¿ªÔÆÌåÓý

Re: HP-8753E Power Cycles at Start Up


 

¿ªÔÆÌåÓý

Peter:

After putting in a new SMPS (Newer SMPS), I have had no failures. I would like to repair the SMPS I pulled but it has not failed yet. This is why I think its a current sense problem.

Rich

On Aug 29, 2022, at 4:56 PM, peter bunge <bunge.pjp@...> wrote:

?
Rich,
Mine are all Sprague and the more I test the?less I think they?are the?problem,?but.....? I should test one.
Right now it is running rock solid.
Has your's failed again? You said you had eliminated A8 as the problem? I have seen dipped tantalums short but they stay shorted.
Peter

On Mon, Aug 29, 2022 at 4:38 PM Rich Miller via <av8torrich=[email protected]> wrote:
Peter:

While I cannot answer your question directly, the capacitor's in mine are Nippon¡¯s. I have seen those last forever without fault. They were sparing no expense in this department when making the supply.?

Off the top of my head, I can¡¯t think of anything I have seen inside this supply that would short intermittently and then recover.?

I had to leave the office, so I am not in front of the supply any longer.?

Rich


Join [email protected] to automatically receive all group messages.