks in the lab, too. Been a while, fortunately. More than 25 years ago we had a technician who installed a few tant caps backwards into a VMEbus chassis. Blam, molten
By
Jim Ford
·
#150686
·
|
HP PN 04194-81803 What is it?
I am trying to find out if there is an equivalent commercial PN to replace an 04194-81803.
Thank you, Francesco Ledda
By
Francesco Ledda
·
#150685
·
|
Re: HP3325A Trouble, Calibration FAIL 2 & 3
Meanwhile i found a bad (high resistance) hf transistor Q219, located in the amp. It is a MM4018, which is not available. Can I use a 2N5160?
This is from the specs a similar type: PNP, 40V,hfe 10,
By
Ralf Rosche
·
#150684
·
|
Re: Manual for HP 466A amplifier
Thanks to all for the information on the HP-466A!
Hope I can help others from my HP experience ('77-'2001).
By
WB2UAQ
·
#150683
·
|
Re: HP 478 A? Thermistor mounts
I have, as listed in the 1994 Options List, an HP-478A-H63, which is spec'd
to operate from 100 KHz to 1 GHz and it really does respond that low and
lower. It belonged to HP Division 4 (does anyone
By
Jeremy Nichols
·
#150682
·
|
Re: HP 478 A? Thermistor mounts
Hi David? Thank you found the information so made a new file HP 478A? and put the information in there
Regards :Paul
By
Paul Bicknell
·
#150681
·
|
Re: HP 3403C True RMS Voltmeter display fault
Hey again,
Here are the photos. The one with the big blob is the blown one. It does melt with the iron so it's indeed solder. Thanks for the info on possible replacements.
By
AArnaud
·
#150680
·
|
Re: HP 478 A? Thermistor mounts
Here are the 478A options listed in the 1994 options list, that is on the web somewhere.
David
By
factory
·
#150679
·
|
HP 478 A? Thermistor mounts
Re HP 478 A Thermistor mounts could someone –please direct me to a listing of the different options for example HP 478 A - H75
also interested in any one that uses the H75 option
Regards Paul
By
Paul Bicknell
·
#150678
·
|
Re: HP 3403C True RMS Voltmeter display fault
Ok, tgat's good to know.
DaveD
KC0WJN
[email protected]> wrote:
By
Dave Daniel
·
#150677
·
|
Re: Manual for HP 466A amplifier
I have a scanned copy.? It's not great, and I don't remember where I got it, but I've attached it anyway...
~~
Mark
By
Mark Moulding
·
#150676
·
|
Re: E5052A Power On Test (3.3V Bus Supply)
Isn’t one solution to side modes using multiple (fiber) loops of different lengths, such that fewer frequencies fit?
(Borrowed thinking in terms of laser resonators and etalons where sometimes
By
Ed Marciniak
·
#150675
·
|
Re: HP 3403C True RMS Voltmeter display fault
The OSCON type capacitors aren’t available in higher voltage ratings.
________________________________
Sent: Tuesday, March 4, 2025 11:30:05 AM
To: [email protected]
By
Ed Marciniak
·
#150674
·
|
Re: HP 3403C True RMS Voltmeter display fault
I'll add that since, in general, modern caps are smaller than vintage caps,
make sure the the voltage rating of the new cap provides sufficient margin
over the voltage rail. For instance, Tektronix
By
Dave Daniel
·
#150673
·
|
Re: HP 3403C True RMS Voltmeter display fault
Arnaud, you can replace Ta caps with alternatives in most applications, so there's no need to spend too much for an original type. Usually ones that burn out are in supply bypassing spots, which are
By
ed breya
·
#150672
·
|
Re: HP 54512B (pair) repair
Mike,
No worries, everything I'm doing is at my own risk and expense. I welcome
any ideas - Razvan, Jim thanks to you too!
Radu
[email protected]> wrote:
By
Radu Bogdan Dicher
·