Keyboard Shortcuts
ctrl + shift + ? :
Show all keyboard shortcuts
ctrl + g :
Navigate to a group
ctrl + shift + f :
Find
ctrl + / :
Quick actions
esc to dismiss
Likes
- Tinysa
- Messages
Search
Re: Linux Mint USB
#tinysa
Thanks Jim, I did that, the program opens but won't connect to the TinySA. I get this in the terminal window 2023-01-18 02:08:23,923 - TinySASaver.TinySASaver - ERROR - Tried to open /dev/ttyACM0 and failed: [Errno 13] could not open port /dev/ttyACM0: [Errno 13] Permission denied: '/dev/ttyACM0' For some reason in the TinySA program the USB port is /dev/ttyACM0 but there is no communication with the unit and I did rescan and I get /dev/ttyACM0 again no communication when I press Connect. I have the TinySA Ultra on the way from RL Electronics and would like for that to work. Thanks Ed kj4fgi On Tue, Jan 17, 2023 at 12:30 AM Jim Allyn - N7JA <jim@...> wrote: I am able to use both my tinySA and my tinySA Ultra with Linux Mint 20.3 Cinnamon and tinySA-saver 0.2.2.? Are you using the latest version of tinySA-saver from github? |
|
Re: "CALIBRATE above 5.34GHz" throws a "Signal level too low or not on frequency" error
#bug_report
Forgot to mention this wiki entry that explains how calibration above 5.34GHz is done:?
-- For more info on the tinySA go to https://tinysa.org/wiki/ |
|
Re: My tinySA Ultra Signal level Critical
#ultra
Simplest is not to enable the LNA and everything will be correct
-- For more info on the tinySA go to https://tinysa.org/wiki/ |
|
Re: "CALIBRATE above 5.34GHz" throws a "Signal level too low or not on frequency" error
#bug_report
Did you input a 5.34GHz signal between -50 and -10dBm when calibrating above 5.34GHz?
-- For more info on the tinySA go to https://tinysa.org/wiki/ |
|
"CALIBRATE above 5.34GHz" throws a "Signal level too low or not on frequency" error
#bug_report
I just purchased a tinySA from one of the vendors listed at tinysa.org. When running the "CALIBRATE above 5.34GHz", the test throws a "Signal level too low or not on frequency" error.
?
"CALIBRATE 100kHz to 5.34GHz" passes.
?
Did Config/Clear Config/Config/calibration several times using both provided leads, resulting in same error when calibrating above 5.34GHz.
?
Version: tinySA4_v1.4-40-g2f63e1c
Build Time: Jan 13 2023 - 17:56:59
HW Version: V0.4.5.1 (167)
?
Any ideas on how to proceed prior to returning it to the vendor? |
|
Re: My tinySA Ultra Signal level Critical
#ultra
If I temporarily disable LNA, will all my measurements still be correct?
|
|
Re: File /TinySA Ultra Menu-Tree Chart - v1.4-40 - Updated January 16, 2023.vsdx uploaded
#file-notice
¿ªÔÆÌåÓýGreat job, David! That's going to be very helpful. Thank you for all the work that went into it.On 1/17/2023 12:51 PM, Group
Notification wrote:
|
|
Re: My tinySA Ultra Signal level Critical
#ultra
I forgot to mention that you can trace the signal through the LNA using a scope
-- For more info on the tinySA go to https://tinysa.org/wiki/ |
|
File /TinySA Ultra Menu-Tree Chart - v1.4-40 - Updated January 16, 2023.vsdx uploaded
#file-notice
Group Notification
The following files and folders have been uploaded to the Files area of the [email protected] group. By: David Massey <wd4owa@...> Description: |
|
Re: File /Visio-TinySA Ultra Menu-Tree Chart - v1.4-40 - Updated January 16, 2023.pdf uploaded
#file-notice
You're welcome!? I'm going to post the Visio file in the FILES sections today for those that want to take what I've done and expand upon it.
|
|
Re: File /Visio-TinySA Ultra Menu-Tree Chart - v1.4-40 - Updated January 16, 2023.pdf uploaded
#file-notice
Mr. Massey, huge thank you for this document. It is of tremendous help.
-- Steve Johnson? ? ? K0OIL |
|
Harmonic measurement. Question for Eric
Eric,
When I measure harmonics I always use? well defined high pass filters that suppresses the fundamental by >50db. The commercial spectrum analyzer I use at work lets me input the attenuation of the fundamental as a value in db. This permits the measurement of harmonics that are very small (-110 to -120dbc) without overloading the front end. Would it be possible to do with the ultra? It should be just a mathematical function that doesn't impact the function of the analyzer it self. The read out would be the reduced fundamental + offset , and harmonics in dbc to the combined fundamental value. John |
|
Re: My tinySA Ultra Signal level Critical
#ultra
There is either something wrong with the first switch or with the LNA
The new FW detects this and refuses to calibrate -- For more info on the tinySA go to https://tinysa.org/wiki/
|
|
Re: My tinySA Ultra Signal level Critical
#ultra
When I turn off LNA? I generate CAL out at 30Mhz it work fine .
But I turn on LNA look like this. I backed to the old firmware. |
|
Re: My tinySA Ultra Signal level Critical
#ultra
After I update firmware?
Connector work normal |
|
Re: My tinySA Ultra Signal level Critical
#ultra
With the latest FW the selftest will do a better check on the LNA
-- For more info on the tinySA go to https://tinysa.org/wiki/ |
|
Re: New TinySA-ULTRA Case (3D print) on Thingiverse
#ultra
On 16/01/2023 21:03, Larry McElhiney via groups.io wrote:
Hi Folks,Where can this be bought, please, Larry? Not that it's likely to be economic for the UK, though. I did search and find: but no purchase price. 73, David GM8ARV -- SatSignal Software - Quality software for you Web: Email: david-taylor@... Twitter: @gm8arv |