¿ªÔÆÌåÓý

ctrl + shift + ? for shortcuts
© 2025 Groups.io

Re: Nanovnasaver plots


 

Thanks Roger for pointing this out. Knowing one's measurement tools is the
basic rule here. Thanks for pointing out.
I have also checked my NanoVNA.
If I measure the Zload of port 2 (using Port 1, so using a calibration
Through connection and measuring port2's S11). I get the following |Zload|
of port 2:
[image: afbeelding.png]
This is ok-ish IMHO (variation some 0.1ohm (less than 0.2%). There is a
deviation from 50ohm (around 05%), but this could be partly due to
inaccuracy calibration Load (measured with DVM: 49.7ohm).
I don't think the Zload (|Z| of port2/load) is problematic.
If I measure metal resistors (averaged between 300 and 1800kHz) using
Reflection and Series S21 I method, I get the below |Z| values. I also give
the R values measured with DVM

[image: afbeelding.png]

[image: afbeelding.png]
So indeed |Zrefl| does not deviate much from the DVM value (<2%). The
|ZseriesS21| though differs quite a lot (~9%) from DVM value and looks to
be proportional to the resistor value.

Where is this 9% error coming from? A 9% error is large! I don't think the
deviation of the port2 load (Zload) from 50ohm is problematic (only some
0.2%). Looking at the Zsourse (the impedance of port 1): would that not be
close to Zload (impedance of port2, I hope/assume...). But that one is
compensated using the SOLTI method, or not?
I understand the missing of 12-term correction is the other problem (as
stated by Owen Duffy: ). Who has a
procedure for this? I could not find this on the web (Owen Duffy refers to
it, but I don't see a procedure).
Something must deviate some 9% from the 'my norm', but what? I tried to
simulate this by varying Zo, Zload, Zsource; but I can't find really
realistic values to cater for the variation see in ZseriesS21 (I need to
bring Zload back to some 40ohm to cater for this 9%, but then some other Z
values (|ZseriesS11|) don't match properly [and it is not measured, see
above!];-).

Any help to understand this is appreciated. Thanks.
So at this moment I agree that 2port Series method does not produce good
values.

All the best,

Victor


Op di 31 mei 2022 om 23:07 schreef Victor Reijs via groups.io
<pe1atn.victor.reijs@...>:

Hello Roger,

Op di 31 mei 2022 om 19:40 schreef Roger Need via groups.io <sailtamarack=
[email protected]>:


The intent of my post was to reply to comments by Victor Reijs and G3TXQ
(sk) that the S11 mode was not suitable for measuring high impedances.
When it comes to the NanoVNA this is not true and actual measurements
like
the one I posted can be used to prove my point.
It is not me but


or


who derives the accuracy.
Can you provide where they go wrong? That would be very helpful. For me.
Thanks.

All the best,

Victor





Join [email protected] to automatically receive all group messages.