Re: TEK 466 Analog Storage
A good start would be to have the controls thus: POWER Off INTEN Fully ccw FOCUS Midrange SCALE ILLUM As desired VERT MODE Ch 1 POSITION Midrange VOLTS/DIV 5V VAR VOLTS/DIV Calibrated detent AC-GND-DC
By
Colin Herbert
·
#150563
·
|
Re: TEK 466 Analog Storage
I have a feeling that it depends on whether or not it is in storage mode at startup. With storage mode turned off, IIRC, any storage scope ought to look "normal". In storage mode, then you'll get a
By
Harvey White
·
#150562
·
|
Re: TEK 466 Analog Storage
Thanks, but that link you provided only seems to point to an extremely high resolution scan of the front cover of the manual (69Mb's worth!) unless there's something I'm doing wrong! But no worries, I
By
Paul B.
·
#150561
·
|
Re: Tektronix 465,NO Trigger
Hi Bert No its not the Tunnel Diodes that are faulty its the Trigger driver JFET's have blown, in fact one of them was cracked which I was unable to see until I removed the Heatsink, so I hope someone
By
John Stoole <john@...>
·
#150560
·
|
Re: Tektronix 465,NO Trigger
I'm really surprised that no one else has mentioned the infamous tunnel diode problem. This often shows up on a lot of the older Tek products and the are many, many previous discussions on this and
By
Bert Haskins
·
#150559
·
|
Re: tektronix 214 option 94
This scope is currently on eBay. If it is of any significance, the probes are permanently wired in, according to the description. Colin.
By
Colin Herbert
·
#150558
·
|
Re: tektronix 214 option 94/ Li-Ion fork
Hi! My tektronix 214,221 and 222 is now working with lithium batteries. /g/TekScopes/photo/66992/1?p=Name,,,20,1,0,0 /g/TekScopes/photo/66992/2?p=Name,,,20,1,0,0
By
Miguel Work
·
#150557
·
|
Re: Tek 2467b test 05 error
Excellent point, Artekmedia. I'll know soon enough! -Stefan
By
Stefan
·
#150556
·
|
Re: Tektronix 465,NO Trigger
Well Success at last Fabio !!!!! I did not realize that the FET's were in sockets under the heatsinks, I decided to change them and I still had no Trigger, as they were out I tested the FETt's on my
By
John Stoole <john@...>
·
#150555
·
|
Re: TEK 466 Analog Storage
This should be covered in the service manual: e.g. https://docs.telegraphics.com.au/Tektronix_070-1753-01_466_Storage_Oscilloscope_with_option_Service_Instruction_Manual_Revised_Nov_1982.tif
By
toby@...
·
#150554
·
|
Looking for ...
120-1087-00 transformer or complete HV module for a 7603. Chris
By
W9RP
·
#150553
·
|
TEK 466 Analog Storage
Hi all, I've finally got a mo to look at the 466 I bought sight-unseen a while ago. I'm not really familiar with analog storage scopes, though. I've been given to understand that after switching on
By
Paul B.
·
#150552
·
|
Re: Concentric A and B time-base knobs/interlocking
Interesting... Raymond
By
Raymond Domp Frank
·
#150551
·
|
Re: Tek 2467b test 05 error
Also in this day and age it is not sufficient to look at the just the unit serial number. There are so many organ transplant units floating around where a previous? owner swapped an early A5 in a
By
Dave / NR1DX
·
#150550
·
|
Re: Tek 2467b test 05 error
As Siggi has mentioned, the A5 board (processor control) is a frequent source of failure in these scopes.? But Tektronix went through several design iterations of the A5 board over the years
By
machineguy59
·
#150549
·
|
Re: Tektronix 465,NO Trigger
Good morning John, Yey! Seems like you've made some progress there! A reading of -7.9V on the pins 2,3 of U460 is not normal, if the input is set to EXT and the input jack is grounded... the output of
By
Fabio Trevisan
·
#150548
·
|
Re: Concentric A and B time-base knobs/interlocking
I had a flash of inspiration over this; I wondered if the Z-axis modulation was functioning properly. I tried putting a sine-wave into the CH1 vertical and a square-wave into the external Z input.
By
Colin Herbert
·
#150547
·
|
Re: Tektronix 465,NO Trigger
Good Morning Fabio, I have carried out the tests to the U640 as you suggested for both the A trigger and the B trigger here are the results A TRIG 14/15 20.4 mv 2/3 -7.9 volts !!!!!!!!!?????? B TRIG
By
John Stoole <john@...>
·
#150546
·
|
Re: Tek 2467b test 05 error
Thanks guys for the responses. I'm definitely new to scope repair. The 2467 could not be passed up but it did come with the faults when I purchased it. I do have an inexpensive digital scope that I
By
Stefan
·
#150545
·
|
Re: tektronix 214 option 94/ Li-Ion fork
I wouldn't pick to much on Li-Ion chemistry. There are a lot of deployed chemistries and construction methods. Some have problems some don't, just like NiCd & NiMH. I've seen plenty of NiCD rotting in
By
Kevin Oconnor
·
#150544
·
|