开云体育

ctrl + shift + ? for shortcuts
© 2025 开云体育

TinySA Ultra+ - Signal generator level error


NelHau
 

System: v1.4-192-g73fe677

While checking the output levels of the signal generator, I noticed an odd behavior at signal levels below -70 dBm.

When checked against a bench top spectrum analyzer, the signal levels were measured to be within about 0.5 dBm from -18.5 dBm to -70 dBm. However, when going from -70 dBm to -80 dBm, the output signal was only about -76 dBm. And then farther below that level the measured output levels were also generally off.

I first when down by -10 dBm steps from -50 dBm to the minimum at -115 dBm, and then went up in +10 dBm steps to -55 dBm.

Photos below.

Nelson
Seattle WA



-50 dBm output




-60 dBm output




-70 dBm output




-80 dBm output




-90 dBm output



-100 dBm output




-110 dBm output




-115 dBm output



-105 dBm output



-95 dBm output




-85 dBm output




-75 dBm output



-65 dBm output




-55 dBm output

Join [email protected] to automatically receive all group messages.