Locked Re: Skimming angle XRF- Experiment
.mca files for the galena encrustation. I should have labeled the picture of the sample better- that is a mm ruler, not inches! Geo
Locked Re: Skimming angle XRF- Experiment
Back to collimated beam and sensor. Examining encrustation on corner of a galena crystal.
By GEOelectronics@... · #2110 ·
Locked Re: Skimming angle XRF- Experiment
" How do you control the beam angle? What accuracy can you set it to and how is that measured. *Can you run a shot with and without the detector collimator I¡¯d like to see the difference.* Dud" Same
By GEOelectronics@... · #2108 ·
Re: Am241 beam collimator
Good idea Taray. Geo To: [email protected] Sent: Sunday, January 31, 2021 10:42:52 AM Subject: [XRF] Am241 beam collimator Hi guys Two common ways for Am241 housing is coaxial and converging plates Here
By GEOelectronics@... · #2107 ·
Am241 beam collimator
Hi guys Two common ways for Am241 housing is coaxial and converging plates Here is an another way I am toying round. Using radionuclide lead pigs I manage to squeeze 4 buttons and made a pair The
By taray singh <sukhjez@...> · #2106 ·
Re: This Groups.io memory problems
Don't panic Taray, your stuff is good to retain original,becuase it shows the steps for students. Only duplicates should be removed, or of possible reduced in size if that even applies. George To:
By GEOelectronics@... · #2105 ·
Re: This Groups.io memory problems
GeoI think we can minimize on?sending pics ?of Mca files?I think ?some of my??pic ?mca?files ?in the database can be deleted?Taray
By taray singh <sukhjez@...> · #2104 ·
This Groups.io memory problems
Yesterday I tried to upload some documents concerning the Tel-X-Ometer and its copper target tube generator, the whole post was rejected by the system with a warning about exceeding the free memory
By GEOelectronics@... · #2103 ·
Re: Supplements exposed :filling materials of pharmaceuticals
Am is not cleaner than a tube with all the Np interference. A tube allows you to target the K edges and functions to selectively target a Z range of elements. To do that with a isotopic source you
By Dude · #2099 ·
Locked Re: Skimming angle XRF- Experiment
Randall said: "Somehow, I had the impression that the exciting beam was broadside and the low, skimming angle was from the sample shim towards the sensor." To answer this point, yes that arrangement
By GEOelectronics@... · #2097 ·
Locked Re: Skimming angle XRF- Experiment
This is the story of the blue rock. I think this is from the Blanchard mine from a field trip there (many thanks to our hosts, the NM Rock and Mineral Club who have the patent on that mine). . It is
Re: Supplements exposed :filling materials of pharmaceuticals
Pic of AmX8 on Si-PIN with the tungsten collimator still attached and results. The target was a pure silver stamp, pressed against the front of the RapCap.
By GEOelectronics@... · #2093 ·
Re: Supplements exposed :filling materials of pharmaceuticals
Taray I'm experimenting with a tube generator for education, and that's the "beam". Otherwise I use the AmX8 extensively. It is still the best way from a number of standpoints, and I will continue
By GEOelectronics@... · #2092 ·
Locked Re: Skimming angle XRF- Experiment
Mo-Re metal alloy "stamp" .002" thick. Molybdenum-Rhenium Skimming, tried at 2 ( 22kV@20uA ) 22kV @ 20uA then again at 15uA. Same patterns for both, the higher current just gave higher peaks. Pictures
By GEOelectronics@... · #2091 ·
Re: Supplements exposed :filling materials of pharmaceuticals
Geo This thing is very sensitive??indeed The size,distance??and??Z of??any mount or base matters??besides the Am button arrangement? ?I have removed all supporting clips now ?I am using
By taray singh <sukhjez@...> · #2090 ·
Re: Supplements exposed :filling materials of pharmaceuticals
Thanks for that update Taray. Your system is very sensitive to Fe. Have you tried Iron supplement pills or iron-added cereal food? I tried calcium pills before but no luck, now with the new
By GEOelectronics@... · #2089 ·
Re: Supplements exposed :filling materials of pharmaceuticals
I repeated the generic clopidogrel tablets This time I crushed them to make it homogeneous and a longer count time Chose a linear scale to make it. proportional . This time only Fe is seen But this
Wavelength Dispersive X-Ray Fluorescence Spectrometer (WDXRF Spectrometer)
Up until now, the main subject being discussed on this group has been technically called ED-XRF. The ED is for energy dispersive. Specifically simultaneous ED, since the excitation produces a spectrum
By GEOelectronics@... · #2087 ·
Re: Supplements exposed :filling materials of pharmaceuticals
Dud I too have some difficulty differentiating generics from original . I rely??on big company names ,price and country of manufacture. India and China famous for cheap generics . ?The weight
By taray singh <sukhjez@...> · #2084 ·
Re: Supplements exposed :filling materials of pharmaceuticals
Very interesting on the aspirin Both have about the same Rh but the Ca is way different, amongst other things. I may have to try this although the levels may be below my delectability. I have some
By Dude · #2079 ·