Re: Tektronix 2465A... With error code Test 05 Fail 24
Well I have replaced the U500 and the scope is triggering just fine. However; when I changes Volts/Div scale without changing the input signal the display jumps off center. If I recenter the
By N7ZMG · #207734 ·
Re: FOUND: broken Tektronix WM490f (to 140GHz) and/or WM490d (to 170GHz) mixer(s)
Fyi: The jlcpcb cnc factory on youtube...not a small operation. https://youtu.be/YuSApjBn_zY?si=e38M-DFYBbrykJgV
By Edward Prest · #207733 ·
Re: Substitute parts for 12au6s on 575 curve tracer
You can use 6AU6 in place of 12AU6. They have identical characteristic, the only difference being the heater voltage, 12.6 V @ 150 mA for the 12AU6 vs 6.3 V @ 300 mA for the 6AU6. You can compensate
By unclebanjoman · #207732 ·
Trouble Getting Screenshots From Tektronix TDS 754D
Hello, I have a TDS754D. Works fine but I¡¯m having trouble getting some simple screenshots out. I used the hardcopy function, export as .bmp mono color and export to floppy. For some reasons it will
By Nick Stanley · #207731 ·
Re: Substitute parts for 12au6s on 575 curve tracer
There were actually serious attempts at replacing some tubes with JFET alternatives in very-high-volume applications like telephony (1960s-70s). I think the most well known is Teledyne's "FETRON,"
By ed breya · #207730 ·
Re: Substitute parts for 12au6s on 575 curve tracer
And other examples of replacing a 6AK8 tube ( triple diode/triode tube ), with a solid state replacement in HF Rx receiver using HV MosFets. https://www.youtube.com/watch?v=KlXMnNx7ZDg
By garp66 · #207729 ·
Re: Substitute parts for 12au6s on 575 curve tracer
Any possibility that a HV MosFet might be substituted ? LND150 Mosfets have been substituted for a different tube ( the 12ax7 ; which is Not the 12au6 discussed herein for the Tek 575) See:
By garp66 · #207728 ·
Re: Replacement Fan for 2465B/2467B
Common NIDEC or Mineba fan, usually the original can be recussitated if correct procedure is used. Jon
By Jean-Paul · #207727 ·
Re: SC503 -12V power supply
Hi Barry, I think you fixed the root cause of why shorting across C-E of Q8051 didn't work. I was expecting to see the same voltage as if Q7042 was removed (+0.6V) but at this point the regulator
By Ozan · #207726 ·
Re: Replacement Fan for 2465B/2467B
Is there a good solid replacement? Thanks, Monte I have bought many of these for 2465/67B scopes over the years with good results. eBay listing: Nidec Beta V FAN TA300DC 12V 0.16 Amp A33315-16
By Craig Cramb · #207725 ·
Replacement Fan for 2465B/2467B
I've read through a lot of posts regarding the replacement fans. Is there anything outside of the parts description from the service manual that I need to take into account? Both my 'scopes need the
By Monte56 · #207724 ·
Re: Tektronix 454A Oscilloscope No Trace but Beam Finder works fine?
I switched to x1 and the MAG light tuned off. Only UNCAL is still on, even though I have already cleaned the pot. Switched the triggering back and forth between AC and DC. Set the Source switch to
By Jeffrey Wang · #207723 ·
Re: SC503 -12V power supply
Ozan, I found a broken trace which I repaired. I also swapped out a few transistors. Today I am measuring -0.48 volts when I short across C-E of Q8051. I understand that 2N2222A is a lower
By Barry Breaux · #207722 ·
Re: Tektronix 454A Oscilloscope No Trace but Beam Finder works fine?
Yes I just downloaded it. Will spend some time reading it.
By Jeffrey Wang · #207721 ·
Re: Tektronix 454A Oscilloscope No Trace but Beam Finder works fine?
CH 2 is the same. Vertical Position knobs work in Beam Finder mode.
By Jeffrey Wang · #207720 ·
Re: Tektronix 454A Oscilloscope No Trace but Beam Finder works fine?
Thank you Ben for the detailed info - I really appreciate it! Here are my measurements for the reference points you listed: Junction between C369 and L369: 12.04V L343 (any side is fine for just
By Jeffrey Wang · #207719 ·
Re: Repair Tektronix TDS 520
Thanks for the reply. Today I unsoldered about 20 chips that were near the capacitors to see if there were any interruptions in the conductor tracks. I didn't find any interruption. How do you think
By Paul · #207718 ·
Re: Tektronix 454A Oscilloscope No Trace but Beam Finder works fine?
- Adjusted Ch1 var and the "uncal" light is off, as I mentioned in post #207706 - The var for A & B time base is in detent, as you can see clearly in my video. That said, I cleaned the pot and it
By Jeffrey Wang · #207717 ·
Re: Tektronix 454A Oscilloscope No Trace but Beam Finder works fine?
Jeffrey, The uncal lights and x5 lights are still on. The uncal lights should turn off when you go to the detent position which you will feel. As others have said, the switch could be dirty making
By Mark Vincent · #207716 ·
Re: Tektronix 454A Oscilloscope No Trace but Beam Finder works fine?
Still trying to get a bit further without having to take measurements may yield more info. If both channels show the same behavior: Since bad contacts are rather frequent with old equipment like
By Raymond Domp Frank · #207715 ·