Re: how to save the calibration data for 6G-12G for SA and SG in the high freq for Ultra


 

You have to enter upfront the actual output level of the generator. Then you select each point, set the generator to the correct frequency and check if the signal is visible and the marker is correctly place on the peak. Then you clock OK to use the measure peak level to correct. When you select the same frequency again the peak should be on the actual output level from the generator
?
harm_lna is the table used when in harmonic mode with LNA enabled.
?
--
Designer of the tinySA