¿ªÔÆÌåÓýNice clean spectra, lookin¡¯ good.? You can take the conversion gain up to 2048 channels which will get a bit more resolution at the expense of a longer count time. If you are talking about a peak please put the cursor on it so we can read the energy and counts. This program really needs to have the x-axis numbers to be below the spectra as it¡¯s really hard to see the energy which is covered up by the spectra baseline noise. Also post the.mca files along with the photos so some of us can play with the data. I¡¯m surprised we don¡¯t see a large Am peak.? How do you do the energy cal? Dud ? From: [email protected]
[mailto:[email protected]] On Behalf Of taray singh via groups.io
Sent: Monday, September 28, 2020 4:00 AM To: [email protected] Subject: [XRF] Si pin xrf ..1st go ? Hi guys Using 6 Am241 /ludlum??lead disc??glued to cardboard box Si pin??near orifice Due to the shape of detector and fear of samples potentially damaging Be window,I opted for this way Anyway it is only a prototype. Accidentally glued one the Am buttons upside down and it is really stuck Will fix it next time.. Samples as follows? 1 Selenium Was discussed in the old forum My sample is a centrifuged suspension of selenium sulphide(selsen) in a??plastic sample container .It has remained the same since preparation about 2 years ago It sticks like a coating so without a container,alpha is active. ? 2 X ray apron polymer Nowadays they are lead free and consist of polymers mixed??with antimony,tungsten or bismuth etc to make it light weight? I always thought there was antimony present in my sample from my previous xrf experiments.It appears to Tin here?? 3 Bipp a bismuth iodine medical paste ? 4 Mn 99.7% purity in a plastic container? ? Thanks Taray |