Re: Making a Q-meter /
On Thu, 4 Aug 2022 at 14:00, Navmar <navmar@...> wrote: You raise the issue here of how to generate a signal with very low source resistance. Could you please sketch it out, as I am not following it. Dave |
Re: Making a Q-meter /
You raise the issue here of how to generate a signal with very low source resistance.
How about this idea. Consider a signal source feeding a resistive divider consisting of Rs (source) and Rl (load).? This gives the relationship? Rs/Rl = Vs/Vl-1. So Rs approaches 0 as Vs and Vl become equal. If you introduce an ALC control loop that monitors Vl and attempts to make Vs and Vl equal then the effective Rs becomes small. A worthy idea or not ? I played with this a bit recently and could quite easily get Rs down to one ohm at 50MHz.? Its part of an idea to modernize the old Boonton 250 RX meter.? Bill |
Re: Making a Q-meter /
On Thu, 4 Aug 2022 at 14:34, Jeffrey Pawlan <pawlan@...> wrote: I have a real Boonton Q meter.? Yes it was made with tubes (valves).? But it was a truly outstanding meter in its day.? David, I would consider giving it to you for trade for something else small but getting it to you is a a problem. I think it would be cheaper for me to fly there with it in a suitcase than to ship it. Well Jeffrey, you bring up an interesting point. I bought a 160-A from the USA a few days ago. The shipping was a suspiciously low $21.72.? It's going via the eBay Global Shipping Program which is run by Pitney Bowes. The tracking shows it was picked up in California, and is still in
California, but is due for delivery tomorrow by FedEx in Kentucky, which I assume is the Global Shipping Centre. Now what happens there I do not know. If they realise they are only being paid $21.72 to ship it to the UK, they might refuse to ship it. The other possibility, and one I fear, is they will ship it with inadequate carriage, and I will be expected to pay for the excess carriage charge. In which case I will decline. If I'm going to have to pay $150-$200 in shipping charges, it is not worth my while. The meter cost me $75. |
Re: Making a Q-meter /
On Thu, 4 Aug 2022 at 13:22, jfphp via <jfphp=[email protected]> wrote:
Do you know of a model number(s)? My search with Google didn¡¯t bring up anything useful. There are plenty of LCR meters, but unless significantly different from my 4284A and 4285A, they are pretty poor at measuring Q above about 20.? Dave?
Dr. David Kirkby, Kirkby Microwave Ltd, drkirkby@... Telephone 01621-680100./ +44 1621 680100 Registered in England & Wales, company number 08914892. Registered office: Stokes Hall Lodge, Burnham Rd, Althorne, Chelmsford, Essex, CM3 6DT, United Kingdom |
Re: Making a Q-meter /
On Thu, 4 Aug 2022 at 11:53, Chuck Moore via <wd4hxg=[email protected]> wrote: If you have not already done so, you may want to read through the Boonton 260 Q Meter manual. Thank you. That is the best 260-A manual I have seen, although I have seen a ¡°manual¡± that covered a few of the Boonton models. I always felt that there should be one specific to each model. Reading that, I do get the feeling that some of the improvements made between the 160-A and 260-A, would not be necessary if an instrument was made with a microprocessor doing some number?crunching. For example, they reduced the injection resistor from 40 milli ohm to 20 milli ohm. But still it¡¯s a source of error at high Q which one can correct with the aid of a calculation. If an instrument was made nowadays, one might just as well correct every reading ?Maybe the resistors used in the PW design would be okay if a microprocessor did the corrections, rather than?
I don¡¯t understand why those thermocouples are so easily damaged.?
If I bought one I would send it to Keysight - assuming that they will still calibrate it, which I suspect that they will. I have a healthy scepticism of calibration labs.? Dave Dr. David Kirkby, Kirkby Microwave Ltd, drkirkby@... Telephone 01621-680100./ +44 1621 680100 Registered in England & Wales, company number 08914892. Registered office: Stokes Hall Lodge, Burnham Rd, Althorne, Chelmsford, Essex, CM3 6DT, United Kingdom |
Re: Making a Q-meter /
I have a real Boonton Q meter.? Yes it was made with tubes (valves).? But it was a truly outstanding meter in its day.? David, I would consider giving it to you for trade for something else small but getting it to you is a a problem. I think it would be cheaper for me to fly there with it in a suitcase than to ship it.
73, Jeffrey Pawlan? WA6KBL |
Re: Making a Q-meter /
On? Thu, 4 Aug 2022 9:19? Dr. David Kirkby wrote:
"A step-down transformer seems the most obvious way...? I have not yet looked at the HP 4343A meter service manual. That will probably give me some ideas. "
------------------------------------------------------------
The
HP4342A Service manual is vague about the
injection transformer, but it is driven by a cascaded emitter follower.?
There is a 75 Ohm resistor across the primary winding to stabilise the
driver load impedance outside DUT resonance.HP
Journal September 1970
(at https://www.hpl.hp.com/hpjournal/pdfs/IssuePDFs/1970-09.pdf) is more
forthcoming and has some other interesting details about the HP4342A design.?
I do wonder what ferrite was used and how
they wound the toroidal transformer to avoid self resonance and obtain "a nearly flat frequency response" over the design bandwidth 22kHz to 70MHz.? The turns ratio was 50:1.
PeterS?? ???? G8EZE
|
Re: Making a Q-meter /
On Thu, 4 Aug 2022 at 13:22, jfphp via <jfphp=[email protected]> wrote:
Dr. David Kirkby, Kirkby Microwave Ltd, drkirkby@... Telephone 01621-680100./ +44 1621 680100 Registered in England & Wales, company number 08914892. Registered office: Stokes Hall Lodge, Burnham Rd, Althorne, Chelmsford, Essex, CM3 6DT, United Kingdom |
Re: Making a Q-meter /
Have a look at the manual of the Rhode & Schwartz? QMeter
On Thursday, August 4, 2022 at 11:55:19 AM GMT+2, Dr. David Kirkby, Kirkby Microwave Ltd <drkirkby@...> wrote:
On Thu, Aug 4, 2022 at 02:14 AM, Kenneth Greenough wrote: Re' a Q meter. I have seen something like that before. One problem is that it would not work with coils in screened enclosures, some of which are.?I have ordered a couple of Boonton Q-standards, which I was hoping to measure. Those are in screen enclosures, with 2 banana plugs 25.4 mm (1¡±) apart.?See attached photograph.?
|
Re: Making a Q-meter /
If you have not already done so, you may want to read through the Boonton 260 Q Meter manual.
It is a wealth of information and provides a lot of insight on where errors creep into measurements. The manual can be found at the following link:? Boonton's approach was to use a 20 milliOhm precision resistor with an attached thermocouple that drove a panel meter. By measuring the power dissipation of the resistor, they obtained a known voltage drop across the resistor. The integral VTVM measured the voltage across the tuned circuit and the display meter was scaled in Q based on the results of the ratio of the tuned circuit voltage to the voltage across the precision resistor. HP bought out Boonton and marketed the 260A for a few years before presenting the HP4342 series of Q Meters. I have not confirmed but was told that HP abandoned use of the precision resistor in the 4342A. I used the 4342A for years in the 80's through the early 2000's. It is a tough box and in my opinion, if you can buy one with unmolested innards, it is well worth the cost of repair and calibration in a metrology lab. |
Re: clock calibrator
Since I don;t have a VNA and constructing one in less than a couple of months is a major undertaking, I would probably take the following approach, using components in my junk box.
1) Construct 2 input channels, each with a MAX913 comparator, 2 74HC40102 dividers and a 74HC series gate (for truing up the counter output to be synchronous with the 10 MHz input). This will pruduce a 10 Hz train of pulses (1 kHz if only one 74HC40102 is used). 2) One channel is the start input of a Time-to Amplitude Converter (TAC). The other is the stop input. 3) The TAC outputs (at a 10 Hz or a 1 kHz rate) can be fed to a MCA or examined in some other way to determine the frequency difference. An Ortec 566 TAC can achieve 15 pS resolution (100 nS range) or 10 pS (50 nS range). Jitter in the MAX913 and the 74HC logic would add some more uncertainty (currently being evaluated for another project). Since the 566 TAC is a general purpuse instrument, a dedicated TAC, built specifically for this purpose might achieve the same or better resolution. Measurement resolution (15 pS/10S) in the 10-12 area might be possible (15 uHz) and 100uHz might be readily in reach. Stephen Menasian |
Re: Making a Q-meter /
I think the best method for measuring loaded and unloaded Q and at the same time the coupling factor
is by using the resonance circle that is created on the smith chart of VNA. Read the article by late Darko Kajfez from Miss State Univ (olemiss) it becomes a bit tricky for very high unloaded Q or very weak couplings but still usable and I had seen several imporvements on that method but i cannot remember where...I have only used his method with a simple computer program that I wrote to analyze the S11 collected from the VNA |
Re: Making a Q-meter /
On Thu, Aug 4, 2022 at 02:14 AM, Kenneth Greenough wrote:
Re' a Q meter. I have seen something like that before. One problem is that it would not work with coils in screened enclosures, some of which are.?I have ordered a couple of Boonton Q-standards, which I was hoping to measure. Those are in screen enclosures, with 2 banana plugs 25.4 mm (1¡±) apart.?See attached photograph.?
|
Re: Making a Q-meter /
I see another problem that I was unaware of. There are two definitions of Q I am aware of? Here¡¯s a proof that the two definitions are equal?
|
Re: Making a Q-meter /
¿ªÔÆÌåÓýRe' a Q meter.This is only a thought idea, not tested hi. Have a small enclosure with sockets on opposite sides. each socket has a single turn loop attached with sufficient distance between the two loops to allow the coil under test to be positioned between them. one turn is fed from your sig' gen' and the other feeds a scope or RF meter.? A sufficient level from the sig gen may require an amplifier. With the coil to be tested placed between the loops tune the sig gen for maximum throughput, ie resonance. Then tune the sig gen lower until the throughput is 3dB down, ie 0.71 reduction in voltage or 1/2 power. Repeat by tuning the sig gen higher than resonance for a similar drop in throughput. the resonant frequency divided by the bandwidth just ascertained is the Q of the coil. The coil being tested can, of course, have a capacitor in parallel if wished. 73 Ken G8BEQ ? On 04/08/2022 09:19, Dr. David Kirkby,
Kirkby Microwave Ltd wrote:
|
Making a Q-meter /
I am looking at the possibility of making a Q meter for the frequency range 1.5 MHz to 150 MHz. One common way to do this is to inject a voltage from a LOW impedance source, into a series tuned circuit, then measure the voltage developed across the tuned circuit with a high input impedance voltmeter. The Q is the ratio of those two voltages.?
/g/Test-Equipment-Design-Construction/files/Q-factor%20%28Q%20is%20the%20inverse%20of%20dissipation%20factor%20DF%29. 1) The meter has measurement ranges of only 0-20 and 0-100, but many inductors have Q¡¯s much higher than 100. The HP 4342A can measure Qs from 5 to 1000. The old Boonton 160-A measures Q up to 640 (from memory).? ? I have not yet looked at the HP 4343A meter service manual. That will probably give me some ideas.? I suspect the answer is to not try to get such a low output impedance, accept that the voltage generated across the LC combination will be less than the Q, and correct for that in software.? Any other thoughts??
?
|
Re: clock calibrator
¿ªÔÆÌåÓýThis is the method used for the VNWA frequency meter. Also for generation Allan deviations on the fly. Kind regards Kurt ? Fra: [email protected] <[email protected]> P? vegne af Dr. David Kirkby, Kirkby Microwave Ltd ? Another way I have heard about to align two 10 MHz oscillators it to use a VNA. Perform a 2-port calibration of the VNA, which must be fed from a GPSDO or similar. Set the VNA to measure S21, and feed the oscillator you wish to align into port 2. If?the phase of S21 remains constant, then the two oscillators are at the same frequency.? ? ? |