Hi Ken,
?
The PEMPro ON data appears to show a fairly large reduction in periodic error. There's a bit of a rotation error in the data, which may mean the PEMPro calibration wasn't as good as it could be. But overall, after derotating the data, worm PE was measured at less than an arcsecond. I can't tell what happened with your PHD2 guide log. There doesn't seem to be any improvement, but I would expect (from the PEMPro plot) that the improvement should be significant.
?
?
There is something wrong with the worm or your measurement, despite PEC being able to correct for it. That large peak is very unusual, and well exceeds any other periodic error I've seen from Losmandy mounts. At the very least, I'd try to clean, regrease, and adjust the worm/gear mesh, but also, possibly swap the DEC and RA worms to see if you get a better result with another worm.
?
?
?