Hi ?in your case ?from 238-247 if you haven¡¯t done it already? ..201 onwards? ?backing up correction constants ? One thing to check is the level of your LO feedthrough amplitude ?..the signal at 0hz on the left hand edge of the display Do frequency 0hz? span 1mhz¡ and see what the level of the LO is? --good ones are around -10db or better. ? By the way you¡¯re 8591? is a bit different on the front end than the 8590e series , yours is more prone to mixer failure but can be fixed . Lots of info on you tube ? Regards Mike g4wam ? ? ? ? ? ?
toggle quoted message
Show quoted text
From: [email protected] [mailto:[email protected]] On Behalf Of KeesHofwegen Sent: 18 March 2023 17:47 To: [email protected] Subject: Re: [HP-Agilent-Keysight-equipment] low amplitude readings below 10MHz on a 8591E ? got it and will start reading.?
Any recommendations about what paragraph? ? You need 08590-90316? from? 2001 8590 series Service Guide ? Are large tome but worth downloading or buying ¸é±ð²µ²¹°ù»å¡¯²õ Mike g4wam ? ? From: [email protected] [mailto:[email protected]] On Behalf Of KeesHofwegen Sent: 18 March 2023 14:41 To: [email protected] Subject: Re: [HP-Agilent-Keysight-equipment] low amplitude readings below 10MHz on a 8591E ? Hi, Are we talking about the user manual 08590-90301, because there is not much cal data in it other than the normal stuff. I also have the installation, verification & Operation manual. Or are there other manual I need?
|