I just purchased a tinySA from one of the vendors listed at tinysa.org. When running the "CALIBRATE above 5.34GHz", the test throws a "Signal level too low or not on frequency" error.
?
"CALIBRATE 100kHz to 5.34GHz" passes.
?
Did Config/Clear Config/Config/calibration several times using both provided leads, resulting in same error when calibrating above 5.34GHz.
?
Version: tinySA4_v1.4-40-g2f63e1c
Build Time: Jan 13 2023 - 17:56:59
HW Version: V0.4.5.1 (167)
?
Any ideas on how to proceed prior to returning it to the vendor?