Re: Tek 7704a Incorrect Readout Position
CLA pot?
check if defective and replace pot?
J
By
Jean-Paul
·
#207306
·
|
Re: TDS744A flicking
Hi, I own a TDS694C which has a similar power supply. Worked fine but after
sitting on the shelf for a long time died on the first power up.It produced
modulated high pitch sound on standby power,
By
Andrei Yershov
·
#207305
·
|
Tek 7704a Incorrect Readout Position
Hey everyone, I’ve got a Tek 7704a where the readout that SHOULD be at the top and bottom is offset and near the center.
Adjusting the vertical separation pot didn’t fix it, it won’t adjust
By
talynprior@...
·
#207304
·
|
FS: Tektronix Type 1101 accessory power supply
FS: Tektronix Type 1101 accessory power supply
LEMO connectors (4) on front panel.
+15VDC -15VDC +5VDC
TESTED, Works as it should.
Serial # B042132
$150 by PayPal or Zelle
Ships from Dallas,
By
Ron
·
#207303
·
|
Re: PCB for 465B Electrolytic Capacitors
Those work well and I've used them in a couple of Tek scope power supplies. The thing I don't much like about them, at least for this application, is once you solder them in place, then if you ever
By
n4buq
·
#207302
·
|
Re: PCB for 465B Electrolytic Capacitors
Strange, I emailed a reply and it never showed up.
On Ebay, search for Tektronix capacitor adapter. Somebody is selling a set for the 465 for $10 or best offer. I don't know the seller but they've
By
Paul Amaranth
·
#207301
·
|
Re: Free... G-8688-04 plotter board
Hi George,
I just bought a Tek 2232 with GPIB option. I see that you intentionally installed this option on your 2232 a couple years ago. Maybe for transferring data to a computer? It seems the
By
HF
·
#207300
·
|
Re: TDS744A flicking
[email protected]> wrote:
The TVSes should measure open in both directions, until you reach somewhere
beyond their threshold voltage. This is 220/250V respectively, which is why
they're challenging
By
Sigur?ur ?sgeirsson
·
#207299
·
|
Re: TDS744A flicking
Hello everyone,
Thank you for your quick responses. I will follow your suggestions as soon as I return from work. I’m afraid I don’t have a curve tracer. Perhaps there’s another way to test
By
rafael suarez
·
#207298
·
|
Re: TDS744A flicking
Hello everyone,
Thank you for your quick responses. I will follow your suggestions as soon as I return from work. I’m afraid I don’t have a curve tracer. Perhaps there’s another way to test
By
rafael suarez
·
#207297
·
|
Re: TDS744A flicking
Hey Rafael,
I second Jared's advice; you'll probably want to have another look at the
PSU. The standby supply in these scopes is notorious for giving up, leading
to dead or hiccuping supply. AFAIK
By
Sigur?ur ?sgeirsson
·
#207296
·
|
TDS380 Scope Stopped Booting - PSU?
I turned on my TDS380 yesterday and the CRT flashed some garbage and the front panel LEDs all came on and stayed on but that's it. I'll pull it apart this weekend and start looking for obvious failed
By
Richard Peterson
·
#207295
·
|
Re: Any real difference between 7904 & 7904A
7904 psu service is a nightmare. I had to do a full recap and hated it. Instead of disturbing the custom transformers, I clipped the leads and soldered to them. Still love the 7904 for it's
By
Andreas Matz
·
#207294
·
|
Re: TDS744A flicking
It looks like the PSU is hiccuping. First up, I'd try unplugging the grey ribbon cable to the CRT board to isolate it from power to check it's not still bad somehow.
Have you replaced C17 in the PSU
By
Jared Cabot
·
#207293
·
|
Re: Any real difference between 7904 & 7904A
In particular, the ability to raise the upper half of the later models onto the integrated "kickstand" in order to more easily access the controls and components in the lower half of the 'scope makes
By
Dave Daniel
·
#207292
·
|
P6042 repair
Hi all,
I have P6042 current probe that works well but the thumb button has popped off the slide assembly. The probe is still perfectly usable by manually sliding the head open and closed. The thumb
By
Morris Odell
·
#207291
·
|
Re: PCB for 465B Electrolytic Capacitors
Barry
That's very useful - thanks!
Menahem
By
M Yachad
·
#207290
·
|
Re: Replacing Electrolytics in a 465/465B
Barry,
I read that I should see a link to your full email address. However, I don't see one. I'd like to purchase a cap replacement board. Please email me at jonwoellhaf@....
Thanks!
Jon -
By
jonwoellhaf
·
#207289
·
|
Re: A6302 Current Probe - AM503B 588 Error, No DC function
For the A6302 probe, you are correct, the AC signal is generated by the coil, while the DC signal is created by the hall-effect sensor. Might be a broken wire inside the probe, or possibly inside the
By
groups.io@...
·
#207288
·
|
Re: Replacing Electrolytics in a 465/465B
Excellent!
Please send an email to <jonwoellhaf@...>, and we'll work out payment and shipping.
And -- duh -- I just noticed the link in your first post in this thread that has a plethora of
By
jonwoellhaf
·
#207287
·
|