I got as far as step 5-5, YO retrace compensation.
I am using an HP 8566A Spectrum Analyzer but cannot see the signal shown.
I assume the top trace is the chan B input sweep from the 8350B, I see that and can delay along it.
The bottom trace would be the video from the 8566A but which video. There is one jumpered at the back but if disconnected there is no display on the 8566A. The other is the plotter video and I doubt it is this one, it does nothing anyway. I suppose it does not matter that I disconnect the jumper, however I see nothing on the ¡®scope but a straight line. I get that I am looking for the 7 GHz switch point that shows up at 3.5 GHz because the Aux is half freq. What should the span be?
?Hi Peter, I have done this calibration several ?time , I think ?the A vs B is the X-Y display on the scope , I was using the HP 1740A oscilloscope to do this. The intensity must set to the high position because it is difficult to see the crest of the ramp to 0V.