¿ªÔÆÌåÓý

ctrl + shift + ? for shortcuts
© 2025 Groups.io
Date

Re: The T-Check confusion

 

Hello Kurt,

Do you have a definition of this 10/12 term correction method? I am interested in understanding it. Thanks.

All the best,

Victor


Re: NanoVNA-H "Fail Write" to 16GB SD card #nanovna-h

 

Use a different card. Some manufacturer's cards just don't support the
protocol used by the nanovna firmware.
See this post:
/g/nanovna-users/message/26892


On Tue, Jun 7, 2022, 8:29 AM gfmucci via groups.io <gfmucci=
[email protected]> wrote:

Below are the test results from my 16 GB SD card...

Warning: Only 14753 of 14754 MByte tested.
Test finished without errors.
You can now delete the test files *.h2w or verify them again.
Writing speed: 13.9 MByte/s
Reading speed: 28.2 MByte/s
H2testw v1.4

Next trouble shooting step?







Re: NanoVNA - Signal Generator

 

In Expert settings there is the reference frequency you may change to your like, so I had to decrease that value a little bit (by 45 units afaik) such I now read 300.00000 an my counter.


Re: NanoVNA-H "Fail Write" to 16GB SD card #nanovna-h

 

Below are the test results from my 16 GB SD card...

Warning: Only 14753 of 14754 MByte tested.
Test finished without errors.
You can now delete the test files *.h2w or verify them again.
Writing speed: 13.9 MByte/s
Reading speed: 28.2 MByte/s
H2testw v1.4

Next trouble shooting step?


Re: Nanovnasaver plots

 

Here is the Zsource I derived (using the method of Owen Duffy). I used one
100ohm SMD resistor (99ohm using DVM) soldered between two female SMA PCB
connectors:
[image: afbeelding.png]
Zsource indeed close to 43ohm... So it might be that I can explain the
differences. I really had not expected that Zsource was so different from
50ohms....
Need to think about this all;-)

All the best,

Victor

Op di 7 jun. 2022 om 15:10 schreef Victor Reijs via groups.io
<pe1atn.victor.reijs@...>:

Hello ROger,

I check again and if I use a Zsource of around 43ohm and a Zload as
measured using the S11reflect of port2.
Using that Zsource gives me more or less comparable graphs for the
differently derived Zdut's.
[image: afbeelding.png]
From (Fig. 10)
2 port series-through:
Z2pseries(S11)=(Zsource+Zload)*S11/(1-S11) (red)
Z2pseries(S21)=(Zource+Zload)*(1-S21)/S21 (green)
Z2pseriesrefl(S11)=Zo*(1+S11)/(1-S11)-Zload (blue)
1port reflection:
Z1prefl(S11)= Zo*(1+S11)/(1-S11) (orange)


[image: afbeelding.png]
So I really will need to determine Zsource for my NanoVNA. Will work on
that now.

All the ebay,

Victor


Op di 7 jun. 2022 om 12:27 schreef Victor Reijs via groups.io
<pe1atn.victor.reijs@...>:

Hello Roger,

I try to have zero extension after the calibration planes (at least for
30MHz). So I use SMA connectors with open, closed, load (2-100ohm SMD in
parallel) and through pins, see:

I do of course all my (Excel) calculations in the complex environment.
And
I get the same results as the NanoVNA saver program for all values. I
plotted the |Z|, as It provides enough detail (certainly when having 10%
deviation;-).
For my S21 evaluation; my DUTs are normal resistors soldered on centre of
female SMA connectors (I don't care if the leads are long, all is my DUT)
and for comparing I use low frequencies (to minimise stray effects).

I use a slightly extended formula:
ZDUT(s) = 2*(Zsource+Zload)*((1/S21)-1)
Zload is the one measured with through connection to port 2 (S11
reflection).
Zsource can be manually varied in Excel for the evaluation (default
50ohms,
but can be any complex number).
Besie this, I also have Zo which is the reference of pure 50ohm (used in
the calucations/definitions). My Zo calibration load is 49.7ohm using
DVM.

[image: afbeelding.png]

I need to measure impedances larger than 3 or 4kohm. as I am looking at
choking devices and I hope they have impedances above 5 or more kohm. And
frequencies upto the end of the HF bands (30MHz).
If I look at my DVM comparison with 1port S11 (reflection) results; even
21kohm provides ok-ish results (within 2%, which is good enough for me).

But I still want to understand why my 2port S21 Zduts are off some 9%.

When doing the 2port measurements, I get (as expected) similar impedance
values for impedance measuring Series S11 (minus Zload) and 1 port S11
reflection measurement. So this (S11) part of the 2port measurements can
be
mapped nicely on the 1 port measurements.

Op di 7 jun. 2022 om 01:09 schreef Roger Need via groups.io
<sailtamarack=
[email protected]>:

Next I attached the cable to Port2 (CH1) and measured the complex
impedance R +/-jX fro 10 kHz to 250 MHz. The attached plots show that
there is a considerable change in resistance over this range and an
unwanted capacitive reactance.

So in my case the difference is (using the through connector of my
calibration kit and from 330kHz to 30MHz).
[image: afbeelding.png]

My results are comparable to your 6dB attenuator results (I though have
this dip near low freqeuncies, I don't see this in a 50ohm calibration
load, so this is soemthing of my port2, I think...).

What I have not tested is how the impedance of Port1 (Ch0) varies with
frequency. This is not as easy a measurement and I have not done it.
The
attenuator "trick" could be used here as well but needs further
investigation.
Owen Duffy mentions a method (). I
also
did not do it (I though variate in my Exel to see the results and see if
I
can map the NanoVNA measurmeent with the Excel).


What happens with the S21 series method is that as you increase the DUT
resistance you get more error in the S21 gain AND S21 phase. This
translates to larger differences in the expected complex impedance of
the
DUT.
But 9%!? I am not able to simulate this in my Excel spreadsheet (except
by
lowering Zload or Zsource to some 40ohm)... And would you not see the
same
type of error in S11? (while that one is ok-ish).


I believe there are many sources of error such as the following:
- Port1 and Port 2 impedance is not 50 +j0 as pointed out above
- Stray inductance and capacitance of the test jig have a considerable
effect on the magnitude and phase of the S21 measurement. I have tried
several different test jigs and keep trying to improve. The attached
photo
shows one I use but still needs work because the S11 reference plane
is
too far from the DUT and I get an unwanted phase shift
For now I keep the freqeuncies of interest (for me) low: upto 30MHz.


- To calculate S21 the NanoVNA needs to know the power level generated
on
Port 1. I have a hunch that this power level fluctuates as the Return
Loss
decreases due to a higher DUT impedance.
Something like that must indeed happen.
The behavoir looks to be similar if the Zload is paralleled (so on port2
side) with an impedance (at least that is what I can see/simulate in my
Exel spreadsheet). But I don't think the port2 impedance does not change
much due to Zdut (I really hope!).
I tried to simulate volttage change of port 1 by varying Zsource (so a
change on port1 side): I can get a Zdut(2pseriesS21) equal to
Zdut(2pseriesS11) when Zsource=40ohms.
But the Zrefl (using 1 port S11 measurement) also will be changed due to
this change in Zsource. While that one was measured ok-ish...
So it is something, IMHO, on the port2 side...

I tried a 10dB at port2 to see the effects on Zdut. After recalibration I
get thee for 2port measurements:
[image: afbeelding.png]

The right bottom graphs compares Zdut with 10dB (red/darkblue) and
without
10dB (orange/light blue). No significant change, so still this 9% error
compared to Zdut using S11 reflection measurement...

For more discussion on this topic here are some old post links...
/g/nanovna-users/topic/67738316#7993

/g/nanovna-users/topic/hardware_deficiencies_when/80639862?p=

With care it is possible to get S21 results that compare well with S11
measurements. Here is a post I made earlier on the subject.
/g/nanovna-users/message/24390. But I just use S11
method most of the time for ease of use.
Will digest these links.

Thanks.

All the best,


Victor










Re: Nanovnasaver plots

 

Hello ROger,

I check again and if I use a Zsource of around 43ohm and a Zload as
measured using the S11reflect of port2.
Using that Zsource gives me more or less comparable graphs for the
differently derived Zdut's.
[image: afbeelding.png]
From (Fig. 10)
2 port series-through:
Z2pseries(S11)=(Zsource+Zload)*S11/(1-S11) (red)
Z2pseries(S21)=(Zource+Zload)*(1-S21)/S21 (green)
Z2pseriesrefl(S11)=Zo*(1+S11)/(1-S11)-Zload (blue)
1port reflection:
Z1prefl(S11)= Zo*(1+S11)/(1-S11) (orange)


[image: afbeelding.png]
So I really will need to determine Zsource for my NanoVNA. Will work on
that now.

All the ebay,

Victor


Op di 7 jun. 2022 om 12:27 schreef Victor Reijs via groups.io
<pe1atn.victor.reijs@...>:

Hello Roger,

I try to have zero extension after the calibration planes (at least for
30MHz). So I use SMA connectors with open, closed, load (2-100ohm SMD in
parallel) and through pins, see:

I do of course all my (Excel) calculations in the complex environment. And
I get the same results as the NanoVNA saver program for all values. I
plotted the |Z|, as It provides enough detail (certainly when having 10%
deviation;-).
For my S21 evaluation; my DUTs are normal resistors soldered on centre of
female SMA connectors (I don't care if the leads are long, all is my DUT)
and for comparing I use low frequencies (to minimise stray effects).

I use a slightly extended formula:
ZDUT(s) = 2*(Zsource+Zload)*((1/S21)-1)
Zload is the one measured with through connection to port 2 (S11
reflection).
Zsource can be manually varied in Excel for the evaluation (default 50ohms,
but can be any complex number).
Besie this, I also have Zo which is the reference of pure 50ohm (used in
the calucations/definitions). My Zo calibration load is 49.7ohm using DVM.

[image: afbeelding.png]

I need to measure impedances larger than 3 or 4kohm. as I am looking at
choking devices and I hope they have impedances above 5 or more kohm. And
frequencies upto the end of the HF bands (30MHz).
If I look at my DVM comparison with 1port S11 (reflection) results; even
21kohm provides ok-ish results (within 2%, which is good enough for me).

But I still want to understand why my 2port S21 Zduts are off some 9%.

When doing the 2port measurements, I get (as expected) similar impedance
values for impedance measuring Series S11 (minus Zload) and 1 port S11
reflection measurement. So this (S11) part of the 2port measurements can be
mapped nicely on the 1 port measurements.

Op di 7 jun. 2022 om 01:09 schreef Roger Need via groups.io <sailtamarack=
[email protected]>:

Next I attached the cable to Port2 (CH1) and measured the complex
impedance R +/-jX fro 10 kHz to 250 MHz. The attached plots show that
there is a considerable change in resistance over this range and an
unwanted capacitive reactance.

So in my case the difference is (using the through connector of my
calibration kit and from 330kHz to 30MHz).
[image: afbeelding.png]

My results are comparable to your 6dB attenuator results (I though have
this dip near low freqeuncies, I don't see this in a 50ohm calibration
load, so this is soemthing of my port2, I think...).

What I have not tested is how the impedance of Port1 (Ch0) varies with
frequency. This is not as easy a measurement and I have not done it.
The
attenuator "trick" could be used here as well but needs further
investigation.
Owen Duffy mentions a method (). I also
did not do it (I though variate in my Exel to see the results and see if I
can map the NanoVNA measurmeent with the Excel).


What happens with the S21 series method is that as you increase the DUT
resistance you get more error in the S21 gain AND S21 phase. This
translates to larger differences in the expected complex impedance of the
DUT.
But 9%!? I am not able to simulate this in my Excel spreadsheet (except by
lowering Zload or Zsource to some 40ohm)... And would you not see the same
type of error in S11? (while that one is ok-ish).


I believe there are many sources of error such as the following:
- Port1 and Port 2 impedance is not 50 +j0 as pointed out above
- Stray inductance and capacitance of the test jig have a considerable
effect on the magnitude and phase of the S21 measurement. I have tried
several different test jigs and keep trying to improve. The attached
photo
shows one I use but still needs work because the S11 reference plane is
too far from the DUT and I get an unwanted phase shift
For now I keep the freqeuncies of interest (for me) low: upto 30MHz.


- To calculate S21 the NanoVNA needs to know the power level generated on
Port 1. I have a hunch that this power level fluctuates as the Return
Loss
decreases due to a higher DUT impedance.
Something like that must indeed happen.
The behavoir looks to be similar if the Zload is paralleled (so on port2
side) with an impedance (at least that is what I can see/simulate in my
Exel spreadsheet). But I don't think the port2 impedance does not change
much due to Zdut (I really hope!).
I tried to simulate volttage change of port 1 by varying Zsource (so a
change on port1 side): I can get a Zdut(2pseriesS21) equal to
Zdut(2pseriesS11) when Zsource=40ohms.
But the Zrefl (using 1 port S11 measurement) also will be changed due to
this change in Zsource. While that one was measured ok-ish...
So it is something, IMHO, on the port2 side...

I tried a 10dB at port2 to see the effects on Zdut. After recalibration I
get thee for 2port measurements:
[image: afbeelding.png]

The right bottom graphs compares Zdut with 10dB (red/darkblue) and without
10dB (orange/light blue). No significant change, so still this 9% error
compared to Zdut using S11 reflection measurement...

For more discussion on this topic here are some old post links...
/g/nanovna-users/topic/67738316#7993

/g/nanovna-users/topic/hardware_deficiencies_when/80639862?p=

With care it is possible to get S21 results that compare well with S11
measurements. Here is a post I made earlier on the subject.
/g/nanovna-users/message/24390. But I just use S11
method most of the time for ease of use.
Will digest these links.

Thanks.

All the best,


Victor






Re: NanoVNA - Signal Generator

 

On 6/7/22 5:25 AM, Rich NE1EE wrote:
On 2022-06-07 00:07:-0700, you wrote:

FYI - I made an output voltage measurement based on the "Px" settings (on my H4, fw v1.2 from the release .dfu file, o'scope w/ 50ohm input, CW=10MHz, square wave:
The CW frequency was -520Hz off at 300MHz here, so I had to adjust the 26MHz reference frequency a little bit..
Thanks for posting those. I guess that Auto would somehow adjust the current based on reflected (or thru) values to try to keep the reflected (or thru) values in some range?

520 Hz in 300 MHz seems like a lot...are they all that far off? I should see if I can check mine. That would definitely have an impact on using it for a signal generator.
that's a bit less than 2 ppm. That's pretty good for an inexpensive TCXO.? If you want to hack, you might be able to put a voltage on the trim input, but that could actually make things worse. You're probably better off just programming it for a frequency that is 520 Hz different.


Re: NanoVNA - Signal Generator

 

On 2022-06-07 00:07:-0700, you wrote:

FYI - I made an output voltage measurement based on the "Px" settings (on my H4, fw v1.2 from the release .dfu file, o'scope w/ 50ohm input, CW=10MHz, square wave:
The CW frequency was -520Hz off at 300MHz here, so I had to adjust the 26MHz reference frequency a little bit..
Thanks for posting those. I guess that Auto would somehow adjust the current based on reflected (or thru) values to try to keep the reflected (or thru) values in some range?

520 Hz in 300 MHz seems like a lot...are they all that far off? I should see if I can check mine. That would definitely have an impact on using it for a signal generator.


Re: NanoVNA - Signal Generator

 

FYI - I made an output voltage measurement based on the "Px" settings (on my H4, fw v1.2 from the release .dfu file, o'scope w/ 50ohm input, CW=10MHz, square wave:

Pauto 160mV
P2mA 160mV
P4mA 320mV
P6mA 480mV
P8mA 600mV

The CW frequency was -520Hz off at 300MHz here, so I had to adjust the 26MHz reference frequency a little bit..


Re: NanoVNA-H "Fail Write" to 16GB SD card #nanovna-h

 

On 2022-06-06 18:14:-0700, you wrote:
It is recommended in the NanoVNA beta group that you power off to install or remove SD cards. IIRC some cards have been permanently damaged by "hot swapping".

Roger
Thanks for the tip.


Re: spreadsheet for importing snp files, Excel 2010

 

I updated the spreadsheet with a note, and removed the column that was confusing (just one of my experiments).

--
~R~
72/73 de Rich NE1EE
The Dusty Key
On the banks of the Piscataqua


Re: spreadsheet for importing snp files, Excel 2010

 

On Sun, Jun 5, 2022 at 12:45 PM, AG6CX wrote:


Formula for cell ¡°M4¡± shows = 20 x log10 (0.16), which actually calculates
to 20 x (-0.79588) = -15.917 dB
Actually, it's
-20 x log10 (0.016)
--
~R~
72/73 de Rich NE1EE
The Dusty Key
On the banks of the Piscataqua


Re: Nanovnasaver plots

 

Hello Roger,

I try to have zero extension after the calibration planes (at least for
30MHz). So I use SMA connectors with open, closed, load (2-100ohm SMD in
parallel) and through pins, see:

I do of course all my (Excel) calculations in the complex environment. And
I get the same results as the NanoVNA saver program for all values. I
plotted the |Z|, as It provides enough detail (certainly when having 10%
deviation;-).
For my S21 evaluation; my DUTs are normal resistors soldered on centre of
female SMA connectors (I don't care if the leads are long, all is my DUT)
and for comparing I use low frequencies (to minimise stray effects).

I use a slightly extended formula:
ZDUT(s) = 2*(Zsource+Zload)*((1/S21)-1)
Zload is the one measured with through connection to port 2 (S11
reflection).
Zsource can be manually varied in Excel for the evaluation (default 50ohms,
but can be any complex number).
Besie this, I also have Zo which is the reference of pure 50ohm (used in
the calucations/definitions). My Zo calibration load is 49.7ohm using DVM.

[image: afbeelding.png]

I need to measure impedances larger than 3 or 4kohm. as I am looking at
choking devices and I hope they have impedances above 5 or more kohm. And
frequencies upto the end of the HF bands (30MHz).
If I look at my DVM comparison with 1port S11 (reflection) results; even
21kohm provides ok-ish results (within 2%, which is good enough for me).

But I still want to understand why my 2port S21 Zduts are off some 9%.

When doing the 2port measurements, I get (as expected) similar impedance
values for impedance measuring Series S11 (minus Zload) and 1 port S11
reflection measurement. So this (S11) part of the 2port measurements can be
mapped nicely on the 1 port measurements.

Op di 7 jun. 2022 om 01:09 schreef Roger Need via groups.io <sailtamarack=
[email protected]>:

Next I attached the cable to Port2 (CH1) and measured the complex
impedance R +/-jX fro 10 kHz to 250 MHz. The attached plots show that
there is a considerable change in resistance over this range and an
unwanted capacitive reactance.

So in my case the difference is (using the through connector of my
calibration kit and from 330kHz to 30MHz).
[image: afbeelding.png]

My results are comparable to your 6dB attenuator results (I though have
this dip near low freqeuncies, I don't see this in a 50ohm calibration
load, so this is soemthing of my port2, I think...).

What I have not tested is how the impedance of Port1 (Ch0) varies with
frequency. This is not as easy a measurement and I have not done it. The
attenuator "trick" could be used here as well but needs further
investigation.
Owen Duffy mentions a method (). I also
did not do it (I though variate in my Exel to see the results and see if I
can map the NanoVNA measurmeent with the Excel).


What happens with the S21 series method is that as you increase the DUT
resistance you get more error in the S21 gain AND S21 phase. This
translates to larger differences in the expected complex impedance of the
DUT.
But 9%!? I am not able to simulate this in my Excel spreadsheet (except by
lowering Zload or Zsource to some 40ohm)... And would you not see the same
type of error in S11? (while that one is ok-ish).


I believe there are many sources of error such as the following:
- Port1 and Port 2 impedance is not 50 +j0 as pointed out above
- Stray inductance and capacitance of the test jig have a considerable
effect on the magnitude and phase of the S21 measurement. I have tried
several different test jigs and keep trying to improve. The attached photo
shows one I use but still needs work because the S11 reference plane is
too far from the DUT and I get an unwanted phase shift
For now I keep the freqeuncies of interest (for me) low: upto 30MHz.


- To calculate S21 the NanoVNA needs to know the power level generated on
Port 1. I have a hunch that this power level fluctuates as the Return Loss
decreases due to a higher DUT impedance.
Something like that must indeed happen.
The behavoir looks to be similar if the Zload is paralleled (so on port2
side) with an impedance (at least that is what I can see/simulate in my
Exel spreadsheet). But I don't think the port2 impedance does not change
much due to Zdut (I really hope!).
I tried to simulate volttage change of port 1 by varying Zsource (so a
change on port1 side): I can get a Zdut(2pseriesS21) equal to
Zdut(2pseriesS11) when Zsource=40ohms.
But the Zrefl (using 1 port S11 measurement) also will be changed due to
this change in Zsource. While that one was measured ok-ish...
So it is something, IMHO, on the port2 side...

I tried a 10dB at port2 to see the effects on Zdut. After recalibration I
get thee for 2port measurements:
[image: afbeelding.png]

The right bottom graphs compares Zdut with 10dB (red/darkblue) and without
10dB (orange/light blue). No significant change, so still this 9% error
compared to Zdut using S11 reflection measurement...

For more discussion on this topic here are some old post links...
/g/nanovna-users/topic/67738316#7993

/g/nanovna-users/topic/hardware_deficiencies_when/80639862?p=

With care it is possible to get S21 results that compare well with S11
measurements. Here is a post I made earlier on the subject.
/g/nanovna-users/message/24390. But I just use S11
method most of the time for ease of use.
Will digest these links.

Thanks.

All the best,


Victor


Re: NanoVNA-H "Fail Write" to 16GB SD card #nanovna-h

 

On 06/06/2022 19:54, gfmucci via groups.io wrote:
Version 1.1; Build Dec 21 2021.
Verbatim 16 GB micro SD card.
First formated Fat 32 via Windows 10 system. Wouldn't write from NANO.
Next installed "SD Card Formater" on my desktop. Did a full Fat 32 format. Wouldn't write from NANO.
Made sure to have NANO "off", then installed card, then turned NANO "on." No joy.
Suggestions?
Check the SD card with (for example):



Cheers,
David
--
SatSignal Software - Quality software for you
Web:
Email: david-taylor@...
Twitter: @gm8arv


Re: NanoVNA-H "Fail Write" to 16GB SD card #nanovna-h

 

On Mon, Jun 6, 2022 at 06:07 PM, Rich NE1EE wrote:


I have found (-H4, v4.3, DiSlord firmware v1.2)
1. I can hot swap the SD cards. No mounting, unmounting. I don't rush the
write ops, but it is clearly done when it says it is.
It is recommended in the NanoVNA beta group that you power off to install or remove SD cards. IIRC some cards have been permanently damaged by "hot swapping".

Roger


Re: NanoVNA-H "Fail Write" to 16GB SD card #nanovna-h

 

On 2022-06-06 11:54:-0700, you wrote:

Version 1.1; Build Dec 21 2021.

Verbatim 16 GB micro SD card.
First formated Fat 32 via Windows 10 system. Wouldn't write from NANO.
Next installed "SD Card Formater" on my desktop. Did a full Fat 32 format. Wouldn't write from NANO.
Made sure to have NANO "off", then installed card, then turned NANO "on." No joy.
I have found (-H4, v4.3, DiSlord firmware v1.2)
1. I can hot swap the SD cards. No mounting, unmounting. I don't rush the write ops, but it is clearly done when it says it is.
2. I tried a newer 32 GB card, and no joy. I happened to have an older one, and that worked. So now I am a one-card nVNA guy ;-) and I share it among my nVNAs.
3. There have been various discussions about the SD card requirements, but even if I understood them, the card specs don't seem to list what is needed.

~R~


Re: Inconsistent VSWR readong #calibration

 

I normally use the NanoVNA without the computer connected, and it matches
well with my MFJ analyzer.

When connecting to the computer today, I saw a similar issue when measuring
an HF multiband vertical (seemed to affect frequencies around 7MHz more
than others). The issue was resolved by putting a good ferrite bead on the
USB cable. Evidently, as others have noted, there can be RF on that cable
that needs decoupled.


On Mon, Jun 6, 2022, 4:51 PM Roger Stierman via groups.io <rogerstierman=
[email protected]> wrote:

Is the MFJ running on batteries? Or a wall wart?
Personally seen erroneous readings MFJ on wall wart, batteries are better.
Then check adapter/cable for NanoVNA.
Something is different in the RF path. Search, Grasshopper. WA0VYU

Sent from the all new AOL app for Android

On Mon, Jun 6, 2022 at 5:29 PM, Stefan<spommere@...> wrote:
Thanks.

How do you explain this:

MFJ: 1.8
NanoVNA not connected to laptop via USB-C: 2.7

Forget about the laptop, the USB-C cable, the NanoVNA Saver app .. but
just unplugging/plugging the coax among MFJ/NanaoVNA.

The MFJ is correct; I used it at another site last weekend, and compared
the reading to a $10k spectrum analyzer someone used there.

I went through several cycles of calibrating the NanoVNA, all with the
same result.












NanoVNA-H "Fail Write" to 16GB SD card #nanovna-h

 

Version 1.1; Build Dec 21 2021.

Verbatim 16 GB micro SD card.
First formated Fat 32 via Windows 10 system. Wouldn't write from NANO.
Next installed "SD Card Formater" on my desktop. Did a full Fat 32 format. Wouldn't write from NANO.
Made sure to have NANO "off", then installed card, then turned NANO "on." No joy.

Suggestions?


Re: Inconsistent VSWR readong #calibration

Roger Stierman
 

Is the MFJ running on batteries?? Or a wall wart?
Personally seen erroneous readings MFJ on wall wart, batteries are better.
Then check adapter/cable for NanoVNA.
Something is different in the RF path.? Search, Grasshopper. WA0VYU

Sent from the all new AOL app for Android

On Mon, Jun 6, 2022 at 5:29 PM, Stefan<spommere@...> wrote: Thanks.

How do you explain this:

MFJ: 1.8
NanoVNA not connected to laptop via USB-C: 2.7

Forget about the laptop, the USB-C cable, the NanoVNA Saver app .. but just unplugging/plugging the coax among MFJ/NanaoVNA.

The MFJ is correct; I used it at another site last weekend, and compared the reading to a $10k spectrum analyzer someone used there.

I went through several cycles of calibrating the NanoVNA, all with the same result.


nanoVNA_saver issues #nanosaver

Anne Ranch
 

1. I cannot get "VSWR" to start / display anything reasonable , BUT "return loss" works as expected
2. How do I delete "marker" - say delete all markers? (I hate clutter)
3. moving mouse wheel zooms "vertical display " , how do I "unzoom "? "
4. is the app missing "save" buttons ( any change ) intentionally ?
5. Can the "sweep " be sped-up ? (serial port baud rate ?)