¿ªÔÆÌåÓý

ctrl + shift + ? for shortcuts
© 2025 Groups.io
Date
Re: Dead 7603
You've got a dead 7603 with no lights. Definitely look at the low voltages first. Check the 0.15A (150mA) fuse, F855, just because its 'filament' is so delicate it can easily become disconnected from
By Al Holt · #149173 ·
Re: Dead 7603
Does it make a ticking sound - like 1/sec? You may need to put your ear close to the cover to hear it. If so, it's something in the PSU - rectifier board, LV board, inverter - likely a shorted
By JJ · #149172 ·
Re: Dead 7603
That diagnosis gives no information regarding the location of any HV fault. If the -3kV is not present, the fault could either be the HV transformer, something in the string of components associated
By Craig Sawyers <c.sawyers@...> · #149171 ·
Re: Tek 465 no display
You probably meant .3 * 18, or 54 W? It's not entirely fair to think of 18 V in both cases since the voltage drop Vce will increase dramatically when the line is short circuited. In the 465 the
By Albert Otten · #149170 ·
Re: Dead 7603
That model, if the graticule lights work you got juice to the Xfmr. Other lights are a different story. If you get absolutely nothing, turn it on and see if the blades of the power plug read infinity
By Jeff Urban · #149169 ·
Dead 7603
Hi, my name is Jim and I'm a Tek-aholic. ("Hi, Jim.") Long time lurker (about a year and a half), first time poster. My latest Tek project is a completely dead 7603 that I bought on the 'Bay. I did
By Jim Ford · #149168 ·
OT: LED lamp bulbs at ridiculously low prices at (some) Costco stores
This is OT, but I just have to say I happened to stop by the Fremont Costco today, and stumbled upon some amazing deals on certain types of LED bulbs, like six-packs of PAR-xx/xxW equivalent spots for
By Ed Breya · #149167 ·
Re: Tek 465 no display
One thing that wasn't said on this thread ( unless I missed it) is that these scopes can be LETHAL !!! If you get "stung" by CRT high voltage (-1450) if it doesn't kill you, you'll be one hurtin
By HBcubed · #149166 ·
Re: Tek 465 no display
True, and I've made some of them. However, as I remember it, the voltage across the sense resistors should be the VBE drop of the trigger transistor. I'm also going on general theory without looking
By Harvey White · #149165 ·
Re: Novice question - 475, what mode is this?
Hello Dimitri, To you and all, first of all, apologies if my answer is maybe mission the history (your previous messages) but I'm writing from my phone and couldn't let this go. Itseems no one
By Fabio Trevisan · #149164 ·
Re: Novice question - 475, what mode is this?
Should help, but that jitter might be settings. If B sweep has an auto trigger, turn it off. Also switch it to DC coupling and turn off any low or high filters. Adjust the trigger level slowly and at
By Jeff Urban · #149163 ·
Re: Novice question - 475, what mode is this?
I meant Bob's computer monitor. If LCD which most are these days, some are not quite perfect. My 23" on my good old PC, I think it is an Acer has great viewing angle, about like a CRT. But my laptop
By Jeff Urban · #149162 ·
Re: I have two Tek manuals I don't think I need
There were people on here recently looking for donations for the cost of scanning a bunch of manuals and making PDFs out if them. I would have kicked in a few bucks but I do not have a credit card, so
By Jeff Urban · #149161 ·
DuraProbe 123 adjustment
i find myself with a Duraprobe 123 100 mHz probe, which is very nice but doesn't say tektronix on it anywhere. It has H and L adjustments which i assume are intended to get rid of the spikes using the
By John Ferguson · #149160 ·
Re: Tek 465 no display
Hi Harvey, I think you (and Dave before) forget here that the supplies are protected by foldback current limiting. In case of a nearly dead short the current will be much smaller than in normal
By Albert Otten · #149159 ·
Re: Agilent keysight HP group
Thanks everyone I just must have missed it or typed it in wrong
By Pete Lancashire · #149158 ·
Re: Agilent keysight HP group
https://groups.yahoo.com/neo/groups/hp_agilent_equipment/info -- Richard Knoppow dickburk@... WB6KBL
By Richard Knoppow · #149157 ·
Re: TDS 380 won't trigger
On Sat, 23 Jun 2018 at 20:45 Gary Appel <garyappel@...> wrote: > Thank you Siggi. > > The gentleman on EEVBlog has a 'scope that reported exactly the same > errors > as I have, so it sounds like
By Sigur?ur ?sgeirsson · #149156 ·
Re: Agilent keysight HP group
https://groups.yahoo.com/neo/groups/hp_agilent_equipment/info;_ylc=X3oDMTJmYjg3amVpBF9TAzk3MzU5NzE0BGdycElkAzEwOTI5MzMzBGdycHNwSWQDMTcwNTA4MzY2MwRzZWMDdnRsBHNsawN2Z2hwBHN0aW1lAzE1Mjk4NTY0NTI-
By Tom Miller · #149155 ·
Re: Agilent keysight HP group
This one? https://groups.yahoo.com/neo/groups/hp_agilent_equipment/info