Ozan,
Thank you for the tips.? My observations:
1. I was able to verify +15VF, +5A and +10UF voltages directly on A7.? I need to get an extender to really dig into it :-(
2. In the system diagnostics, +10V Reference (presumably +10UF?) is quite unstable
3. In verifying +10UF, I measured at one end of R105, which connects +10UF to V(REF)+ of the DAC in the third converter ref_cal circuit.? Not knowing which end was which, it was easy enough to find the +10UF (and checking with an o-scope it's quite stable, unlike the system diagnostic), but as it happens the other side of R105 has a very small voltage on it (less than 100 mV).? This doesn't seem right to me.??
As I await my extender, any further insights would be appreciated!