On Mon, Aug 9, 2021 at 09:22 AM, hwalker wrote:
One issue I noted while using the TABLE menu is that the table values under trace 3 (red trace) have a couple of presets that don't appear to be intentional (see below).? When I tried to disable the [0Hz 108zdBm] button a "FATAL ERROR" message appeared on the display screen and required me to power cycle the tinySA.?
new beta FW release that should solve this bug.
The table is now limited to 6 entries. The 8 in previous FW was incorrect causing memory corruption. Increasing to 8 entries may be possible in the future
?
--
------------------------------------------
For more info on the tinySA go to https://tinysa.org/wiki/