Re: 5086-7337 *Yig Oscillator issue*
Dear, RPG stand for "Rotary Pulse Generator". This is the endless knob on front panel for changing parameters. Regards.
By
Roberto Vegliach
·
#133304
·
|
Re: 5086-7337 *Yig Oscillator issue*
Don - Thanks - I have not played much with the 8350 plug YIGS (mostly 834X stuff) - good information to know. I do find it a bit surprising though. Reasoning is as follows, correct me if I'm wrong: 1)
By
Bruce
·
#133303
·
|
Re: 5086-7337 *Yig Oscillator issue*
Ruben - The RPE is likely to be an easy fix - there is an incandescent lamp that illuminates the rotating defraction grid. It can be replaced by either another lamp (available on E-bay and other
By
Bruce
·
#133302
·
|
Re: HP 8595E SA DB-25 RS232 is not functioning.
Thank you for response. The translator is an old Chinese one blocked by Windows but there is a workaround to install older drivers and make it work. That is what I had available at the time. I did
By
candrian
·
#133301
·
|
Re: HP 8595E SA DB-25 RS232 is not functioning.
Hi. Who's chipset is used in your USB-Serial adapter??? If it is FTDI, good.? Anything else, especially Prolific, can be "troublesome" to say the least.? Unless you take a lot of extra steps and
By
Dave_G0WBX
·
#133300
·
|
Re: 5086-7337 *Yig Oscillator issue*
Hi from an earlier posting there might be confusion PCB stands for printed circuit board But RPG is a new one for me are you referring to the plug in Key pad ? As far as I can remember there is not a
By
Paul Bicknell
·
#133299
·
|
Re: 5086-7337 *Yig Oscillator issue*
No CCT Dues NOT stand for calibration constants CCT Stands for Circuit Diagram in the old days HP included them in the Service manual later on they were sold separately as CLIPS component level
By
Paul Bicknell
·
#133298
·
|
Re: 5086-7337 *Yig Oscillator issue*
Hello again: I assume that 'CCT' stands for calibration constants. I haven't that table. As I told you, I have only the OSM available on Keysight site, which is the same in the files section here.
By
RubenRubio
·
#133297
·
|
Re: 5086-7337 *Yig Oscillator issue*
Absolutely yes! I have been there on this issue. Even a brand new one from the factory may not give an output, or a very low output until it is aligned with the proper endpoint drive voltages. Don
By
Don Bitters
·
#133296
·
|
Re: 8566B attenuator fix
Comparing the older electromechanical attenuators with the newer circuit card attenuators, I saw several failures on the circuit card. I never saw a schematic for the card and could not pry the info
By
Don Bitters
·
#133295
·
|
Re: 8566B attenuator fix
The 2 push rods on each attenuator/solenoid assembly have ball and notch rings under the ball that fits into a keyhole slot on the contact flap. The key to disengaging the push rod from the flap is
By
Don Bitters
·
#133294
·
|
Re: HP 8595E SA DB-25 RS232 is not functioning.
Your full-handshake hookup is correct, although it is missing the analyzer DTR to pc DSR (paralleled with pc CTS), but since you are configuring for no-handshake, it may not matter. You say the
By
k1ggi
·
#133293
·
|
Re: HP 8595E SA DB-25 RS232 is not functioning.
You can try the loopback handshaking cable if the interface can work without needing handshake: https://en.wikipedia.org/wiki/Null_modem Ozan
By
Ozan
·
#133292
·
|
HP 8595E SA DB-25 RS232 is not functioning.
Hi all, I want to connect my HP 8595E to my PC using the only available port it has, RS232 via a DB-25 connecter (option 023). As i couldn't find out the correct pinout i managed to make a quick
By
candrian
·
#133291
·
|
Firmware for HP 54200D
Hi all. Sorry it has been a while since I have been on here - life, you know, can get in the way of fun sometimes. Anyway, I have been feeling sorry for this puppy for a while and finally bought it to
By
medasaro
·
#133290
·
|
Re: 8566B attenuator fix
It¡¯s pulse drive. There are MM74HC123 dual monostable multivibrators doing the pulsing and the ULN2003A provides the high current drive off of those pulses. I have an issue with some of this on mine
By
bagojfalvibagoj@...
·
#133289
·
|
Re: 8566B attenuator fix
Don - I did not see any of the leaf switches (which interrupt the current after switching) in this attenuator - do they rely on a current limit to protect the coil or is the drive a pulse ? Did I
By
Bruce
·
#133288
·
|
Re: 8566B attenuator fix
The attenuator drive is on A16 brd. in the RF section. It is a current drive to the solenoids in the attenuator. The attenuator is 10dB, 20dBa, 20dBb, 20dBc sections. The attenuator steps are: 0dB =
By
Don Bitters
·
#133287
·
|
Re: 8490XX step attenuator mating connector
Hello JF , It is used on the driver cord 11764-60004 ( but I don't have the parts list of this cord... ) 73s Eric
By
f1ghb
·
#133286
·
|
Re: 8490XX step attenuator mating connector
It looks like IDC - should be easy to find. Cheers@! Bruce Quoting "jfphp via groups.io" <jfphp@...>:
By
Bruce
·
#133285
·
|