Re: Firmware update error Zoom AA-35


 

¿ªÔÆÌåÓý

A flashing analyzer screen indicates that the analyzer is in bootloader mode.
The correct way to take the analyzer out of this mode is to successfully complete the firmware update.

Try manual firmware update.
To do this, download the firmware file here:
Next, connect the analyzer to the PC (do not forget to remove the batteries), launch FlashTool, specify the update from the local file in the settings, and specify the path to the downloaded firmware file.

After that, the flashing process should complete successfully.

If you have a problem, try using an earlier version of FlashTool:

Best Regards, Alex

07.03.2021 23:28, S S §á§Ú§ê§Ö§ä:

On Sun, Mar 7, 2021 at 11:53 AM, Alex RigExpert wrote:
Hi!
?
What version of the analyzer's firmware was before the update?
?
Yes, important information. I recommend removing the batteries from the analyzer before upgrading. In this case, the update process is much more stable.
?
Best regards, Alex
When I turn the unit on version 1.13 is displayed. And the only way I could unlock the device from continuously flashing was to remove the batteries.