Re: Help troubleshooting a 2710 SA
Hi Jaap, This is Rogerio from Brazil. I own some 271x SA's that seem to have errors in the Display storage boards (most of them I could buy for a very low price from users (not companies) so there was
By Rogerio O · #175971 ·
Re: 7613 Mainframe - Readout problem
Look for a signal called x/y shutdown. It originates on the readout board (J2138), goes via coax to the horiz amp (J503), then a 2-pin ribbon (P504) to the main interface board (P38) and then pin BC
By teamlarryohio · #175970 ·
Re: 7613 Mainframe - Readout problem
Attilio, That looks like there is a fault related to the vertical channel switch, U214 (schematic 3 - Vertical Interface). There is an input to U214 labelled 'X-Y shutdown', this disconnects the
By Roger Evans · #175969 ·
Re: 7613 Mainframe - Readout problem
Sorry, I wanted to add that moving the track only horizontally with the Position command of the 7B70 time base the readout does not move. -- Cheers Attilio
By Attilio · #175968 ·
Re: Spectrum analyzer Tektronix 7L13 on mainframe Tektronix 7603
Thank you so much Tom, I will follow your precious advice, acting with the utmost caution. Excellent idea of ??taking a picture of the position of the filter adjustments before acting. --Cheers
By Attilio · #175967 ·
Re: 7613 Mainframe - Readout problem
thank you all I discovered an important thing: if I move the track vertically with the position command of the 7A16A plug-in also the readout moves as if it were a single screen, instead the readout
By Attilio · #175966 ·
Re: Mechanism of CRT Double Peaking (UPDATED)
Hi Jeff, 5103N appears to be only the module mainframe, separate from the CRT unit. However I found the CRT circuit to be similar (if not identical) among the 5000 series models, so no problem there.
By Nenad Filipovic · #175965 ·
Re: Intermittent power on problems with 2465B
Mark: Doubt it is the A5 SMD cap, BUT that MUST be checked and fixed on any old 2465B. Any relation to the mains voltage, Check turn on with a variac? Suggest to open it and fix, and NOT exercise
By Jean-Paul · #175964 ·
Re: Intermittent power on problems with 2465B
Hi Mark, I would start by recapping the PS, the intermittend not starting makes me feel it's in there. The SMD caps on the A5 board would cause all sorts of funtions on the frontpanel not working as
By satbeginner · #175963 ·
Intermittent power on problems with 2465B
I have a tek 2465B (serial# B077725, so could have the infamous SMD cap issues) that sometimes does not start. Typically, this is in the morning, and after I hit the switch 2-3 times, it starts. This
By Mark Hatch <mark2382@...> · #175961 ·
Re: Quick question 564B
It is a pretty well known phenomenon... well discussed on this group over the years... often by me. Around about the time the 547/545B scopes got designed, tektronix decided to make a universal HV
By Chuck Harris <cfharris@...> · #175960 ·
Re: Quick question 564B
Chuck, Please tell me more about the 564B HV transformer epoxy disease! Thanks! Pete On Sunday, January 3, 2021, 8:07:20 a.m. EST, Chuck Harris <cfharris@...> wrote: It is a cute little scope,
By petertech99h · #175959 ·
Re: 7613 Mainframe - Readout problem
If I remember the threads properly, the thermal compensation (dancing readouts) fixed a small wobbling problem.? This sounded like a much larger deviation, and the picture I was thinking of was, say,
By Harvey White · #175958 ·
Re: 7613 Mainframe - Readout problem
and I must be thinking of another 7K frame as I couldn't find it in the manual I have. It's also been called thermal compensation, or it may not be there on this model. But the symptom was dancing
By teamlarryohio · #175957 ·
Re: Spectrum analyzer Tektronix 7L13 on mainframe Tektronix 7603
No problem, Attilio! :) As to tuning the filter, it's very important to take a good picture (or several) showing clearly the initial state of all of the adjustments before you started adjusting them.
By Tom Lee · #175956 ·
Re: 7613 Mainframe - Readout problem
There is a low frequency compensation adjustment on the vertical output amplifier. If your power supplies are ok, check that next. -ls-
By teamlarryohio · #175955 ·
Re: Spectrum analyzer Tektronix 7L13 on mainframe Tektronix 7603
Thanks Tom, sorry, but I'm a little old and sometimes I miss something. I have read the cavity filter tuning procedure, the note that warns me that if you lose the signal it becomes almost impossible
By Attilio · #175954 ·
Re: 7613 Mainframe - Readout problem
Thanks Harvey, I have not yet tested the voltages and their ripple on the power supply, I thought it was a fault on the Readout board, but I replaced it and the defect remained. Tomorrow I'll check
By Attilio · #175953 ·
Re: 7613 Mainframe - Readout problem
Have you checked power supply voltages and ripple with the two sets of plugins inserted?? Wandering across the screen suggests both X and Y problems, and changing with load suggests power supply.?
By Harvey White · #175952 ·
Re: Spectrum analyzer Tektronix 7L13 on mainframe Tektronix 7603
Hi Attilio, I encourage you to read the article I linked you to earlier. You do have to retune the filter only a few percent. You cannot skip that step¡ªthe filter has a very steep rolloff so a few
By Tom Lee · #175951 ·