¿ªÔÆÌåÓý

ctrl + shift + ? for shortcuts
© 2025 Groups.io
Date
2465B Parametric Freq Measurement Error
After successfully calibrating my 2465B, it reports "All tests passed" on start up. The question marks are all gone! My issue is with the parametric measurement of frequency. When I attempt to measure
By joe@... · #156513 ·
My TDR Evaluation of the SG503 012-0482-00 Cable has been uploaded
I uploaded the results of my testing the SG503 012-0482-00 cable. In addition I compared it to two other similar cables: a Tektronix 012-0057-00 50¦¸ cable, and a Pomona Electronics 50¦¸ cable. It is
By Dennis Tillman W7pF · #156512 ·
New file uploaded to [email protected]
Hello, This email message is a notification to let you know that a file has been uploaded to the Files area of the [email protected] group. *File:* Evaluation of the Tektronix 012-0482-00 cable for
By [email protected] Notification <TekScopes+notification@...> · #156511 ·
2246 no display/odd behavior
Ok. I've removed the battery - have a new one on order. When reassembling I found that J2208 was installed backwards initially. Corrected but no change in behavior. All further testing is done with
By Matt · #156510 ·
Re: SG504 or 503 cable
Ok, compared the RG-58 Tektronix cable to an RG-400 I built of similar length. At 200Mhz and 4Vpp (x1 amplitude) we have .... The OEM Tektronix cable clocked in at 0.20 dBm low. DIY RG-400 cable was
By Ancel · #156509 ·
Re: 2246 no display/odd behavior
How can you edit a post on this system? Anyway- Have you replaced the battery yet? Should do that first and reboot the scope several times to get the ram initialized. Pay attention to the adjustments
By Tom Miller · #156508 ·
2246 no display/odd behavior
Ok so some updates: I haven't found any previous posts that address my issues. All voltages on the low side are in spec. Looking back at the ripple measurements in my first post I think they are all
By Matt · #156506 ·
Re: Powering up a 577 for the first time
The 577 used one of the Tek custom potted SMPS transformers if my memory is correct. Next to a tube failure, that is your biggest worry. I agree with others. Clean, repair power cord, fuse, power,
By Kevin Oconnor · #156505 ·
Re: Powering up a 577 for the first time
I think a good visual inspection followed by the use of an ohm meter would be my first plan. Then repair the cord and install the correct fuse. If you have an ESR meter, check the bulk capacitors and
By Tom Miller · #156504 ·
Re: Powering up a 577 for the first time
If you use the correct fuse then the fuse will protect the equipment. If the fuse blows then you need to go into troubleshooting mode. -- Sent from my Android device with K-9 Mail. Please excuse my
By Jim Potter · #156503 ·
Re: Wanted: Earth leads with croc-clip for P6139A
I'm still wanting these if anyone can help. Thanks David
By David C. Partridge · #156502 ·
Powering up a 577 for the first time
All, While I was searching eBay for a transistor adaptor for my 576, I came across a parts 577 in good condition with both the adaptor and an original safety cover for a low price... and ended up
By groups@... · #156501 ·
Re: A Probing Question
Hello Everyone, Thank all of you who responded to this query and Raymond in particular.? I hope that there are a few other people in the group who benefited from this discussion.? I certainly did.?
By Jack Reynolds · #156500 ·
11801 10x rise time measurement error
I just got bit by this yet again. Does anyone know if it was fixed and what ROM version fixes it? Doing an "autoset" fixes it, but as I have not been able to identify what triggers it, it's hard to
By Reginald Beardsley · #156499 ·
Re: [Test-Equipment] Suggestions for going through Tek2465DVS/DMM?
Thank you Siggi for your email and information. I'll work on the scope today, despite needing to assemble few computers for my customers..... I was working way to hard and sometimes I need to
By Tony Fleming · #156498 ·
Re: SG503 amplitude precision
OK, thanks for this information Dennis! I think I wanted to make sure nobody's feelings are hurt.... But I'll stop answering the emails that aren't directly related to me. This setup is very different
By Tony Fleming · #156497 ·
Re: CRT Hood for 7000 series 'scopes?
Keep an eye out for a plastic flower pot at Home Depot.Cut or enlarge the hole in the bottom. HankC, Boston, WA1HOS
By HankC, Boston, WA1HOS · #156496 ·
Re: A Probing Question
Have a look here for something special, a bit more extreme: <http://www.ggb.com/> Raymond
By Raymond Domp Frank · #156495 ·
Re: SG504 or 503 cable
And a lot lower loss at higher frequencies. Loss can be a little hard to gauge with the usual TDR. The whole point behind the fancy tek cable is that the amplitude accuracy of the SG503 requires the
By Chuck Harris <cfharris@...> · #156494 ·
Transfer option cards from TDS 320 to TDS 350?
Hello from Dublin, Ireland. I have a question some of you knowledgeable folks might be able to advise me on. I have two apparently similar scopes, a TDS 320 (100 MHz, 500 M samples/sec) and a TDS 350
By Peter Mulvey · #156493 ·