model Ultra calibration signal issues (and issues on the RF input)


 

Regarding the calibrator output, I have been aware of an artifact. Because I have been doing some measurements (related to the new Fault Finding page, I also decided to catch the artifact now):


This is the calibrator at 30 MHz. Every approx. 250 or 400 ?s, the artifact appears.
The calibrator output also has a steep negative slope but a rounded positive slope (HW 0.4.5.1):


The level is okay, fft midscale dBm reference = -55 dBm, and 10 dBm/div. Hence level is -35 dBm - okay.

=== RF input ===

The RF input also has artifacts:


It has been verified that the generator is switched off, but it seems that the RF spectrum analysis operation is not affected. Disclaimer: I'm talking about my defective tinySA with defect LNA and quite sure also defective input switch, but it is hard to believe that this sort of artifact would occur just because of this.

Rob