Alan,
Why not trying to increase the span by 10mhz at a time to see where you get correct readings? That may help to narrow down the issue.?
On Thu, 1 Oct 2020 at 10:10 AM, Alan ¡°W2AEW¡± Wolke<alan.wolke@...> wrote: Hello DiSlord,
Thank you for the reply.? I tried this procedure, and still have the same problem.? I am not sure why.
When I run through the calibration procedure over a wider span, it works perfectly as expected.
For some reason, only when I repeat the exact same procedure over this narrow span (460-475MHz), I have the problem.
I have tried your process of just doing ISOLN and THRU and saving, with the same result.
(this is on a -H4, hardware version 4.2.1)
I also have your 1.0.39 firmware on my new SAA-2N, and it works perfectly under the same conditions.
Do you have any other suggestions?
Thank you,
Alan
toggle quoted message
Show quoted text
On Thu, Oct? 1, 2020 at 12:11 AM, DiSlord wrote:
This is calibration issue.
Reset device (off/on), load your calibration slot, and not made calibration
reset, do only ISOLN/THRU after press done and save to slot.
PS Firmware allow made apart calibration OPEN/SHORT/LOAD and ISOLN/THRU (if
not reset before).