Re: Question: Is it possible to use calibration imperfections to detect the quality of calibration standards??? #calibration


 

Doing a manual optimization using
openC0=300e-15 F
openC1=-250e-24 F
did result in an reduction of the amplitude variations (blue lines, see attached picture)
Changing shortL0 seemed to have a similar impact as openC0 so I do not know how to reduce both amplitude and phase variations (red lines)