Tek 485 Scope, Very strange Z axis modulation query
Here¡¯s my story, Tek 485, a ham fest purchase, it was dirty and nonfunctional. It's a pretty common issue with ham fair test equipment. After cleanup of the 485, I found it had a shorted Tant in the
By Michael - AF7U · #208068 ·
File /Tektronix TDS420_460 Schematics.pdf uploaded #file-notice
The following items have been added to the Files area of the [email protected] group. * /Tektronix TDS420_460 Schematics.pdf (
By Group Notification <noreply@...> · #208067 ·
Re: TDS420A capacitor rot and ACQ failure
Files put into a single PDF and uploaded to the files section.
By Tim Wright · #208066 ·
2465B focus at high brightness
My 2465B suffers from a miss-tracking of focus with intensity. When the intensity is turned up (75 to 100%) I have to re-adjust focus. When I do this the readout is out of focus (because it is a lower
By vk2bea · #208065 ·
Re: TDS420A capacitor rot and ACQ failure
I have been working on getting the schematics for the TDS 420/460 converted from production engineering schematics to a form which can be used without trying to create a manual. I have no background
By Ferrous Steinka · #208064 ·
Re: Tek Type 130 L-C Meter - confirming the meter capacitor 'error'...?
My caps were backwards as well. Looking at W140.com I see that they were added at SN 6040.
By circlotron · #208063 ·
Re: Tektronix TDS744A power supply repair - part help needed
Hello Chris, I am afraid there is no reasonable way how to rewind the transformer. It is very small (just 3x3x3cm), has 6 independent windings and broken one is closest to the core. It is also
By Vladimir Filip · #208062 ·
Re: Tek Type 130 L-C Meter - confirming the meter capacitor 'error'...?
It's quite hard to believe that every 130LC was drawn and assembled incorrectly, and not picked up during what appears to be one ofTek's longest runs of production. I do recall a piece of Marconi
By Richard Kelly · #208061 ·
Re: Tek 465M saga
Replying by memory, I believe it is correct.? In most cases, if the fuse that blows is the .250 mA fuse, check the tantalum caps nearby. The culprit mostly will be the HV block. There was another
By Ananda · #208060 ·
Re: Tek Type 130 L-C Meter - confirming the meter capacitor 'error'...?
I don't recall for sure, but on my two 130s I think those caps were also reversed from the "proper" polarity, but it didn't seem to make much difference to operation. The net DC voltage on the meter
By ed breya · #208059 ·
Re: Tektronix TDS744A power supply repair - part help needed
Hi, not what you want to hear, but have you considered re-winding it ? Either on the same or a similar former. If you're very lucky, the primary winding might be the outer one, or a separate former ?
By fenugrec · #208058 ·
Re: TEK 492 Spec Analyz advice, parts
Bad SMD? Not likely due to age. However, for the long term every board should be pulled (only one or two at a time) and inspected anyway, see. See
By Edward Prest · #208057 ·
Re: TDS420A capacitor rot and ACQ failure
Bummer. It was worth trying. [email protected]> wrote:
By Radu Bogdan Dicher · #208056 ·
Re: TDS420A capacitor rot and ACQ failure
Just checked, and it does not. Different ACQ board altogether. I have a 540, and it is similar to that. Makes sense, same series.
By Tim Wright · #208055 ·
Re: Tek Type 130 L-C Meter - confirming the meter capacitor 'error'...?
Same here, my caps were backwards. I also replaced a couple of the film caps there because they were DiFilm with high dielectric absorption which warped the low-scale reading. I found that polyester
By Dave Wise · #208054 ·
Re: Chasing an analogue video signal - help
There are many ways to approach it. You should be able to do this with a 7904 + 7B92A in delayed trigger mode. You'll need to display a complete frame with the main sweep and display the scan
By Chris Wilkson · #208053 ·
Re: TDS420A capacitor rot and ACQ failure
The TDS520 component level service manual is available, not sure if it applies to the 420. Radu. [email protected]> wrote:
By Radu Bogdan Dicher · #208052 ·
TDS420A capacitor rot and ACQ failure
All SMD caps replaced, board cleaned and inspected, via's reflowed if showing corrosion signs. Still getting trigComparatorTest FAIL, ERROR ID 151 diagnostic test failure trigComparatorTest, TRIGA
By Tim Wright · #208051 ·
Re: 7904 pre-260000 PSU crackling
When I first got this 7904 it has been stored unused for years. I turned it on and the scope, with some plug-ins, entered tick-mode. I spent months trying to figure out what was wrong to no avail. I
By G?ran Krusell · #208050 ·
Tek Type 130 L-C Meter - confirming the meter capacitor 'error'...?
I read Alan Hampel's article on restoration of the above. He asserts that "Surprisingly, electrolytics C99 (5?F) and C100 (25?F), factory originals, were installed backwards! Not surprisingly,
By Richard Kelly · #208049 ·