Clear your calibration on the device, measure your standards
while the device is in an uncalibrated state.
Essentially check if your standards and device are functioning properly.
toggle quoted message
Show quoted text
On Fri, 4 Feb 2022 at 11:13, Markus Lemke <dl1dsn@...> wrote:
Hi,
I'm facing the same issue with my new NanoVNA V2 and NanoVNA Saver version
V0.3.10 under Win64.
In 9 out of 10 calibration attempts I get the following error message:
¡°two of short, open and load returned the same values at frequency ...¡±
This is really annoying - does anybody have an idea if this is a hardware
or a software issue on my device?
Calibration on the device itself works fine but with the V0.3.10 on the PC
it's a nightmare.
BR,
Markus