This is what I get with two -15 dBm input signals 20 kHz apart.
?
?
?
With 20 dB of internal attenuation there is still a bit of overloading as the scan improves when I manually set 30 dB attenuation
?
?
You may be running a very old FW which is missing the later performance improvements in the automatic attenuator setting.
--
Designer of the tinySA
For more info go to