¿ªÔÆÌåÓý

ctrl + shift + ? for shortcuts
© 2025 Groups.io

Re: Si-PIN with a view


 

That's a nice graphic display too Dud, simple, direct.?

"Why does the DppMCA software limit you to 411? Is it a gain setting problem?"

Correct, the last allowed click position for course gain is something like 3.8. The 411 comes from course gain= 4 and fine gain all the way down.

Sensitivity falls off fast above 100 keV with CdTe, but when the signal is abundant enough, the resolution is still really fantastic. I'm chasing peaks that return 1 pulse per 3 hours, another, 1 per 6 hours......

\\
Geo

----- Original Message -----
From: Dude <dfemer@...>
To: [email protected]
Sent: Sun, 08 Mar 2020 13:43:43 -0400 (EDT)
Subject: Re: [XRF] Si-PIN with a view

Geo,

Very good on the cal procedure and like every other laboratory
procedure spot checks and end of day are essential. The Ba 133 gives other useful
lines to double check the Ecal with. A Fe-55 and the Ba133 should be in everyone¡¯s
XRF tool kit for sure. Cals and cal checks are easy for us with X-ray sources to
light up something with but for the Am and natural source guys they have
limited ability due to weak source and ?count time. Trying to do an energy cal
with those makes for a very noisy cal signal. ?Taray¡¯s experiment with a non x-ray
tube is interesting and may help get light the way. (if you pardon the pun).

?

Why does the DppMCA software limit you to 411? Is it a gain
setting problem?

I have been using Spectragryph which was made for optical
spectroscopy but I also use it for LIBS, Gamma Spec, and XRF. It ?has spectral line
look up for optical, gamma, and XRF. ?It can control the optical spectrometers
and I can import XRF and Gamma spec data and do peak ID. It free for the basic
program.

?

?

As you say we definitely need some more export programs and FP
data reduction programs as well.? Charles I believe is a programmer and maybe
he can take a look at making up a program for formatting the DppMCA data file
to an Ecal¡¯d Channel vs Energy file that we can use for importing to other
programs. What was the Peekaboo program. Groups IO blocked me for bouncing and
I missed some messages.

?

Dud

?

From: [email protected]
[mailto:[email protected]] On Behalf Of GEOelectronics@...
Sent: Sunday, March 8, 2020 6:18 AM
To: [email protected]
Subject: Re: [XRF] Si-PIN with a view

?

Thanks
Dud,?

With
Si-PIN the evolved calibration procedure for general scans here has evolved
into what may become my standard.

The
"stage is set" for a range of 0-61 keV, 4096 channels.

5.9
keV the Mn Ka X-Ray from a Fe-55 calibration disc (Spectech) is entered in the
E calibration chart to anchor the low end, 59.5 from an Am source anchors the
top end. In the middle 17.75 is convenient, also from Am. A Ba-133 cal disc
from Spectech gives?31 keV (among others) for the 4th point.

On
long runs, an after-run calibration test is done and noted, just for reference.
Amptek's Si-Pin has proven to be remarkably calibration-stable all the way out
to 100 hours of runtime. When multiple short runs are done, a randomly
interspaced calibration verification test run is thrown in for the record, as
well as at end of session.?

?

?

The
setup for CdTe sensors has evolved to two different ranges, 1-220 and 1-411
keV, both are 4096 channels. The stage is set using Fe-55 @5.9 and Ba-133 for
31, 81, 356. Resolution, while not as good as the Si-PIN is orders of magnitude
better than the best possible NaI(Tl). Future plans are to experiment wit
expanding its top end and exploring beyond 411, but for now that's as high as I
can get the acquisition program to allow. I'm hoping for 511 (e annihilation
energy) or ultimately 662 (Ba-137m y-Ray)

?

The
DppMCA and ADMCA acquisition control programs by Amptek and pretty much just
that, acquisition-and-control-their display, analysis and ability to compare
are severely limited.

Vey
much more can be accomplished by amateur science exporting those files into a
better display and compare program. Not to mention some FP (fundamental
parameter) freeware programs

?

Bill
Kolb recently sent in a scan made on his SpecTech UCS-30 MCA using a 5"
NaI(Tl) in a graded shield, and it knocked my socks off. Will ask for?
permission to post that here.

?

I
must encourage Charles' efforts and watch his progress, to me it is amazing,
very few of us can do that, least of all myself, and it is important that it be
done. We don't know where it will lead, but consider us luck to explore with
these new tools at long last.

?

Ours
is a small group here for now, but we are setting a new paradigm which is being
widely read as this is an open group. Hopefully some of the others lucky enough
to obtain the Si-PIN and other advanced TE cooled sensors will join in and help
us add to the knowledge base.

?

Geo/K0FF

?

?

?

-----
Original Message -----

From: Dude <dfemer@...>

To: [email protected]

Sent: Sat, 07 Mar 2020 20:27:54 -0500 (EST)

Subject: Re: [XRF] Si-PIN with a view

?

Geo,

Bringing
in an old .mca file will use the old cal. It seems the Ecal

in the View / preferences tab only works for a new acquisition. If you need to

change an old cal you¡¯ll need to reenter the channel and energy table and save

it.

?I
like 4 pt cals. If you look at the new cal the? quadratic has a

much better fit compared to linear.

Dud

?

From:
[email protected]

[mailto:[email protected]] On Behalf Of Charles David Young
Sent: Saturday, March 7, 2020 9:48 AM
To: [email protected]
Subject: Re: [XRF] Si-PIN with a view

?

Here is the .mca that I am now pointed to in View menu /
Preferences
.? BTW, just pointing to a
.mca file here is kind

of buggy.? I eventually had to enter it manually.

?

The E cal is 4 peaks and it is now pretty close over the

entire range.? Not up to Dud's standards I am sure but much closer than

before.

?

Charles

?

On Sat, Mar 7, 2020 at 10:36 AM <GEOelectronics@...>

wrote:

"Instead it uses

the E cal of a specific .mca file that you have to point to in the?View

menu / Preferences.
"

?

GOOD TO KNOW THAT!

?

I've been struggling with that same issue...¡­.THANKS

Charles!

?

George?

?

?

----- Original Message -----


From: Charles David Young <charlesdavidyoung@...>


To: [email protected]


Sent: Sat, 07 Mar 2020 09:44:18 -0500 (EST)


Subject: Re: [XRF] Si-PIN with a view

?

All that being said I can see the value of having a precise

E cal saved in my .mca files so that others can view them in DppMCA with the

minimum of hassle. What had eluded me until now was how DppMCA saves the E

cal.? It is not done automatically so that it will start up with the most

recent E cal.? Instead it uses the E cal of a specific .mca file that you

have to point to in the?View menu / Preferences.

?

So I will do my best to calibrate in DppMCA using the data

(not the E cals!) in these .mca files and store that so my .mca files will have

a more precise an E cal going forward.

?

Charles

?

?

On Sat, Mar 7, 2020 at 6:05 AM Charles David Young <charlesdavidyoung@...>

wrote:

"
These

spectra are wrong as well. It appears you have a bad energy


cal that you were using.
"

?

Sorry, but my Theremino image of these spectra speaks for

itself.? The energy cal over the entire range from AgLa at 2.98 up to

CeKb2 at 40.23 is very precise.

?

As I have tried to explain, the E cal in DppMCA is

irrelevant for my work.? My version of Theremino imports the bin data from

the .mca files and ignores the E cal embedded in the files.

?

Dud, I appreciate your valiant efforts to explain your

professional process to an amateur rock hound like myself.? For my

purposes though I like


DppMCA for collecting data (really no choice with this Amptek device) and

Theremino for analyzing and presenting the results.? The crucial factor

for me is that I have complete control over the Theremino software and have

customized it extensively to support my hobby.? I find this to be fun and

satisfying and a good way to share information with my friends.

?

Charles

?

On Fri, Mar 6, 2020 at 9:52 PM Dude <dfemer@...> wrote:

Charles,

These

spectra are wrong as well. It appears you have a bad energy


cal that you were using.

?

This is

the Ag cal on 2/24/202

START_TIME -
02/24/2020

04:58:50

SERIAL_NUMBER - 0

<<CALIBRATION>>

LABEL - Channel

312 3.605

1275 14.96

±õ¡¯»å

surely like to know how you got that cal Sb and Y?

?

±á±ð°ù±ð¡¯²õ

the Ce cal Its using Ag from 3/2/2020

<<CALIBRATION>>

LABEL - Channel

293 3.151

2031 22.163

?

±á±ð°ù±ð¡¯²õ

the Nb cal its using Ag as well

<<CALIBRATION>>

LABEL - Channel

293 3.151

2031 22.163

?

Using

the Ag cal, Ka is at ch 1881 with a 22.1 keV which is


correct. Then looking at ?channel 2953 (the Ce peak) it shows 34.75 keV

which


is also correct

However

using the Nb file the Ka peak at channel 2953 is 32.25?


keV

The Ce

file Ka at channel 2953 also shows ?a 32.5 keV which is the


Ka of Ba . Ce should be at 34.7 keV ?

So

while the Ag cal file is correct the Ce and Nb files are energy


shifted implying a gain change or a different cal file was used. Looking at the


2 point data that¡¯s pretty clear.

?

Looking

at the Ag file at channel 2953 gives 34.75 keV

Ch 2953

for the Nb file is? 32.25 - Nb and Ce use the same cal. The


Ag file is different

Ch 2953

for Ce is 32.25? - Nb and Ce use the same cal. The Ag file


is different

The

energy offset is 32.25? ¨C 34.75 = -2.5 keV error

Your Ce

should have come in at 34.77 keV but the cal is -2.5 keV to


low due to using an old Ag erroneous cal which ID¡¯s it as Ba at 32.25

keV.? The


same effect occurs with the Ce and Nb La¡¯s producing an offset making Ce look

and


ID like Ba.

?

This is

a classic example of what you are doing with Theremino is


wrong. Look at the attached Ag Cu Ce ?.png file. Ag and Ce are lined up


correctly but nothing could be further from the truth.? Look at the raw
.mca

¨C files


nothing lines up correctly. The problem is you have taken ?erroneous data

without


even examining it, quickly moved it over to theremino and manually fudged it

with


¡°sliders¡± to fit a preconceived notion making it look ¡°pretty¡±. ?

The

fault here ?lies in your comment - ¡°
These
were


scanned on different days.? It takes hours to create each scan so there is


no point in trying to recalibrate each day.? If this Si-PIN has any drift


I have not been able to detect it
¡±.

I

understand the issue of a really weak source and the time


it takes, but skipping absolutely essential steps in spectroscopy is just not


done and this is a perfect example of why.? If you had looked at and

interpreted


the raw data spectra in DppMCA it might have been apparent ( it wasn¡¯t for me I


thought it was Ba and it fit Ba , even DppMCA thought it was Ba.) but looking


at the known Np peaks should have said something was wrong as they were shifted


too. Yeah, I didn¡¯t do that either.

?

The

essential steps in spectroscopy that always need to be


done are pretty straight forward.

1.?????
Determine
what energy range you want to look at and set


up the gain to do that. In your case your source is Am, so a good energy range


would include 59.5 keV. For the low end you can get an source that covers the


low end like Fe55 and use the Am Np peaks. Use a 3pt or better cal as anything


fits 2 pts and if one is off you¡¯ll never know it.

2.?????
For THAT
energy range and Gain run a hot source calibrate.


Check other peaks that you didn¡¯t use to make sure it¡¯s a good cal fit. Save


that cal and use it for that energy/gain range the next time. But ¨C see #3

3.?????
At the
beginning of the day or before the first


spectra, do a Cal Check. Use the hot cal source and verify that all channels


and their energy line up and are correct. This doesn¡¯t take any time as it¡¯s a

hot


cal source you are using. At the end of the days run do another Cal Check ¨C

QA/QC


you know. If it doesn¡¯t match throw out the data and start over. That¡¯s pretty


rare though as these things shouldn¡¯t drift.

4.?????
Using DppMCA
start ID¡¯ing the big peaks working down using


your eyeballs and an energy table looking for close interferences. Put a manual


ROI on it and verify Ka,b- La,b and when you have a positive ID leave the ROI


on it. A positive ID needs both a and b lines. A not determined has only one


line so don¡¯t bet on that horse. Do not use computer picks as that is only


there to mislead and embarrass you.

?

If you

must use the Theremino pretty picture thing then Cal it


correctly. The very fact that it doesn¡¯t have a linear or quadratic regression


in it is a real clue it¡¯s not ready for prime time.? Being that you¡¯re an


accomplished software geek with this program use the linear or quadratic


equation from DppMCA under the Calibrate tab

(Energy

=A+Bx ?or for a quadratic E= A+B*x + C*x^2 where x is


channel number) to calculate the channel vs energy table then import that


directly into the Thermino and avoid the problematic slider thing.? Or

?program


it in to Theremino and have it calc it.

The

slider thing is probably needed for the sound cards non linear


response or missing codes. It¡¯s, at best, an approximation that works for NaI


but shouldn¡¯t be used ?for high res detectors.

?

Dud

?

?

?

From:[email protected]


[mailto:[email protected]] On

Behalf Of
Charles David Young
Sent: Thursday, March 5, 2020 10:28 AM
To:[email protected]
Subject: Re: [XRF] Si-PIN with a view

?

Dud,

here are the .mca files along with another one that I


did of Nb last night.? I would be interested in seeing what calibration


points you can come up with.? These were scanned on different days.?


It takes hours to create each scan so there is no point in trying to


recalibrate each day.? If this Si-PIN has any drift I have not been able


to detect it.

?

Charles

?

On

Thu, Mar 5, 2020 at 6:56 AM <GEOelectronics@...>


wrote:

"Dud, I made


a special decluttered plot just for you.? And in the process I discovered


that the SmKa1 peak that I had previously identified is actually CeKb2 so this


Ce metal is actually a little purer than I thought!

?

Charles"

?

Be sure to pay attention to


L-gamma and M- shell lines too.?

?

Geo

?

?

?

-----

Original Message -----



From: Charles David Young <charlesdavidyoung@...>



To: [email protected]



Sent: Wed, 04 Mar 2020 23:07:58 -0500 (EST)



Subject: Re: [XRF] Si-PIN with a view

?

Ok,

so the deal is that using AgKa1 as one of the points on


the E cal line is not very accurate by the time it gets up to CeKa1 so the peak


finder in DppMCA

?

assumed

that peak was BaKa1.? Use these cal points


instead in DppMCA:

?

2955

34.72

270

2.98

?

You

may ask why I never noticed this discrepancy?? The reason


is that I use DppMCA strictly for collecting data.? The E cal in DppMCA is


not critical.? All analysis is done in my highly customized Theremino


where I can label things nicely.? The table format of



DppMCA just doesn't cut it for me.

?

Dud,

I made a special decluttered plot just for you.?


And in the process I discovered that the SmKa1 peak that I had previously


identified is actually CeKb2 so this Ce metal is actually a little purer than I


thought!

?

Charles

?

On

Wed, Mar 4, 2020 at 2:44 PM Dude <dfemer@...> wrote:

Charles,

Check


your ¡°²õ±ô¾±»å±ð°ù¡± calibration. This is Ba not Ce. Go back to the



.mca file and look at it with the original cal using the DppMCA program. It



looks like you used Ag for a cal but when was that cal done? Before you ran the



¡°Ce¡± or was it an old one?

Please


clean up the theremino plot, you have so many lines and



labels on their it¡¯s hard to tell what is what. Get rid of anything that¡¯s not



ID¡¯d or an interference anything else is clutter.? One shouldn¡¯t ?do


an interp



based on pre-established labels. Treat each peak separately starting at the low



energy and work your way up ruling out close interferences. Then put an ROI and



?ID label on it and use a real MCA program.? Thermino is great for


NaI but it¡¯ll



get you in trouble with these high res detectors.

There


is no Sm , La or Ce in here and it looks like you were at



83% dead time which is too high which can shift energy and FWHM.

Think


scientifically. The hypothesis was its Ce. You need to prove



that.? I¡¯ll guess you used your pre-established Ce labels and did the


¡°²õ±ô¾±»å±ð°ù¡±



thing to line it all up to a preconceived notion. ??Wrong approach


and guaranteed



to get you in trouble- Don¡¯t use a secondary Cal - use the data acquisition



program to ID peaks.? Look at the periodic table ¨C Ba, La, and Ce line up


so



they are going to have the same spectral structure close energies and look



alike if the cal isn¡¯t right.? An interp starts by looking at each


individual peak



energy and any surrounding interferences with no pre-conceived notions, then


decide



based on Ka,b and La,b and move on to the next. If it doesn¡¯t take you an hour



or more you¡¯re not doing it right. Computer picks suck, stay a long way away ¨C



use your eyes, brains and an energy table.

Dud

?

From:[email protected]



[mailto:[email protected]] On


Behalf Of
Charles David Young
Sent: Wednesday, March 4, 2020 4:25 AM
To:[email protected]
Subject: Re: [XRF] Si-PIN with a view

?

Steve,

?

This


is from Gunnar Faerber who did his own analysis.?



He is also the one that identified the Betafo speciman as brannerite.

?

?

Meanwhile,


this is an interesting scan of Ce metal.? It



is evidently hard to purify it of the other lanthanides because it shows plenty



of Sm and even detectable La.? The La1 and Lb1 peaks echo the Ka1 and Kb1



peaks nicely.? Taken together this set of peaks is quite useful for



calibration from low to high.

?

Charles

?

On


Tue, Mar 3, 2020 at 9:03 PM WILLIAM S Dubyk <sdubyk@...> wrote:

Charles,


who did you get your specimen of ishikawaite from,



and do you have a microprobe analysis for it? The reason I ask is that out of



curiosity I looked at some of the samarskite analyses from the Petaca district,



and several of the Queen mine specimens may be that mineral. I will ask Al



Falster, who is somewhat of an expert on the samarskite minerals, about what



analyses to use to determine this - oxides versus cation ratios. From my



understanding, this mineral has U+Th greater than Y+REE, and that is what I see



in the Queen mine specimens.





Steve

?

?

?

?

?

?

?

?

?

?

?

?

?

?

?

?

?

?

?






Join [email protected] to automatically receive all group messages.