Re: Repairing LEVEL Control from a 7B50/7B90?
It turns out those aren't all that difficult to disassemble - at least not down to the pot that's at the furthest from the front. I was able to reinforce the connection that was intermittent and the
By n4buq · #190868 ·
Re: 7904A trigger selector again
Hi Mark, Thanks much for the tip. I'll give that a try. Neil
By Neil Barlow · #190867 ·
Re: 7904A trigger selector again
Neil, Replace the 9V zeners with BZX85B9V1. The originals will be open. The other 9V zeners in the scope will be open also. I replaced the axial type condeners in parallel with the zeners with film
By Mark Vincent · #190866 ·
Re: Red Filter on 7A13
The bezel I ordered came in today. I cut a small piece of the red filter and it works great! Thanks again for that suggestion. Thanks, Barry - N4BUQ
By n4buq · #190865 ·
Re: Repairing LEVEL Control from a 7B50/7B90?
Subject should have stated 7B50/7B70 instead of 7B50/7B90. Thanks, Barry - N4BUQ
By n4buq · #190864 ·
Repairing LEVEL Control from a 7B50/7B90?
While troubleshooting the triggering problems in my 7B50, I discovered that the LEVEL pot is intermittent on one side. This is one of those stacked, plastic pots with a switch at the end and serves as
By n4buq · #190863 ·
Re: 7904A trigger selector again
Hi Craig, I see that this is an old topic, but did you ever resolve this issue? If so, how? I have a similar story in process with a 7904A. Neil
By Neil Barlow · #190862 ·
Re: TDS5/6/700 application software packages?
Sounds like an "eye" diagram. It's a diagram of multiple overlays of data and clocks that show when data is stable (with respect to the clock). You display data while syncing to clocks, but allow long
By Harvey White · #190861 ·
Re: 7904 problems
I have added 16 images to the album showing the cal signal via a properly adjusted probe at 1 ms per division (horiz) and 1 volt per division (horiz). Each image has the trace position half a division
By Jeff Dutky · #190860 ·
Re: 7B50 - Schematic Question
I think I found the prime problem. The LEVEL pot is defective. I don't know what made me decide to check that but, when I did, I found that the end that connects to P260-7 isn't making good contact
By n4buq · #190859 ·
Re: 212, 213, was Re: [TekScopes] New kid on the block...
Roy, I described a modification to use a lithium-ion 18650 cell plus a boost converter in an earlier message: /g/TekScopes/message/165764 --John Gord
By John Gord · #190858 ·
Re: TDS5/6/700 application software packages?
I was idly looking at TekWiki the other day, and I noticed that on the TDS784 page (https://w140.com/tekwiki/wiki/TDS784) there's now a link to "Telecom Test and I-Pattern Software TTiP". I haven't
By Sigur?ur ?sgeirsson · #190857 ·
Re: 7904 problems
Hi all, I have uploaded some pics of the 7904 problem to Jeff's photo folder of distortion on his 465B . They show two 7904 traces, one with the calibrator square wave and one without. They are OK on
By Morris Odell · #190856 ·
Re: 7B50 - Schematic Question
It looks like DC level is different in different mainframes. R7903 shows signals centered around 0V. Diff pair following TP101/TP111 has good common mode rejection, 50-100mV DC shift at the input
By Ozan · #190855 ·
Re: 7904 problems
If you have a Variac which can handle the current, you can do even better. Connect the coil to the output of the Variac. When degausing, turn the Variac up to the highst level you are using. Then,
By SCMenasian · #190854 ·
Re: Tek 468 GPIB board
I have now written a Python program for reading the data and creating a snapshot. It uses a Prologix style Gpib adapter to interface with the scope. I would like it to interface with Linux GPIB, but
By John · #190853 ·
Re: 7904 problems
Oh yes! Been there, done that many times. And it usually works great. You can make a very effective degaussing coil by purchasing a 100 foot roll of #14 NM with ground electrical cable at your local
By Ralph Hartwell · #190852 ·
Re: 212, 213, was Re: [TekScopes] New kid on the block...
The batteries are 5/4 AA size nicads. The holders are far more difficult to find then the batteries. I don't know if anyone has made a 3D printed version of them. It is very bad for the power supply
By greenboxmaven · #190851 ·
Re: Super Clean 570 Curve Tracer on Craigslist
There is a picture of one of the early (gray) ones like this one on Tekwiki that has the same box/panel. I think they moved the box to the other side and made it longer and skinnier on the later blue
By Wayne · #190850 ·
Re: 212, 213, was Re: [TekScopes] New kid on the block...
Got it, thanks. I don't know why I had such a hard time with that other instance of the google drive file... Sometimes it works for me, and sometimes it doesn't. Either of you guys have any thoughts
By Roy J. Tellason, Sr. · #190849 ·