¿ªÔÆÌåÓý

ctrl + shift + ? for shortcuts
© 2025 Groups.io

Re: Supplements exposed


taray singh
 

Geo
This is from Amptek troubleshooting guide?

If the resolution is worse than you expect, then ...
Check for good cooling
In Amptek¡¯s thermoelectrically cooled detectors, the back of the detector is the heat sink. In Amptek¡¯s XR100 and X123 products, this is connected to the case. The detector is cooled relative to the back of the detector. If the back of the detector gets warm, so will the detector, and resolution will be degraded. It is very important to provide a good heat sink from the preamp case or detector back to ambient. Many customers have found that much more heat sinking was required than they initially expected.

Taray





On Sunday, October 25, 2020, 10:54 PM, GEOelectronics@... wrote:

" Tip for calibration
It can be difficult to find 2 peaks at extreme end within a specimen or combination.
Sometimes with? this method it is difficult to identify peaks
Easier way is use 2 different samples
After entering data into the calibration box,delete data and reset?
Enter next sample and repeat"

Good idea, and the calibration points can be radioactive source, or XRF made by exciter.

This last way you can place the cal peak anywhere you wat, up to the limit of your exciter, simply by using pure elements. One very low and one at 59.5 is really all that is needed due to Si-PIN linearity.

Try this:?After an important long run, save the .mca and a BAK.mca of it, change DPPMCA time to longer and add 59.5 source for how long it takes to get a good solid peak at 95.5 and 13.95 or 17.750keV, all the while leaving the original scan on the screen. Save and label this as?"xxxxx"-Post-Run-Am-Cal.mca and use it to tweak the calibration of earlier saved .mca, or as proof of linearity later.?

That's why to include 0-62? keV even though you might have no peaks above ~32 etc. Just set RANGE boxes to highlight the area where the peaks are, maybe 2 to 35 keV.

Also must add, any "drift" in calibration is usually not drift at all, but the maturing of the peak from start to finish to a point where the actual peak is at more clear. In most cases after days of running it may only be a +/- 1 channel change.


Is there an issue/question? with cooling/heatsink?? What you feel on case should be warm like a puppy. This is the 1 Watt or so of heat being removed from the sensor chip inside the tip behind the Be window, which operates in a vacuum at? ~230K = minus 43C.

My SDD sensor has a more sophisticated cooler, it uses a liquid? filled "heat-pipe" to bring heat inside the main box from the tip, a heatsink and fan inside the box to cool it down and blow it away from the sensor end. In a way heat pipe is same idea as a cold finger on HPGe sensors. Heat pipe also = Drinking Bird toy tech.

Geo




From: "taray singh via groups.io" <sukhjez@...>
To: [email protected]
Sent: Sunday, October 25, 2020 2:53:13 AM
Subject: Re: [XRF] Supplements exposed

Here is pic of selenium sulphide(dried centrifugal sample of selenium sulphide suspension / Selsun) using 2048 with threshold correction.
The selenium peaks are seen clearly? and not hidden like the one with the supplements.

Tip for calibration
It can be difficult to find 2 peaks at extreme end within a specimen or combination.
Sometimes with? this method it is difficult to identify peaks
Easier way is use 2 different samples
After entering data into the calibration box,delete data and reset?
Enter next sample and repeat
Can delete previous ROI? after entering data into calibration box
Eample Fe and Am241 59.5?
?
Also a pic from Engineer's handbook about heat sink
Maybe a little small to see on a pc
Taray

Join [email protected] to automatically receive all group messages.