Re: 1S1 Serial No. 100396 Guernsey
When troubleshooting analog sampling systems, I've found that it is most efficient to start with the timing circuitry. I don't worry about sampling gate until after I confirm that the timing circuitry
By Kurt Rosenfeld · #111740 ·
Re: Help in using TEK Differential Amp to check Switchmode Supplies
I meant for connecting to the 340 volt peak supply which is typical on capacitor input switching power supplies. 120 volt AC input power supplies use a voltage doubler to get to 340 volts DC. As
By David Hess · #111739 ·
Re: Help in using TEK Differential Amp to check Switchmode Supplies
I have a red 250 MHz x10 Chinese probe rated to 1.2 kVolts and they are not kidding about the voltage rating; 2 kVolts will cause it to break down. It does deliver a 1.4 nanosecond rise and fall time
By David Hess · #111738 ·
Re: 1S1 Serial No. 100396 Guernsey
I find that setting the display mode to "man" (manual) mode simplifies things. TekScopes@...> wrote: [Non-text portions of this message have been removed]
By Kurt Rosenfeld · #111737 ·
Re: 1S1 Serial No. 100396 Guernsey
Sampling scopes have DC offset, which may be a multi-turn or a coarse/fine control.This could drive the trace way off screen.Does it behave the same in both plugin slots ? Since you said you were not
By HankC, Boston, WA1HOS · #111736 ·
Re: Help in using TEK Differential Amp to check Switchmode Supplies
I have one, I use it with a 485. Pretty satisfied with it. I consider it "basic equipment". I have taken it up to 2.6KV working on Z axis circuits with no ill effects. I see them with higher voltage
By Mike McGinn · #111735 ·
Re: Help in using TEK Differential Amp to check Switchmode Supplies
wrote:
By Bert Haskins · #111734 ·
Re: Help in using TEK Differential Amp to check Switchmode Supplies
On a flyback supply that is the truth, yes. Up to 800V Mosfets, are not uncommon. Apart from the voltage rating, x10 may mean no more than 50V/DIV on some scopes, making it completely impractical to
By Stefan Trethan <stefan_trethan@...> · #111733 ·
Re: Help in using TEK Differential Amp to check Switchmode Supplies
"... adequate for 240 VAC work which is unlikely to produce higher than +/- 340 volts DC...." Erm, guys, if these are switched mode supplies on a 240V system, when the switching transistor switches
By David Thomas · #111732 ·
1S1 temperamentalisms re #100396
from Tim P (UK) re the difficulty triggering these beasts, I have four - one works perfectly, one 'sort-a-works' with a fast trigger, one is very iffy and the other one has never worked, and is
By Tim Phillips <tim@...> · #111731 ·
Re: Windows (OT)
Hmm, Windows Update is temporarily (?) broken, at least for some people: http://www.msfn.org/board/topic/173049-windowsmicrosoft-update-not-working-on-windows-2000xp2003/ Automatic updates seem to be
By David DiGiacomo · #111730 ·
Re: Windows (OT)
On 11/19/2014 12:07 PM, David DiGiacomo telists@... [TekScopes] wrote: . . . &#92;POSReady So, is that a Microsoft update for the point of sale crowd that is still supported? If so, neato!
By John Griessen <john@...> · #111729 ·
Re: Windows (OT)
Type this in a command window: reg add HKLM&#92;SYSTEM&#92;WPA&#92;POSReady /v Installed /t REG_DWORD /d 1 /f Then run Windows Update/Microsoft Update as normal.
By David DiGiacomo · #111728 ·
Windows (OT)
Someone mentioned still using WinXP and updates to it that are available, even though Microsoft has abandoned it. Where to look?
By John Griessen <john@...> · #111727 ·
Re: scanned: Allen-Bradley Ferrites
Thanks Kurt, interesting data! Albert
By Albert Otten · #111726 ·
Re: Testing Electrolythics and ordering replacements
yes, I too thank you for the heads up, and have ordered the kit. Can you tell me how it does as a ESR meter? "Having sex is like playing bridge. If you don't have a good partner, you'd better have a
By Randy Warren · #111725 ·
Tektronix 585A, Free to good home
Free Tek 585A (s/n 009721) on cart. It was working last time I checked but needs some TLC. It is complete, free to anyone who can pick it up. You must take cart also. I have no time to devote to this
By victor.silva · #111724 ·
Re: 1S1 Serial No. 100396 Guernsey
It's important to point out that sampling plug-ins drive the horizontal display of the scope. Make sure it's connected to the scope the right way. There's a free run setting for the trigger. I haven't
By Snappy · #111723 ·
Re: Weird cursor problem on 2465 scope
Congrats on finding your issue. I guess if panos' 2445 has the same issue, a quick way to diagnose it would be to measure the DLY_REF line resistance ground and the -1.25V "rail". The easiest way to
By Sigur?ur ?sgeirsson · #111722 ·
Re: TEK 2236 sweep problem
I see a cable with a toroid on it and the connector is hanging loose. It is near the high voltage lead. I will find out where it goes. Could it be THIS simple?
By e9m47 · #111721 ·