Re: low amplitude readings below 10MHz on a 8591E


 

¿ªÔÆÌåÓý

Hi

?in your case ?from 238-247

if you haven¡¯t done it already? ..201 onwards? ?backing up correction constants

?

One thing to check is the level of your LO feedthrough amplitude ?..the signal at 0hz on the left hand edge of the display

Do frequency 0hz? span 1mhz¡­ and see what the level of the LO is? --good ones are around -10db or better.

?

By the way you¡¯re 8591? is a bit different on the front end than the 8590e series , yours is more prone to mixer failure but can be fixed .

Lots of info on you tube

?

Regards

Mike g4wam

?

?

?

?

?

?

From: [email protected] [mailto:[email protected]] On Behalf Of KeesHofwegen
Sent: 18 March 2023 17:47
To: [email protected]
Subject: Re: [HP-Agilent-Keysight-equipment] low amplitude readings below 10MHz on a 8591E

?


got it and will start reading.?

Any recommendations about what paragraph?

?

Gr Kees

?

?

Op za 18 mrt 2023 om 18:19 schreef Mike Lockley via <mikelockley=[email protected]>:

You need 08590-90316? from? 2001

8590 series Service Guide

?

Are large tome but worth downloading or buying

¸é±ð²µ²¹°ù»å¡¯²õ

Mike g4wam

?

?

From: [email protected] [mailto:[email protected]] On Behalf Of KeesHofwegen
Sent: 18 March 2023 14:41
To: [email protected]
Subject: Re: [HP-Agilent-Keysight-equipment] low amplitude readings below 10MHz on a 8591E

?

Hi,

Are we talking about the user manual 08590-90301, because there is not much cal data in it

other than the normal stuff. I also have the installation, verification & Operation manual.

Or are there other manual I need?

?

Gr. Kees