Re: A comment to the TinySA.exe


 

TinySA internal for calculation use measure?data * 32, for not send negative values additional add 128
This 32 multiplier need for not use slow floating point calculations, and give error 1/32 ~ 0,03 in dBm (measure step is 0.5dBm, and other is calibrations error)

And this values used for get correct data in program, need substract 128 and divide on 32