Re: HP8920A RF input issue #HP8920A


 

¿ªÔÆÌåÓý

Thanks Rab for your prompt reply.

Do you have a contact for RIC? . Preferably Email, which woks well when we are 12 hours apart.

Best Regards

William Macham??? VK6IC


On 16/07/2023 8:23 am, rab@... wrote:

Hi, call rick at amtronix to get the attenuator, label and call data for $125

On Jul 15, 2023 4:48 AM, "William Macham via groups.io" <westwill1@...> wrote:

Hi there,

I have an 8921A that needs a Hi power Attenuator. they seem to be out of production. I live in Perth Australia and wondered if you can advise a supplier for the Stripline type 50 Ohm (?100W) resistor that I can carefully solder into the Heatsink Cavity. I appreciate that I will need to do some Calibration to get the unit back in Specification. I can see plenty of Termination tpes, but they do not have the two Tags.

Any help would be appreciated


Regards

William Macham


On 22/06/2023 8:26 pm, jyelmgren wrote:
since this is an 8920A may be more of a challenge since you need the calibrate routine and data ( usually fixed for the most part anyway) on an Epson type card.

Probably cheaper to just replace what you have .

I have a number of those low power input attenuators from upgrades as well if it is defective and there are plenty out there.

Just use an external high power attenuator and input the value in the settings screen ...will correct the input and output levels ...one of the nice things about the 892X.

Usually the low power option was pretty rare in an 8920A...mostly the standard 60W and the optional 100W out there.? 8920B's seemed to have a lot more variance in that regard.