Re: Help with (another) broken 7B92 please
I can give some general answers, but perhaps not as helpful as you'd want. Considering the trigger circuits:? A good approach is to take the input waveform, compare to a level, then produce a
By Harvey White · #187457 ·
Re: Help with (another) broken 7B92 please
I have a general question about the diagnostic approach to this instrument: how are people figuring out what should be seen at a given test point? There are lots of test points on the sweep board, but
By Jeff Dutky · #187456 ·
Re: Help with (another) broken 7B92 please
Hi Bruce, 1.4V across CR750 is not a good sign, it is most likely in open circuit. You can short across CR750 with an ammeter and make sure the current is not excessive (I think parts list says ~
By Ozan · #187455 ·
Re: Help with (another) broken 7B92 please
/g/TekScopes/photo/260881/3326552?p=Created%2C%2C%2C20%2C2%2C0%2C0 Horizontal is in volts, Vertical 10mA/div Good diode has a resitance <10 ohms, in this example 4 ohms in reverse,
By Miguel Work · #187454 ·
Re: Help with (another) broken 7B92 please
I use a simple x y octoupus to check in board these diodes I have fix recently a 7T11 with bad TD outuput diiode, I replaced it with a 10mA russian diode and it now works.
By Miguel Work · #187453 ·
Re: 4051 Rampack Solid-state Tape Replacement
If enough people want them, a data recovery service might be able to read the tapes and rewrite them on fresh tape cartridges or another medium. I think the cost would be too high for one set of tapes
By stevenhorii · #187452 ·
Re: 4051 Rampack Solid-state Tape Replacement
What we really need are new fresh magnetic tapes loaded with original software for the Tektronix 4041. My old tapes (all 6 of them) are no longer readable by the tape drive. Gary Bosworth
By Gary Robert Bosworth · #187451 ·
Looking to buy Tektronix 661 plug-ins
Hi , I am wanting to find a couple of old Tektronix plug-ins . They are from the rather old 661 sampling oscilloscope . I need a 4S2A and also a 4S3 plugin to complete my system :- please note that I
By Brian · #187450 ·
Re: Help with (another) broken 7B92 please
I am not sure I understand what is happening so these are no more than comments. Have you set the P308/P309 jumper for 'non 79xx' mainframes? If you set the trigger mode to normal then do the
By Roger Evans · #187449 ·
Re: Help with (another) broken 7B92 please
I?m waiting to receive one 7B92 that I have bought cheap beacuse I need pulse diodes, and this time base has 6 fast TD diodes. I don?t understand what you neen to know. what does it mean free run
By Miguel Work · #187448 ·
TM506 + WR501 (without cables) + DD501 + LA501 is looking for a new home
Hello Rick, may you use the e-mail massner@... please. Best regards, Stephan
By Dr.-Ing. Stephan Massner · #187447 ·
Re: tek 2215 trace position issues
the voltages on the vert position pot wipers is as follows fully ccw=3.6v,halfway point =1.2v,fully cw =3.6v.
By Paul M3VUV · #187446 ·
Re: TM500 tester switch - Attn. Jared Cabot.
Here is a PDF of the datasheet for the Greyhill 71BDF30-02-2-AJN switch that was recommended for the tester. The stop pins are shown on page 11. < http://www.farnell.com/datasheets/20309.pdf> If your
By - · #187445 ·
Re: TM506 + WR501 (without cables) + DD501 + LA501 is looking for a new home
Greetings Stephan, A quick message to ask whether the TM506 is still available and, if so, whether I can put in my request ? Thank you, Rick.
By momemeca · #187444 ·
4051 Rampack Solid-state Tape Replacement
The vintageTEK Museum has just released a new version of the 4051 Rampack, It is the same form factor as a rompack, and can store 2MB (equivalent to 7 DC300 tapes) of programs and data in rewritable
By Bob Haas · #187443 ·
TM506 + WR501 (without cables) + DD501 + LA501 is looking for a new home
Dear all, if anybody is interested in the following parts, please write me a PM, please. Mainframe TM506 (working unit) including Word Recognizer/Delay WR501 (without cables) Digital Delay DD501 and
By Dr.-Ing. Stephan Massner · #187442 ·
Re: 7000 series parts hulks
All teh plugins are spoken for as are the 7603 and 7613 -- Oz (in DFW, Texas USA) N1OZ
By Oz-in-DFW · #187441 ·
Tek 308 data analyzer - serial data
I have a small device that uses a 9600 baud rs-232 connection for communication that's not working correctly. I thought I could use my 308 to take a look at the data stream and I can't seem to get the
By Paul Amaranth · #187440 ·
Re: Help with (another) broken 7B92 please
7B92, S/N B052568, No Sweep, Help! Thanks Miguel, TP321, and R765, have ~60mv, ? usec pulses. So pulses but probably not functional. I went back and checked carefully on all pins of U310 and found,
By Bruce Atwood · #187439 ·
Re: 2445B vertical non-linearity other issues.
I'm building a replacement U400 from discrete surface mount components. I'll let you know how it works out.
By Robin Szemeti · #187438 ·