Re: Tek 284 Pulse Gen
I don¡¯t want to spam the list, But I will put it out there since it has been a while. I have video content up on you tube specializing in restoration repair and alignment of test equipment in
By
Zentronics42@...
·
#192811
·
|
Re: Tek 2232 Scope
I've seen somebody on the eevblog forums had the same failure, he ended up soldering the wires directly to the board underneath.
By
Li Gangyi
·
#192810
·
|
Re: Peak Detector to Validate Tek Const. Ampl. Signal Gen. (067-0532-00)
Jim, Both versions of the manual state that the batteries should be replaced ¡°when the voltage drops below 1.33 V, or after one year¡± but I will try adjusting the unit with the current batteries
By
Jeff Dutky
·
#192809
·
|
Re: Tek 284 Pulse Gen
Zen, can you please post the info here, so that everyone can benefit? Thank you very much.
By
JRN
·
#192808
·
|
Re: Tek 284 Pulse Gen
Dave I will shoot you some information off list. Zen
By
Zentronics42@...
·
#192807
·
|
Re: 2465 Calibration
Mark, The spec on the GI ER1400 data sheet does indicate that the data is retained for a minimum of 10 years but does not indicate if that is when the device is never accessed/written to over that
By
Greg Muir
·
#192806
·
|
Tek 284 Pulse Gen
I've recently acquired a 284 that seems to need some TLC. Before I start snooping around in it, I'm wondering if there are any "usual suspects" that I should consider. Besides all the electrolytic
By
DaveH52
·
#192805
·
|
Tek 2232 Scope
Picked up a Tektronix 2232 scope at a Club auction recently. It has a dot with the "beam finder"but no trace. You can move the dot vertically and horizontally with the position knobs. Opening it up I
By
Harrison
·
#192804
·
|
Re: Upload a picture
Hi John, Michael Lynch just taught me how to post a photo. Visit the below included link. The instructions are easy to follow. Regards, Ken
By
ken chalfant
·
#192803
·
|
Upload a picture
Hello Can a picture accompany a message or does it require it being sent to the Photos? How do I do that? Thanks
By
JohnEdward
·
#192802
·
|
Re: Tektronix P6021Current Probe Termination
Looks like it's missing a second page. BTW, I have a spare Model134 Amplifier, if anyone wants to buy one.
By
Jim Adney
·
#192801
·
|
Re: Peak Detector to Validate Tek Const. Ampl. Signal Gen. (067-0532-00)
!.25 V sounds just about right for a mercury cell, but low for an alkaline.
By
Jim Adney
·
#192800
·
|
Re: 2465 Time-Marker Calibration Procedure
By Delay Position, do you mean the deltaREF knob? Thanks so much, guys, for all your help. This is my first Tektronix, or any electronic restoration for that matter, so thank you for being patient
By
Jimbert
·
#192799
·
|
Re: 2465 Time-Marker Calibration Procedure
A couple of <corrections>. Sorry about that. Ross
By
si_emi_01
·
#192798
·
|
Re: 2465 Time-Marker Calibration Procedure
I spent a lot of time calibrating a 2465A for a nephew today. I have all of the Tektronix Calibration equipment to perform the calibrations. I have also interleaved some responses to your questions as
By
si_emi_01
·
#192797
·
|
Re: Peak Detector to Validate Tek Const. Ampl. Signal Gen. (067-0532-00)
Jared, I compared the two versions of the manual for the 067-0625-00 (the early one using mercury cells, and the later one using alkaline manganese oxide cells) and there is no difference in the
By
Jeff Dutky
·
#192796
·
|
Re: 453 trace finder location
Pat, The trace location is under the graticule intensity knob on the left. Likely the contacts are dirty. C1194 and/or C1181 is the one you are talking about with the corroded lead. The first one I
By
Mark Vincent
·
#192795
·
|
453 trace finder location
Where is the trace finder typically located on a 453? Anywhere, centered who knows? Dug out the 453 last week and started looking into maybe or hopefully getting it going again. It looks like the X
By
KD4OBQ
·
#192794
·
|
Re: 2465 Time-Marker Calibration Procedure
Interleaved, at least what I can answer Tektronix time mark generators can put out narrow spikes with a sharp leading edge (so you can tell where it is).? They generally have intervals from 1 ns
By
Harvey White
·
#192793
·
|
2465 Time-Marker Calibration Procedure
1) The procedure calls for a Time-Mark Generator. Is this really just a pulse generator? What is the pulse width? 2) Is the Time-Mark a positive pulse, i.e. it does not go below gnd in its low state?
By
Jimbert
·