¿ªÔÆÌåÓý

ctrl + shift + ? for shortcuts
© 2025 Groups.io

Re: The hardware v0.1


 

VERY IMPRESSIVE!!!!!!!!!? ?I wish I had seen that just a bit ago!!!? You have made fantastic progress!!!!!

I don't understand exactly how your pressure monitor is working.? ?Can you tell me
a)? the smallest pressure change you can measure (RESOLUTION)
b)? the maximum pressure you can measure
with that setup????

We also need a DISCONNECT ALARM.? ?If you add that, you already?meet all the safety requirements.? Disconnect is determined when the inspired pressure fails to significantly rise, despite the fact that you are delivering a breath to the patient.? It means the patient has become disconnected from the tubing.? So instead of seeing 30 cm of H2O pressure....you suddenly?see only 2 or 3 when you deliver a breath.... make sense?

GREAT JOB!? ?You can call me ANYTIME.? ? ?I don't know exactly HOW you call me but I will try.? ?FANTASTIC WORK.
gordon



On Thu, Mar 26, 2020 at 4:42 PM Ashhar Farhan <farhanbox@...> wrote:
Gents,
Please review the candidate hardware. If you guys are fine with the functionality, I will start laying this out.
Gordon, what is a good time to call you?



- f

Join [email protected] to automatically receive all group messages.