Re: 453 trace finder location
Pat, The trace location is under the graticule intensity knob on the left. Likely the contacts are dirty. C1194 and/or C1181 is the one you are talking about with the corroded lead. The first one I
By Mark Vincent · #192795 ·
453 trace finder location
Where is the trace finder typically located on a 453? Anywhere, centered who knows? Dug out the 453 last week and started looking into maybe or hopefully getting it going again. It looks like the X
By KD4OBQ · #192794 ·
Re: 2465 Time-Marker Calibration Procedure
Interleaved, at least what I can answer Tektronix time mark generators can put out narrow spikes with a sharp leading edge (so you can tell where it is).? They generally have intervals from 1 ns
By Harvey White · #192793 ·
2465 Time-Marker Calibration Procedure
1) The procedure calls for a Time-Mark Generator. Is this really just a pulse generator? What is the pulse width? 2) Is the Time-Mark a positive pulse, i.e. it does not go below gnd in its low state?
By Jimbert · #192792 ·
Re: Tektronix P6021Current Probe Termination
Looks like Ken uploaded it to TekScopes; /g/TekScopes/files/P6021%20Amplifier.pdf George
By Glydeck · #192791 ·
Re: Thinking on CRT restorations for double peaking tubes.
He confirmed red glyptal. David J. Brown vintageTEK President https://vintagetek.org
By Dave Brown · #192790 ·
Re: 2432 calibration battery?
Yes, there is information on the web that covers it. I haven't done the modification. I just purchased my 2465BDM in 2016 for $100.00 +$53 shipping off of epay. I replaced the Back-up Battery, fixed a
By si_emi_01 · #192788 ·
Re: Peak Detector to Validate Tek Const. Ampl. Signal Gen. (067-0532-00)
And THAT is exactly how I lucked into this item: the part number was mistyped by the seller and that probably hid it from any other prospective buyers. ¡ª Jeff Dutky
By Jeff Dutky · #192787 ·
Re: Tektronix P6021Current Probe Termination
I uploaded a Electronic Design idea from a few decades ago for a simple amplifier for the P6021. I haven't built it so I can't attest to it. Ken
By khillen@... · #192786 ·
Re: Peak Detector to Validate Tek Const. Ampl. Signal Gen. (067-0532-00)
Jeff, note it's 067-0625-00. Albert
By Albert Otten · #192785 ·
Re: Peak Detector to Validate Tek Const. Ampl. Signal Gen. (067-0532-00)
Jared, Yes, it had occurred to me that adding a couple hundred mV to the battery voltage would require changes to the surrounding resistors to get back to a target voltage that correctly offsets the
By Jeff Dutky · #192784 ·
Re: Peak Detector to Validate Tek Const. Ampl. Signal Gen. (067-0532-00)
If you swap to Alkaline cells, swap R10 and R11 to 200K resistors to bring the adjustment of the trimpots back into a useable range due to the higher cell voltage over the original ones. Also, to
By Jared Cabot · #192783 ·
Re: Peak Detector to Validate Tek Const. Ampl. Signal Gen. (067-0532-00)
I lucked into a 067-0652-00 Peak to Peak Detector on eBay for much less than the going rate. When it arrived I immediately opened it up and checked the batteries. They read only about 1.25 V, while
By Jeff Dutky · #192782 ·
2432, 2232, and 2230 backup batteries
This is what I got for these scopes: 4X EEMB ER17335 Nonrechargeable 3.6V Lithium Battery with Tabs Li-SOCL? 2/3A Size 2100mAh High Capacity UL Certified Single-Use 3.6V Lithium Thionyl Chloride
By Bert Haskins · #192781 ·
Re: 524D RCA customized scope on the big sales platform
Yep, I'm sure those "customer is also supplier" kinds of situations happen a lot. I read about a company making pulse generators. They used IIRC Motorola transistors in the pulsers, which Moto in turn
By Jim Ford · #192780 ·
Re: 524D RCA customized scope on the big sales platform
That¡¯s hilarious, Bob. Thanks for that part of the story. I hadn¡¯t heard before about returning the gifts! Cheers Tom Sent from my iThing; please forgive the typos and brevity
By Tom Lee · #192779 ·
Re: DM501A Input Leakage Source
Roman, I¡¯m seeing somewhat less compared to what you see on the 200 mV DC range. And it varies between plug-ins with one being ~6 mV highest and the lowest being ~2 mV. Of course shorting the input
By Greg Muir · #192778 ·
Re: Tektronix P6021Current Probe Termination
That's too expensive for me.? It's nothing more than the ring and BNC from a standard tektronix probe with the ID pin shorted to ground.? I need to get around to making one. Harvey
By Harvey White · #192777 ·
Re: Tektronix P6021Current Probe Termination
If you are going to use the P6021 with a 7A14 the part number of the adapter is 131-0750-00. There is a guy peddling the adapter in new condition on ePay. The 7A14 can also use the P6019, P6020 and
By Greg Muir · #192776 ·
Re: 524D RCA customized scope on the big sales platform
Steve, You are correct. While working in Denver television at the ripe old age of 18 I worked with RCA TR-3, TR-4 and TR-70 machines and all had built-in video monitors and scopes. The scopes were
By Greg Muir · #192775 ·