One caveat to measuring at the end of a long cable is that the calibration and measurement may need to have the sweep speed reduced. A round trip reflection cal or measurement with a 100 M transmission line with a velocity factor of .66 will be roughly a microsecond.
This can be controlled on a commercial VNA - I'm not sure if the nanovna / saver has that capability.
Bob