Peter,
toggle quoted message
Show quoted text
When FM Measurement is selected, the display will show peak deviation. The steps are outlined in the OP Manual under Section III Operation, Getting Acquainted. Dennis -------------------------------------------- On Mon, 10/16/17, Peter Gottlieb hpnpilot@... [hp_agilent_equipment] <hp_agilent_equipment@...> wrote:
Subject: Re: [hp_agilent_equipment] 8901B Power level zeroisation To: hp_agilent_equipment@... Date: Monday, October 16, 2017, 10:11 PM Ok, so that worked, but what does percent mean for FM?? kHz deviation would be more useful. Peter On 10/6/2017 9:25 AM, Dennis Czelusniak czelusniakd@... [hp_agilent_equipment] wrote: > > > E1 is invalid key sequence if calibration is pressed while not in AM or FM. > 8480 series sensors work with your unit, high priced 11722A not required. > If you have option 10, connect calibrator output to input, select AM, push > calibrate and wait, takes about 20 seconds, do the same with FM, display > reading will be 99.0 to 101.0% > > Dennis > > -------------------------------------------- > On Fri, 10/6/17, matt.ames7bc@... [hp_agilent_equipment] > <hp_agilent_equipment@...> wrote: > > Subject: [hp_agilent_equipment] 8901B Power level zeroisation > To: hp_agilent_equipment@... > Date: Friday, October 6, 2017, 2:57 AM > > > > > > > > > > > > > > > > > > > > > > > > > > > > > > > Hi all, > I have an 8901B here that I use to > measure power with, using the internal power detector and > *not* the fancy external 11722A sensor. > When I have nothing connected to the > input, and I press the zero key, I get an Error > 21. > Does anyone know what I am doing wrong, > and how to zero the internal power level > sensor? > Cheers, > Matt > > > > > > > > > > > > > > > > > > > > > > ------------------------------------ Posted by: Peter Gottlieb <hpnpilot@...> ------------------------------------ ------------------------------------ Yahoo Groups Links Traditional ? ? (Yahoo! ID required) ? ? hp_agilent_equipment-fullfeatured@... |