¿ªÔÆÌåÓý

ctrl + shift + ? for shortcuts
© 2025 Groups.io

Re: Si-PIN with a view


 

Thanks Dud,?
With Si-PIN the evolved calibration procedure for general scans here has evolved into what may become my standard.
The "stage is set" for a range of 0-61 keV, 4096 channels.
5.9 keV the Mn Ka X-Ray from a Fe-55 calibration disc (Spectech) is entered in the E calibration chart to anchor the low end, 59.5 from an Am source anchors the top end. In the middle 17.75 is convenient, also from Am. A Ba-133 cal disc from Spectech gives?31 keV (among others) for the 4th point.
On long runs, an after-run calibration test is done and noted, just for reference. Amptek's Si-Pin has proven to be remarkably calibration-stable all the way out to 100 hours of runtime. When multiple short runs are done, a randomly interspaced calibration verification test run is thrown in for the record, as well as at end of session.?


The setup for CdTe sensors has evolved to two different ranges, 1-220 and 1-411 keV, both are 4096 channels. The stage is set using Fe-55 @5.9 and Ba-133 for 31, 81, 356. Resolution, while not as good as the Si-PIN is orders of magnitude better than the best possible NaI(Tl). Future plans are to experiment wit expanding its top end and exploring beyond 411, but for now that's as high as I can get the acquisition program to allow. I'm hoping for 511 (e annihilation energy) or ultimately 662 (Ba-137m y-Ray)

The DppMCA and ADMCA acquisition control programs by Amptek and pretty much just that, acquisition-and-control-their display, analysis and ability to compare are severely limited.
Vey much more can be accomplished by amateur science exporting those files into a better display and compare program. Not to mention some FP (fundamental parameter) freeware programs

Bill Kolb recently sent in a scan made on his SpecTech UCS-30 MCA using a 5" NaI(Tl) in a graded shield, and it knocked my socks off. Will ask for? permission to post that here.

I must encourage Charles' efforts and watch his progress, to me it is amazing, very few of us can do that, least of all myself, and it is important that it be done. We don't know where it will lead, but consider us luck to explore with these new tools at long last.

Ours is a small group here for now, but we are setting a new paradigm which is being widely read as this is an open group. Hopefully some of the others lucky enough to obtain the Si-PIN and other advanced TE cooled sensors will join in and help us add to the knowledge base.

Geo/K0FF



----- Original Message -----
From: Dude <dfemer@...>
To: [email protected]
Sent: Sat, 07 Mar 2020 20:27:54 -0500 (EST)
Subject: Re: [XRF] Si-PIN with a view

Geo,

Bringing in an old .mca file will use the old cal. It seems the Ecal
in the View / preferences tab only works for a new acquisition. If you need to
change an old cal you¡¯ll need to reenter the channel and energy table and save
it.

?I like 4 pt cals. If you look at the new cal the? quadratic has a
much better fit compared to linear.

Dud

?

From: [email protected]
[mailto:[email protected]] On Behalf Of Charles David Young
Sent: Saturday, March 7, 2020 9:48 AM
To: [email protected]
Subject: Re: [XRF] Si-PIN with a view

?

Here is the .mca that I am now pointed to in View menu /
Preferences
.? BTW, just pointing to a .mca file here is kind
of buggy.? I eventually had to enter it manually.

?

The E cal is 4 peaks and it is now pretty close over the
entire range.? Not up to Dud's standards I am sure but much closer than
before.

?

Charles

?

On Sat, Mar 7, 2020 at 10:36 AM <GEOelectronics@...>
wrote:

"Instead it uses
the E cal of a specific .mca file that you have to point to in the?View
menu / Preferences.
"

?

GOOD TO KNOW THAT!

?

I've been struggling with that same issue...¡­.THANKS
Charles!

?

George?

?

?

----- Original Message -----

From: Charles David Young <charlesdavidyoung@...>

To: [email protected]

Sent: Sat, 07 Mar 2020 09:44:18 -0500 (EST)

Subject: Re: [XRF] Si-PIN with a view

?

All that being said I can see the value of having a precise
E cal saved in my .mca files so that others can view them in DppMCA with the
minimum of hassle. What had eluded me until now was how DppMCA saves the E
cal.? It is not done automatically so that it will start up with the most
recent E cal.? Instead it uses the E cal of a specific .mca file that you
have to point to in the?View menu / Preferences.

?

So I will do my best to calibrate in DppMCA using the data
(not the E cals!) in these .mca files and store that so my .mca files will have
a more precise an E cal going forward.

?

Charles

?

?

On Sat, Mar 7, 2020 at 6:05 AM Charles David Young <charlesdavidyoung@...>
wrote:

"
These
spectra are wrong as well. It appears you have a bad energy

cal that you were using.
"

?

Sorry, but my Theremino image of these spectra speaks for
itself.? The energy cal over the entire range from AgLa at 2.98 up to
CeKb2 at 40.23 is very precise.

?

As I have tried to explain, the E cal in DppMCA is
irrelevant for my work.? My version of Theremino imports the bin data from
the .mca files and ignores the E cal embedded in the files.

?

Dud, I appreciate your valiant efforts to explain your
professional process to an amateur rock hound like myself.? For my
purposes though I like

DppMCA for collecting data (really no choice with this Amptek device) and
Theremino for analyzing and presenting the results.? The crucial factor
for me is that I have complete control over the Theremino software and have
customized it extensively to support my hobby.? I find this to be fun and
satisfying and a good way to share information with my friends.

?

Charles

?

On Fri, Mar 6, 2020 at 9:52 PM Dude <dfemer@...> wrote:

Charles,

These
spectra are wrong as well. It appears you have a bad energy

cal that you were using.

?

This is
the Ag cal on 2/24/202

START_TIME - 02/24/2020
04:58:50

SERIAL_NUMBER - 0

<<CALIBRATION>>

LABEL - Channel

312 3.605

1275 14.96

±õ¡¯»å
surely like to know how you got that cal Sb and Y?

?

±á±ð°ù±ð¡¯²õ
the Ce cal Its using Ag from 3/2/2020

<<CALIBRATION>>

LABEL - Channel

293 3.151

2031 22.163

?

±á±ð°ù±ð¡¯²õ
the Nb cal its using Ag as well

<<CALIBRATION>>

LABEL - Channel

293 3.151

2031 22.163

?

Using
the Ag cal, Ka is at ch 1881 with a 22.1 keV which is

correct. Then looking at ?channel 2953 (the Ce peak) it shows 34.75 keV
which

is also correct

However
using the Nb file the Ka peak at channel 2953 is 32.25?

keV

The Ce
file Ka at channel 2953 also shows ?a 32.5 keV which is the

Ka of Ba . Ce should be at 34.7 keV ?

So
while the Ag cal file is correct the Ce and Nb files are energy

shifted implying a gain change or a different cal file was used. Looking at the

2 point data that¡¯s pretty clear.

?

Looking
at the Ag file at channel 2953 gives 34.75 keV

Ch 2953
for the Nb file is? 32.25 - Nb and Ce use the same cal. The

Ag file is different

Ch 2953
for Ce is 32.25? - Nb and Ce use the same cal. The Ag file

is different

The
energy offset is 32.25? ¨C 34.75 = -2.5 keV error

Your Ce
should have come in at 34.77 keV but the cal is -2.5 keV to

low due to using an old Ag erroneous cal which ID¡¯s it as Ba at 32.25
keV.? The

same effect occurs with the Ce and Nb La¡¯s producing an offset making Ce look
and

ID like Ba.

?

This is
a classic example of what you are doing with Theremino is

wrong. Look at the attached Ag Cu Ce ?.png file. Ag and Ce are lined up

correctly but nothing could be further from the truth.? Look at the raw .mca
¨C files

nothing lines up correctly. The problem is you have taken ?erroneous data
without

even examining it, quickly moved it over to theremino and manually fudged it
with

¡°sliders¡± to fit a preconceived notion making it look ¡°pretty¡±. ?

The
fault here ?lies in your comment - ¡°
These were

scanned on different days.? It takes hours to create each scan so there is

no point in trying to recalibrate each day.? If this Si-PIN has any drift

I have not been able to detect it
¡±.

I
understand the issue of a really weak source and the time

it takes, but skipping absolutely essential steps in spectroscopy is just not

done and this is a perfect example of why.? If you had looked at and
interpreted

the raw data spectra in DppMCA it might have been apparent ( it wasn¡¯t for me I

thought it was Ba and it fit Ba , even DppMCA thought it was Ba.) but looking

at the known Np peaks should have said something was wrong as they were shifted

too. Yeah, I didn¡¯t do that either.

?

The
essential steps in spectroscopy that always need to be

done are pretty straight forward.

1.?????
Determine what energy range you want to look at and set

up the gain to do that. In your case your source is Am, so a good energy range

would include 59.5 keV. For the low end you can get an source that covers the

low end like Fe55 and use the Am Np peaks. Use a 3pt or better cal as anything

fits 2 pts and if one is off you¡¯ll never know it.

2.?????
For THAT energy range and Gain run a hot source calibrate.

Check other peaks that you didn¡¯t use to make sure it¡¯s a good cal fit. Save

that cal and use it for that energy/gain range the next time. But ¨C see #3

3.?????
At the beginning of the day or before the first

spectra, do a Cal Check. Use the hot cal source and verify that all channels

and their energy line up and are correct. This doesn¡¯t take any time as it¡¯s a
hot

cal source you are using. At the end of the days run do another Cal Check ¨C
QA/QC

you know. If it doesn¡¯t match throw out the data and start over. That¡¯s pretty

rare though as these things shouldn¡¯t drift.

4.?????
Using DppMCA start ID¡¯ing the big peaks working down using

your eyeballs and an energy table looking for close interferences. Put a manual

ROI on it and verify Ka,b- La,b and when you have a positive ID leave the ROI

on it. A positive ID needs both a and b lines. A not determined has only one

line so don¡¯t bet on that horse. Do not use computer picks as that is only

there to mislead and embarrass you.

?

If you
must use the Theremino pretty picture thing then Cal it

correctly. The very fact that it doesn¡¯t have a linear or quadratic regression

in it is a real clue it¡¯s not ready for prime time.? Being that you¡¯re an

accomplished software geek with this program use the linear or quadratic

equation from DppMCA under the Calibrate tab

(Energy
=A+Bx ?or for a quadratic E= A+B*x + C*x^2 where x is

channel number) to calculate the channel vs energy table then import that

directly into the Thermino and avoid the problematic slider thing.? Or
?program

it in to Theremino and have it calc it.

The
slider thing is probably needed for the sound cards non linear

response or missing codes. It¡¯s, at best, an approximation that works for NaI

but shouldn¡¯t be used ?for high res detectors.

?

Dud

?

?

?

From:[email protected]

[mailto:[email protected]] On
Behalf Of
Charles David Young
Sent: Thursday, March 5, 2020 10:28 AM
To:[email protected]
Subject: Re: [XRF] Si-PIN with a view

?

Dud,
here are the .mca files along with another one that I

did of Nb last night.? I would be interested in seeing what calibration

points you can come up with.? These were scanned on different days.?

It takes hours to create each scan so there is no point in trying to

recalibrate each day.? If this Si-PIN has any drift I have not been able

to detect it.

?

Charles

?

On
Thu, Mar 5, 2020 at 6:56 AM <GEOelectronics@...>

wrote:

"Dud, I made

a special decluttered plot just for you.? And in the process I discovered

that the SmKa1 peak that I had previously identified is actually CeKb2 so this

Ce metal is actually a little purer than I thought!

?

Charles"

?

Be sure to pay attention to

L-gamma and M- shell lines too.?

?

Geo

?

?

?

-----
Original Message -----


From: Charles David Young <charlesdavidyoung@...>


To: [email protected]


Sent: Wed, 04 Mar 2020 23:07:58 -0500 (EST)


Subject: Re: [XRF] Si-PIN with a view

?

Ok,
so the deal is that using AgKa1 as one of the points on

the E cal line is not very accurate by the time it gets up to CeKa1 so the peak

finder in DppMCA

?

assumed
that peak was BaKa1.? Use these cal points

instead in DppMCA:

?

2955
34.72

270
2.98

?

You
may ask why I never noticed this discrepancy?? The reason

is that I use DppMCA strictly for collecting data.? The E cal in DppMCA is

not critical.? All analysis is done in my highly customized Theremino

where I can label things nicely.? The table format of


DppMCA just doesn't cut it for me.

?

Dud,
I made a special decluttered plot just for you.?

And in the process I discovered that the SmKa1 peak that I had previously

identified is actually CeKb2 so this Ce metal is actually a little purer than I

thought!

?

Charles

?

On
Wed, Mar 4, 2020 at 2:44 PM Dude <dfemer@...> wrote:

Charles,

Check

your ¡°²õ±ô¾±»å±ð°ù¡± calibration. This is Ba not Ce. Go back to the


.mca file and look at it with the original cal using the DppMCA program. It


looks like you used Ag for a cal but when was that cal done? Before you ran the


¡°Ce¡± or was it an old one?

Please

clean up the theremino plot, you have so many lines and


labels on their it¡¯s hard to tell what is what. Get rid of anything that¡¯s not


ID¡¯d or an interference anything else is clutter.? One shouldn¡¯t ?do

an interp


based on pre-established labels. Treat each peak separately starting at the low


energy and work your way up ruling out close interferences. Then put an ROI and


?ID label on it and use a real MCA program.? Thermino is great for

NaI but it¡¯ll


get you in trouble with these high res detectors.

There

is no Sm , La or Ce in here and it looks like you were at


83% dead time which is too high which can shift energy and FWHM.

Think

scientifically. The hypothesis was its Ce. You need to prove


that.? I¡¯ll guess you used your pre-established Ce labels and did the

¡°²õ±ô¾±»å±ð°ù¡±


thing to line it all up to a preconceived notion. ??Wrong approach

and guaranteed


to get you in trouble- Don¡¯t use a secondary Cal - use the data acquisition


program to ID peaks.? Look at the periodic table ¨C Ba, La, and Ce line up

so


they are going to have the same spectral structure close energies and look


alike if the cal isn¡¯t right.? An interp starts by looking at each

individual peak


energy and any surrounding interferences with no pre-conceived notions, then

decide


based on Ka,b and La,b and move on to the next. If it doesn¡¯t take you an hour


or more you¡¯re not doing it right. Computer picks suck, stay a long way away ¨C


use your eyes, brains and an energy table.

Dud

?

From:[email protected]


[mailto:[email protected]] On

Behalf Of
Charles David Young
Sent: Wednesday, March 4, 2020 4:25 AM
To:[email protected]
Subject: Re: [XRF] Si-PIN with a view

?

Steve,

?

This

is from Gunnar Faerber who did his own analysis.?


He is also the one that identified the Betafo speciman as brannerite.

?

?

Meanwhile,

this is an interesting scan of Ce metal.? It


is evidently hard to purify it of the other lanthanides because it shows plenty


of Sm and even detectable La.? The La1 and Lb1 peaks echo the Ka1 and Kb1


peaks nicely.? Taken together this set of peaks is quite useful for


calibration from low to high.

?

Charles

?

On

Tue, Mar 3, 2020 at 9:03 PM WILLIAM S Dubyk <sdubyk@...> wrote:

Charles,

who did you get your specimen of ishikawaite from,


and do you have a microprobe analysis for it? The reason I ask is that out of


curiosity I looked at some of the samarskite analyses from the Petaca district,


and several of the Queen mine specimens may be that mineral. I will ask Al


Falster, who is somewhat of an expert on the samarskite minerals, about what


analyses to use to determine this - oxides versus cation ratios. From my


understanding, this mineral has U+Th greater than Y+REE, and that is what I see


in the Queen mine specimens.




Steve

?

?

?

?

?

?

?

?

?

?

?

?

?

?






Join [email protected] to automatically receive all group messages.