Hi Erik
Thanks for reply regarding C85 - done :-)
?
However, it does not work as expected :-(
There is a lot of noise and in some cases it seems like that an AGC increases the sensitivity when there is no signal.
Have made a test setup where the signal is generated by a signal generator with a 100% AM modulation.
The signal strength at tinaSA is approx. -46dBm
In parallel with tinaSA I have a simple diode detector.
?
First test: Screen dump from tinaSA in Zero-span mode.
Here it is clearly seen that the signal changes between approx. -48dBm and < less than -100 dBm - as expected
?
Second test - attached file: tinySA-ListenMode-50HzAM.pdf
Lots of noise on blue line (Listen signal from tinySA)
?
Third test: - attached file: tinySA-ListenMode-1HzAM
Even more noise on blue line (Listen signal from tinySA)
?
Have tried different settings of REF LEVEL, but it does not affect the signal on the Jack connector in Listen mode.
How to get more clean signal on Jack connector in Listen mode?
?
Thanks in advance, Knud