Re: Zircon- Malawi vs Sri Lanka


 

To all, NEVER TRY TO MEASURE HV going to a X-Ray tube with a meter...¡­...DANGEROUS to do so.

Instead look at the beam with a sensor, with a collimator between (cover the sensor with a lead Pb sheet with a 1mm hole to let the rays through)
look at the bump on your MCA (Bremsstrahlung Rays), where it ends the keV = actual HV kVp.

Geo

----- Original Message -----
From: taray singh via Groups.Io <sukhjez@...>
To: [email protected]
Sent: Tue, 03 Mar 2020 04:24:24 -0500 (EST)
Subject: Re: [XRF] Zircon- Malawi vs Sri Lanka

Geo

My multimeter max is 1000ohm

Theoretically for??HV measurements ,If I place a 1000 ohm resistor in series with my sensor,I should be able to ..

Potential danger of arcing and inexperience is stopping me

Taray

?