Re: H4 + nanovna-saver calibration
And thank you, Stan, for your feedback! 73, Dean W8ZF
By Dean W8ZF · #40119 ·
Re: H4 + nanovna-saver calibration
Thanks Dave, So if the H4 internal cal is *independent* of the nanovna-saver cal, exactly what are you clearing with nanovna-saver's "clear all cals"and when would that be necessary? I ALWAYS either
By Dean W8ZF · #40118 ·
Re: H4 + nanovna-saver calibration
No. They are independent. Dave - W?LEV [email protected]> wrote:
By W0LEV · #40117 ·
Re: H4 + nanovna-saver calibration
Dave, correct information, but you did not address his real question: Is the nanovna-saver 0.4.0 calibration dependent on the nanovna internal cal? I was taught (back a couple of years ago) that
By Stan Dye · #40116 ·
Re: H4 + nanovna-saver calibration
QUOTE: ........ but what happens when the sweep range is different for nanovna-saver than the H4's stored cal? Are results invalid, or are they interpolated (assuming nanovna-saver is sweeping in a
By W0LEV · #40115 ·
Re: H4 + nanovna-saver calibration
Thanks for the replies. There seems to be some disagreement on #2. Appparently for some versions (maybe all), the cal in nanavna-saver IS dependent on the H4 being in a calibrated state before the
By Dean W8ZF · #40114 ·
Re: H4 + nanovna-saver calibration
QUOTE: 1) What is the persistence of calibration done through nanovna-saver? I.e., does it persist if the USB cord is disconnected and the H4 is taken for mobile measurements? Any cal done within
By W0LEV · #40113 ·
Re: H4 + nanovna-saver calibration
#2 is correct. The calibration done in nanovna-saver is solely in the computer. It can't be saved to a slot in the nanovna, partly because -saver has more sweep-point options. You need to do the
By Stan Dye · #40112 ·
H4 + nanovna-saver calibration
I use nanovna-saver with my H4 because it aggregates several useful functions for operating, storing, printing, and exporting measurements. I am confused about the calibration done from
By Dean W8ZF · #40111 ·
Re: H4 calibration
Surprised you said that. Only takes about 2 minutes for the whole procedure. And if you don't care about S21 measurements you can skip the last two steps and only do SOL (short-open-load) Roger
By Roger Need · #40110 ·
Re: H4 calibration
If you use SAVER on a PC or laptop, you can store your verious cals on those and access them later for use. Calibrate once for each frequency range and setup.....store on a "big" PC........recall for
By W0LEV · #40109 ·
Re: H4 calibration
New here. Any tricks to make calibration easier? It is a real hassle as it is.
By KK7OYV · #40108 ·
Re: search for a hidden HF antenna
Re: rain gutter antenna I also have one rain gutter antenna it's about 50 ft long and 8 ft above ground level. For one 40-meter freq i use a 1:1 balun (indoor 10-meter-long radial) plus a matching
By Orrin Winton · #40107 ·
Re: search for a hidden HF antenna
Out of a major forest fire, insurance put us up in a large apartment complex for 2 years. I chose the 3d floor - the top - for RF reasons. Of course, the apartment complex came with its manditory
By W0LEV · #40106 ·
search for a hidden HF antenna
* * *How good can a RAIN GUTTER be on HF bands?* * * *Since moving to a "retirement community" last year I have been searching for a reasonably good HF antenna.* *I have tried a few 10-20 ft
By Barry K3EUI · #40105 ·
Re: Calibration question
Clyde, in HTs used for their purpose, which is portable communications, the body of the HT, and even the body of the operator are all part of the antenna. If you mount a rubber duckie directly on a
By Manfred Mornhinweg · #40104 ·
Re: Calibration question
This can be improved by adding a small Common Mode Choke at the VNA end to help isolate the coax from the case of the device. The frequency being measured also has an impact. *Clyde K.
By Clyde Spencer · #40103 ·
Re: Calibration question
Yes, I have noted that the SWR changes when holding the VNA vs. not holding it, and touching the SMA connector vs. not touching it. The readings generally are better when touching vs. not
By Kenneth Roberts · #40102 ·
Re: Calibration question
Ken, the behavior of rubber duckies depends significantly on the radio they are attached to. So, to get reasonably accurate results when measuring them with the nanoVNA, I suggest that you build a
By Manfred Mornhinweg · #40101 ·
Re: Save change to active trace
make the swr trace active and then press on calib (without recalibrating) and then save to slot 0 (default slot on boot up) ... i do the same if i want other settings to be saved (be it avg setting or
By Siegfried Jackstien · #40100 ·