¿ªÔÆÌåÓý

ctrl + shift + ? for shortcuts
© 2025 Groups.io
Date
Re: 2467B focus/astigmatism anomaly...
I've got a 40kV Glassman 2kW I'm in the middle of converting from positive to negative.
By John Griessen <john@...> · #158993 ·
Re: 2467B focus/astigmatism anomaly...
Who was it that was building the "frankenscope" about 13 years ago? -Dave (another one)
By Dave Seiter · #158992 ·
Re: 2467B focus/astigmatism anomaly...
Hi Chuck The Fluke 40KV probe is 1,000M input, The only other quick solution is an electrostatic meter, but I doubt it would have the resolution that you require. George.
By george edmonds · #158991 ·
Re: 2467B focus/astigmatism anomaly...
You're scaring me Dave.... -Chuck Harris Dave Wise wrote:
By Chuck Harris <cfharris@...> · #158990 ·
Re: 2467B focus/astigmatism anomaly...
That is where I went first, but when my 100M Keithley probe is applied to the focus grid, it pulls the scope out of focus. It is a problem when the measurement equipment adversely affects the measured
By Chuck Harris <cfharris@...> · #158989 ·
Re: 2467B focus/astigmatism anomaly...
If you have another power supply that can make 15KV, you could assemble a differential voltmeter using the supply, a battery-powered voltmeter as null detector, and another voltmeter to measure the
By Dave Wise · #158988 ·
Re: ebay old tektronix equipment
Ha!? I have a buddy who has a 3-car garage filled top-to-bottom with test equipment, wire and cables in containers in the back yard, and at least one storage unit.? Last time I was at his house he
By Jim Ford · #158987 ·
Re: 2467B focus/astigmatism anomaly...
Why don't you use a Fluke 40KV probe these are designed to work with a 10M Ohm input DVM George
By george edmonds · #158986 ·
Re: ebay old tektronix equipment
"Thousands of square feet..."? I wish I had that problem... or maybe it's a good thing I don't!?? I just agreed to buy an early (maybe 20s?) German horizontal milling machine and I really don't
By Dave Seiter · #158985 ·
Re: 2467B focus/astigmatism anomaly...
I thought of using a Fluke differential voltmeter that does have virtually infinite input impedance when it is in balance, but it too is only 1KV. I am going to have to figure out a way to sneak
By Chuck Harris <cfharris@...> · #158984 ·
Re: ebay old tektronix equipment
Yep, something all of us old-timers need to do.Thanks for the reminder.?Jim Ford?Sent from my Verizon, Samsung Galaxy smartphone
By Jim Ford · #158983 ·
Re: 2467B focus/astigmatism anomaly...
Yep, I figured it was a long shot.Thanks, Chuck.?JimSent from my Verizon, Samsung Galaxy smartphone
By Jim Ford · #158982 ·
Re: 2467B focus/astigmatism anomaly...
By Lawrance A. Schneider · #158981 ·
Re: ebay old tektronix equipment
Dad's carefully kept collection now dumped on the kids to figure out what to do with. I think of this dilemma often with my thousands of square feet of equipment storage. But some time has been
By Greg Muir · #158980 ·
Re: 2467B focus/astigmatism anomaly...
The anode voltage is 15KV... I the 34401 tops out at 1KV with 10M input resistance. Not even close. -Chuck Harris Jim Ford wrote:
By Chuck Harris <cfharris@...> · #158979 ·
Re: TOPIC CHANGE: Pro's and Con's of the 576 and 577 Curve Tracers. WAS: 5xx 'Scopes
All: While we are talking about Curve Tracers. Does ANYONE have a spare 575 CRT that they would part with? My brother has a 575 and it has a bad CRT. If you can help, please message me off
By Mlynch001 · #158978 ·
Re: 2467B focus/astigmatism anomaly...
Hmmm... maybe use an HPAK 34401A DMM in high resistance mode?? Don't know what your anode voltage is - it could exceed the 34401A's range.? Just a thought.Jim Ford?Sent from my Verizon, Samsung
By Jim Ford · #158977 ·
Re: 2467B focus/astigmatism anomaly...
Hi Leo, Thanks for doing the test. The problem with making measurements in the focus area of the 2467B is the extreme high impedance of the circuitry. Even my 100M HV probe loads the circuits down
By Chuck Harris <cfharris@...> · #158976 ·
Re: 5xx 'Scopes
Ed, That's pretty much why I bought a 5111, that is to keep my 577 company on the bench.?? I had given my 5223 to the museum and was cleaning out 5000 plugins until I bought the 5111. Bob.
By [email protected] · #158975 ·
Re: Need Help Troubleshooting Tektronix PS280
Hi Dave, I looked at the manual the other day- wow, it's definitely not a real Tek product!? Assuming the displays are working correctly, I would work my way from the transformer secondary, looking
By Dave Seiter · #158974 ·