Re: 8510C questions and cabling
Hi Reg,
Here is a card that could be helpful to you. If it doesn't come across, it is called the HP8510C Pre Flight Checkout.
The 83610B has all sorts of options. No options is best, as I recall. Press its preset button, and it will set itself up in sweep mode at its full frequency range. This will tell you if it is the 45MHz option, or the 10MHz option... or,you can just try tuning it down there.
There are only 6 cables that connect the sweeper:
1) HPIB to the ANA's hub (not the external hub) 2) RF output to the test set's RF input. 3) Stop Sweep 4) Trigger 5) Source Interface 6) 10MHz reference.
I have used the 8350B, and am currently using an 8340B, so I haven't actually tried the above connections.
-Chuck Harris
Reginald Beardsley via groups.io wrote:
toggle quoted message
Show quoted text
I just received an 83631B sweeper to complete my 8510C. In reading the service quick reference guide, 08510-90317, I see there are lots of gotchas and a bunch of cables I need, not all of which have HP part numbers listed clearly. It's also not clear which ones are needed when. In addition there are numerous notes about not making certain connections, but it's not clear which cable the reference is to.
So here is what I have so far:
85101C / 3031A opt 010, 011
85102A / 2546A opt H0 w/ 85103A sticker dated 10/13/1987 and 8530A front panel label
83631B / 3844A w/ Agilent refurbished Jun 2000 sticker
8515A / 3602A
8512A / 2345A
Neither the 8512A nor 8515A have the source interface port shown on p 1-3 of the service quick reference. Does that present a problem?
The only cable I have at present is the 85101C to 85102A connection. I also have an HP floppy with the 8530A FW. The 8510C is what is currently loaded. I've not tried to read the 8530A disk yet.
So now to questions:
What other cables do I need?
What do I need to use it in 8530A mode?
Can the 8512A and 8515A be connected at the same time?
Are any of the units incompatible?
What do I need to do to bring this up?
I've seen the 83631B listed as 10 MHz to 26.6 GHz and 45 MHz to 26.6 GHz. How does one tell which is the case? Is it test set dependent at the low frequency?
I'm a hardcore Unix guy. Is it worth hunting down an HP 9000 series to use with it or can I use more recent computers e.g. HP Z400?
HP-UX would have me trying to admin 5 distinct operating systems. Working entirely alone, I've discovered that it really taxes my memory switching between the 4 I'm already running.
I've read a good bit of 8510 documentation, but the system is so complex that I'm not sure I've even found all of it, much less read everything I need to read.
Thanks, Reg
|
Unfortunately, the Artek scan isn't free, so I will have to pass on that.
Back to the hardware, the readout shows OL on the high ranges and zeros on the low ranges.? I don't understand the relationships with the range settings, autocal, and autozero.? Sometimes TP4 shows -14V and sometimes -13V.? TP3 shows the input from the panel but not always.? This unit is way over my head.
TP3 should be close to whatever is being measured, whether it's the input or one of the auto-cal constants. ? So best to turn autocal off or the relays will be ticking and TP3 changing.
TP4 being so low seems to indicate the main amplifier isn't working correctly.? There is negative feedback through Q18, Q19 or Q21 to Q17 from TP4.? They are controlled by A10U5.? I'd take a look at U5 pins 2, 13 and 14 that control the FET switches.
Orin.?
|
8510C questions and cabling
I just received an 83631B sweeper to complete my 8510C. In reading the service quick reference guide, 08510-90317, I see there are lots of gotchas and a bunch of cables I need, not all of which have HP part numbers listed clearly. It's also not clear which ones are needed when. In addition there are numerous notes about not making certain connections, but it's not clear which cable the reference is to.
So here is what I have so far:
85101C / 3031A opt 010, 011
85102A / 2546A opt H0 w/ 85103A sticker dated 10/13/1987 and 8530A front panel label
83631B / 3844A w/ Agilent refurbished Jun 2000 sticker
8515A / 3602A
8512A / 2345A
Neither the 8512A nor 8515A have the source interface port shown on p 1-3 of the service quick reference. Does that present a problem?
The only cable I have at present is the 85101C to 85102A connection. I also have an HP floppy with the 8530A FW. The 8510C is what is currently loaded. I've not tried to read the 8530A disk yet.
So now to questions:
What other cables do I need?
What do I need to use it in 8530A mode?
Can the 8512A and 8515A be connected at the same time?
Are any of the units incompatible?
What do I need to do to bring this up?
I've seen the 83631B listed as 10 MHz to 26.6 GHz and 45 MHz to 26.6 GHz. How does one tell which is the case? Is it test set dependent at the low frequency?
I'm a hardcore Unix guy. Is it worth hunting down an HP 9000 series to use with it or can I use more recent computers e.g. HP Z400?
HP-UX would have me trying to admin 5 distinct operating systems. Working entirely alone, I've discovered that it really taxes my memory switching between the 4 I'm already running.
I've read a good bit of 8510 documentation, but the system is so complex that I'm not sure I've even found all of it, much less read everything I need to read.
Thanks, Reg
|
Re: First Gen HP 54810A with dead HD, Got ISO 3.5, Any way to write it to a drive?
On Wed, Apr 1, 2020 at 10:58 PM <ujonsson@...> wrote: Hi, My loved 54810A died on me. It is the oldest Gen 1 unit no VIN running 95 and it has not been upgraded. I did read up on the recovery process from the HP documentation and I have the version 3.5 disc image. I can open the ISO and I find the files listed below files.
HP suggested that one use a printer port CD Drive to recover it. The Ghost.EXE wont run on my win10 machine. I do have an XP machine I might get GHOST to run on but it is a laptop. You are getting yourself confused. You only need the parallel port CDROM drive if you want to populate the hard drive while it's installed in the scope. If you connect the new hard drive to another computer, such as your laptop, you can use its regular CDROM drive.
|
The Artek version may not be free, but I can state (from multiple direct experiences) that Dave does an amazing job, IMO worth every penny of his (again, IMO) very reasonable prices.
If the unit is worth repairing, it's worth having a readable manual to help fix it. The decision is, ultimately, up to you but... If it were me, I'd do it in a heartbeat.
toggle quoted message
Show quoted text
On 02-Apr-20 13:25, Bob Albert via groups.io wrote: Unfortunately, the Artek scan isn't free, so I will have to pass on that.
Back to the hardware, the readout shows OL on the high ranges and zeros on the low ranges.? I don't understand the relationships with the range settings, autocal, and autozero.? Sometimes TP4 shows -14V and sometimes -13V.? TP3 shows the input from the panel but not always.? This unit is way over my head.
Bob
On Thursday, April 2, 2020, 12:49:21 PM PDT, Orin Eman <orin.eman@...> wrote:
I took a look at that one.
The text looks OK, but the diagrams and schematics aren't really readable.
I'd get the Artek scan.
On Thu, Apr 2, 2020 at 12:40 PM Bob Albert via groups.io <> <bob91343@... <mailto:[email protected]>> wrote:
My manual is one I downloaded from the Keysight site.
On Thursday, April 2, 2020, 11:20:40 AM PDT, Bob Albert via groups.io <> <bob91343@... <mailto:[email protected]>> wrote:
Orin, TP4 is at about -13 Volts, no relation to the input signal.
Bob
On Thursday, April 2, 2020, 10:31:01 AM PDT, Orin Eman <orin.eman@... <mailto:orin.eman@...>> wrote:
On Thu, Apr 2, 2020 at 9:44 AM Bob Albert via groups.io <> <bob91343@... <mailto:[email protected]>> wrote:
It's complicated by the fact that I don't know where the parts are for different parts of the diagrams. ?
Which manual do you have?
The Artek scan that I have has a diagram of the A10 board on page 222 of the pdf (or at "A10 & A12 Board" under Diagrams in the bookmarks in Artek's scan).
Orin.?
-- --- Bruce Lane, ARS KC7GR
kyrrin (at) bluefeathertech dot com "Quando Omni Flunkus Moritati" (Red Green)
|
Unfortunately, the Artek scan isn't free, so I will have to pass on that.
Back to the hardware, the readout shows OL on the high ranges and zeros on the low ranges.? I don't understand the relationships with the range settings, autocal, and autozero.? Sometimes TP4 shows -14V and sometimes -13V.? TP3 shows the input from the panel but not always.? This unit is way over my head.
Bob
On Thursday, April 2, 2020, 12:49:21 PM PDT, Orin Eman <orin.eman@...> wrote:
I took a look at that one.
The text looks OK, but the diagrams and schematics aren't really readable.
I'd get the Artek scan.
toggle quoted message
Show quoted text
My manual is one I downloaded from the Keysight site.
On Thursday, April 2, 2020, 11:20:40 AM PDT, Bob Albert via <bob91343= [email protected]> wrote:
Orin, TP4 is at about -13 Volts, no relation to the input signal.
Bob
On Thursday, April 2, 2020, 10:31:01 AM PDT, Orin Eman < orin.eman@...> wrote:
It's complicated by the fact that I don't know where the parts are for different parts of the diagrams. ?
Which manual do you have?
The Artek scan that I have has a diagram of the A10 board on page 222 of the pdf (or at "A10 & A12 Board" under Diagrams in the bookmarks in Artek's scan).
Orin.?
|
I took a look at that one.
The text looks OK, but the diagrams and schematics aren't really readable.
I'd get the Artek scan.
toggle quoted message
Show quoted text
My manual is one I downloaded from the Keysight site.
On Thursday, April 2, 2020, 11:20:40 AM PDT, Bob Albert via <bob91343= [email protected]> wrote:
Orin, TP4 is at about -13 Volts, no relation to the input signal.
Bob
On Thursday, April 2, 2020, 10:31:01 AM PDT, Orin Eman < orin.eman@...> wrote:
It's complicated by the fact that I don't know where the parts are for different parts of the diagrams. ?
Which manual do you have?
The Artek scan that I have has a diagram of the A10 board on page 222 of the pdf (or at "A10 & A12 Board" under Diagrams in the bookmarks in Artek's scan).
Orin.?
|
My manual is one I downloaded from the Keysight site.
On Thursday, April 2, 2020, 11:20:40 AM PDT, Bob Albert via groups.io <bob91343@...> wrote:
Orin, TP4 is at about -13 Volts, no relation to the input signal.
Bob
On Thursday, April 2, 2020, 10:31:01 AM PDT, Orin Eman <orin.eman@...> wrote:
It's complicated by the fact that I don't know where the parts are for different parts of the diagrams. ?
Which manual do you have?
The Artek scan that I have has a diagram of the A10 board on page 222 of the pdf (or at "A10 & A12 Board" under Diagrams in the bookmarks in Artek's scan).
Orin.?
|
Re: HP 419A DC Null Voltmeter - INSANE behaviour
Cancel my last post, warmed up, my 419 contiues with the INSANE behavior.
|
Orin, TP4 is at about -13 Volts, no relation to the input signal.
Bob
On Thursday, April 2, 2020, 10:31:01 AM PDT, Orin Eman <orin.eman@...> wrote:
It's complicated by the fact that I don't know where the parts are for different parts of the diagrams. ?
Which manual do you have?
The Artek scan that I have has a diagram of the A10 board on page 222 of the pdf (or at "A10 & A12 Board" under Diagrams in the bookmarks in Artek's scan).
Orin.?
|
It's complicated by the fact that I don't know where the parts are for different parts of the diagrams. ?
Which manual do you have?
The Artek scan that I have has a diagram of the A10 board on page 222 of the pdf (or at "A10 & A12 Board" under Diagrams in the bookmarks in Artek's scan).
Orin.?
|
It's complicated by the fact that I don't know where the parts are for different parts of the diagrams.? All cables have been wiggled, all parts have been poked and tapped.? Nothing is running hot.? TP3 replicates the input signal but TP4 does not.? The 10V reference is spot on.
What is TP4? ?0V?
BTW, best to turn auto-cal off or it will be constantly trying to switch back and forth between the input signal and references.
|
Dan, I have done much of what you suggest.? The front/rear switch is okay.? Even on Ohms it reads all zeros.
On power up it presents OL until I push some buttons.? On TEST it goes through from 11 in sequence down to 0.? In Ohms the front jacks put out a test current.
It's complicated by the fact that I don't know where the parts are for different parts of the diagrams.? All cables have been wiggled, all parts have been poked and tapped.? Nothing is running hot.? TP3 replicates the input signal but TP4 does not.? The 10V reference is spot on.
The plethora of information in the manual is overwhelming.
Bob
On Thursday, April 2, 2020, 09:30:20 AM PDT, Stephen Hanselman <kc4sw.io@...> wrote:
Good point Don, although the real question would be is the front/back switch still a switch. ?We have seen a couple where the switch contacts themselves went bad. ?Nothing gets through, but it does pass self test.? Regards, ? Stephen Hanselman Datagate Systems, LLC
toggle quoted message
Show quoted text
On Apr 2, 2020, at 07:21, Don Bitters via groups.io <donbitters@...> wrote:
?When you power the DMM on it goes through a POP (Power On Preset routine). ?The display does several things including showing all 8¡¯s. ?Are you getting this? ?This is also the ideal time to check for activity on the suspect IC¡¯s. ?Check the manual for what is displayed during the POP. ?If the DMM does not complete POP it should show an error code, see manual for codes and meaning. If the manual is suggesting one of 2 IC¡¯s then you could check for activity on the address and data lines on the 2 IC¡¯s with an o-scope during POP. ?Look for proper levels, TTL or CMOS if a digital IC and good wave shape - data should be square or rectangular shapes with little or no rounding of leading edges, ramps should be straight lines, check for any overly warm IC¡¯s. ?Again check the cabling connections for oxidation and good contact, DeOxit is your friend if used sparingly. ?The display brd has to have a clock signal to operate - check the clock, if I recall correctly it should be 1 MHz. If POP is good and no errors then, try putting a 1VDC into the DMM and then chase the signal through the signal path until it is digitized to be sent to the display (see manual). Don¡¯t be afraid to gently wiggle cables and gently press on brds and components, looking for any changes on your display. A dumb question, where is the front/rear connector switch set to, and if you select ohms mode does your display still show all 0¡¯s (with nothing connected to the inputs)? Don Bitters
|
Good point Don, although the real question would be is the front/back switch still a switch. ?We have seen a couple where the switch contacts themselves went bad. ?Nothing gets through, but it does pass self test.? Regards, ? Stephen Hanselman Datagate Systems, LLC
toggle quoted message
Show quoted text
On Apr 2, 2020, at 07:21, Don Bitters via groups.io <donbitters@...> wrote:
?When you power the DMM on it goes through a POP (Power On Preset routine). ?The display does several things including showing all 8¡¯s. ?Are you getting this? ?This is also the ideal time to check for activity on the suspect IC¡¯s. ?Check the manual for what is displayed during the POP. ?If the DMM does not complete POP it should show an error code, see manual for codes and meaning. If the manual is suggesting one of 2 IC¡¯s then you could check for activity on the address and data lines on the 2 IC¡¯s with an o-scope during POP. ?Look for proper levels, TTL or CMOS if a digital IC and good wave shape - data should be square or rectangular shapes with little or no rounding of leading edges, ramps should be straight lines, check for any overly warm IC¡¯s. ?Again check the cabling connections for oxidation and good contact, DeOxit is your friend if used sparingly. ?The display brd has to have a clock signal to operate - check the clock, if I recall correctly it should be 1 MHz. If POP is good and no errors then, try putting a 1VDC into the DMM and then chase the signal through the signal path until it is digitized to be sent to the display (see manual). Don¡¯t be afraid to gently wiggle cables and gently press on brds and components, looking for any changes on your display. A dumb question, where is the front/rear connector switch set to, and if you select ohms mode does your display still show all 0¡¯s (with nothing connected to the inputs)? Don Bitters
|
Seems I neglected to indicate that I am located in Los Angeles, CA.
Kindly, Michael
toggle quoted message
Show quoted text
On Wed, Apr 1, 2020 at 10:56 PM Michael Yellin via <michaelhq54= [email protected]> wrote: I have the following, all is OBO:
Agilent 16702B logic analyzer with 2 cards, and lots of
accessories $500.00 plus shipping
HP 8560E Spec An with TG, Demod, and Mass Memory ¨C some of
the buttons are intermittent $800.00 plus shipping
HP 53132A Freq Counter with Factory 3ghz option.? $600.00 plus shipping
3x HP 8116A Pulse/Function Generator.? 1 has the GPIB option.? 150 for the one with GPIB, and 100 each for
the others, plus shipping.
Agilent 66332A IT IS HEAVY. 125.00
Pics
|
Does the 3455A have the issue with ROMs beginning to fail like the ones in the 3456A?
Thanks, Barry - N4BUQ
toggle quoted message
Show quoted text
----- Original Message ----- From: "Harvey White" <madyn@...> To: [email protected] Sent: Thursday, April 2, 2020 9:54:54 AM Subject: Re: [HP-Agilent-Keysight-equipment] HP 3455a DVM repair
I'd be glad to help if I could, but I don't have one.? What I've said is based on the 3456As that I have.? Take a low frequency square wave, say 1 second on one second off.? Put that to the input of the meter.? The manual should show you the A/D converter and the inguard processor (IIRC it's inguard), You should be able to see that signal at the a/d inputs. You should be able to identify the outputs of the inguard processor, look to see if anything is changing there.? Then try the outguard processor, that's the one that drives the display.? Again, you're looking at board signals, not signals going so much between chips.
What I suspect, as do others, is that the inguard processor may not be talking to the outguard processor.? The inguard processor may be "broken" as a guess.? That the display changes with range is good.? That it's all zeros may be that it's simply getting all zeros.
Look at the actual boards and check the signals there.? You don't need to start looking at chips until you figure out what's going on.
Typically, the inguard processor is run from an isolated supply (and ALL the voltages must be exact, had that problem with a -30 volt supply in a 3456), the input attenuators, AC converter, ohms current supply, A/D and that processor are all run from that supply.? Then the output of that processor goes through some sort of isolation network rated for many KV to the main processor.? HP used a pair of highly insulated transformers, just a few loops of wire and a core, to get a clock and data signal over in the 3456A.? Likely they're doing the same thing here.
I had a bad inguard processor in one of mine.
Harvey
On 4/1/2020 8:19 PM, Bob Albert via groups.io wrote:
Harvey, my main problem is identifying the various parts.? The manual isn't the easiest for me.? Perhaps you can walk me around to the various places where I can make measurements.? I certainly have no shortage of sources of dc or ac of any reasonable frequency and amplitude.
I did measure the 10V reference with my 3456A at 10.00007 I think. (not sure how many zeros)? So what we have here are all the pieces of a good meter if I can make it play.
Bob
On Wednesday, April 1, 2020, 05:09:17 PM PDT, Harvey White <madyn@...> wrote:
If I might:
generally, a signal goes through attenuators, then is converted to digital.? This is done in a floating processing environment called (by HP) the inguard processor.? That is optically coupled to an "outguard" processor, which is the main processor driving the display.? Any break in the chain can result in zeros.
Firstly, I'd take the signal from the inputs, then see how far it gets.? If you can see anything changing in the digitally coupled outputs (optically isolated), might be working.? You'll need to check on both ends of the optical isolators.? That's assuming they have them.? The 3456 has two very isolated transformers that need to be checked.? What I found is that one of the processors itself was dead when I was trying to resurrect some of them.
So the bottom line is to try to follow the signal. An AC signal would be easiest but you'd have to tell it that it's really DC, otherwise you to the AC converter.? If you should try to do an AC signal into the DC input, what you're looking for is the presence or absence of a signal (to a point where the circuitry gets confused and/or ignores it) and after that, the signal isn't there.? At that point, you need to match the AC signal to an AC input, or DC to DC.
This is based on my experiences with a 3456, so YMMV
Harvey
On 4/1/2020 1:36 PM, Bob Albert via groups.io wrote: Well Chuck, I don't know what to do next with this beast.? It acts as though it works but the display gives nothing useful.? Yet the display electronics seem to be working okay.
If I put a signal in, it ignores it. Open or short on Ohms, voltage or no on volts, gives zeros and no autorange activity.? So perhaps it's an input circuit issue.? In Test I get error 11.? I press test repeatedly and it gives error numbers 10, 9, etc down to 0.
The manual sucks.? It tells me to look on module such and such but nowhere does it seem to show me which module is where.
Bob
On Wednesday, April 1, 2020, 10:22:56 AM PDT, Chuck Harris <cfharris@...> <mailto:cfharris@...> wrote:
Hi Bob,
I don't see how you can come to that conclusion.
The 3455A is highly reparable, and should be well within your abilities.? Lots of people have fixed their own successfully.
What you won't be able to do is send its calibration module out for calibration.? You will have to do that yourself.? But, it is not any more difficult to calibrate than the 3456A.
It is not my meter of choice, as I greatly prefer the 3456A, but for many, it is a useful addition to their labs.
-Chuck Harris
Bob Albert via groups.io wrote:
? I don't see a whole lot of encouragement here regarding the possibility of getting my 3455A working.? Maybe I will just put it aside and wait for a scrap unit to swap out boards or parts.? My 3456A still works so I am grateful for that.
Bob ? ? On Wednesday, April 1, 2020, 06:40:17 AM PDT, Chuck Harris <cfharris@... <mailto:cfharris@...>> wrote:
? The 3455A was a dinosaur when it was first released.
The 3455A and 3456A were sold side by side, with the 3455A being about twice as expensive as the much more capable 3456A.? The 3455A's interesting feature was it was supposed to be cheap to calibrate. All you had to do to do a complete calibration was to unplug the module in the back panel, and replace it with a recently calibrated module.? The idea was you could order up a calibration by mail, and return your old calibration module as a core... never having to send the full meter in for calibration.
Interesting idea, but surely not worth twice the price of a 3456A.
I say it was a dinosaur, because it used all small scale logic rather than large highly capable hybrid circuitry, like the 3456A.? It should be much easier to repair, but it should also be much less reliable.
And, that is my experience with the two units.
-Chuck Harris
paul swedberg wrote:
With respect to the 3455 and 3456 the difference is night and day. I have
numbers of 3456s and was given a seriously troubled 3455. Quite a interesting meter. Replaced at least 11 actually malfunctioning chips on the logic board. The
logics simple so easy to verify the chips were bad. No shot gunning involved. Never really did get it working but ended up with a different 3455 that needed help and the AC board and it is in fine shape. With respect to the old 3455 it has both a logic and voltmeter issue. The 3455
uses an unusual processor and teh DVM uses the same proc and a masked eprom
that is known to go bad. It acts as a state machine. The 3456s are losing there eproms and copies can be found on the net. Thank
heavens. Regards Paul WB8TSL
|
Re: Maximum depth of HP rack mountable gear?
If in an earthquake area, don't forget to tether your racks. Apart from the obvious falling over, a rack can "walk" across the floor and crush you or block access.
Robert G8RPI.
|
I'd be glad to help if I could, but I don't have one.? What I've
said is based on the 3456As that I have.? Take a low frequency
square wave, say 1 second on one second off.? Put that to the
input of the meter.? The manual should show you the A/D converter
and the inguard processor (IIRC it's inguard), You should be able
to see that signal at the a/d inputs.? You should be able to
identify the outputs of the inguard processor, look to see if
anything is changing there.? Then try the outguard processor,
that's the one that drives the display.? Again, you're looking at
board signals, not signals going so much between chips.
What I suspect, as do others, is that the inguard processor may
not be talking to the outguard processor.? The inguard processor
may be "broken" as a guess.? That the display changes with range
is good.? That it's all zeros may be that it's simply getting all
zeros.
Look at the actual boards and check the signals there.? You don't
need to start looking at chips until you figure out what's going
on.
Typically, the inguard processor is run from an isolated supply
(and ALL the voltages must be exact, had that problem with a -30
volt supply in a 3456), the input attenuators, AC converter, ohms
current supply, A/D and that processor are all run from that
supply.? Then the output of that processor goes through some sort
of isolation network rated for many KV to the main processor.? HP
used a pair of highly insulated transformers, just a few loops of
wire and a core, to get a clock and data signal over in the
3456A.? Likely they're doing the same thing here.
I had a bad inguard processor in one of mine.
Harvey
On 4/1/2020 8:19 PM, Bob Albert via
groups.io wrote:
toggle quoted message
Show quoted text
Harvey, my main problem is
identifying the various parts.? The manual isn't the easiest
for me.? Perhaps you can walk me around to the various places
where I can make measurements.? I certainly have no shortage
of sources of dc or ac of any reasonable frequency and
amplitude.
I did measure the 10V
reference with my 3456A at 10.00007 I think.? (not sure how
many zeros)? So what we have here are all the pieces of a good
meter if I can make it play.
Bob
On Wednesday, April 1, 2020, 05:09:17 PM PDT, Harvey
White <madyn@...> wrote:
If I might:
generally, a signal goes through attenuators, then?
is converted to digital.? This is done in a floating
processing environment called (by HP) the inguard
processor.? That is optically coupled to an "outguard"
processor, which is the main processor driving the
display.? Any break in the chain can result in zeros.?
Firstly, I'd take the signal from the inputs, then
see how far it gets.? If you can see anything changing
in the digitally coupled outputs (optically isolated),
might be working.? You'll need to check on both ends
of the optical isolators.? That's assuming they have
them.? The 3456 has two very isolated transformers
that need to be checked.? What I found is that one of
the processors itself was dead when I was trying to
resurrect some of them.
So the bottom line is to try to follow the signal.?
An AC signal would be easiest but you'd have to tell
it that it's really DC, otherwise you to the AC
converter.? If you should try to do an AC signal into
the DC input, what you're looking for is the presence
or absence of a signal (to a point where the circuitry
gets confused and/or ignores it) and after that, the
signal isn't there.? At that point, you need to match
the AC signal to an AC input, or DC to DC.
This is based on my experiences with a 3456, so YMMV
Harvey
On 4/1/2020
1:36 PM, Bob Albert via groups.io wrote:
Well Chuck, I don't know what to do
next with this beast.? It acts as though it works
but the display gives nothing useful.? Yet the
display electronics seem to be working okay.
If I put a signal in, it ignores it.?
Open or short on Ohms, voltage or no on volts, gives
zeros and no autorange activity.? So perhaps it's an
input circuit issue.? In Test I get error 11.? I
press test repeatedly and it gives error numbers 10,
9, etc down to 0.
The manual sucks.? It tells me to look
on module such and such but nowhere does it seem to
show me which module is where.
Bob
On Wednesday, April 1, 2020, 10:22:56 AM PDT,
Chuck Harris <cfharris@...>
wrote:
Hi Bob,
I don't see how you can come to that conclusion.
The 3455A is highly reparable, and should be
well
within your abilities.? Lots of people have
fixed
their own successfully.
What you won't be able to do is send its
calibration
module out for calibration.? You will have to do
that
yourself.? But, it is not any more difficult to
calibrate than the 3456A.
It is not my meter of choice, as I greatly
prefer
the 3456A, but for many, it is a useful addition
to
their labs.
-Chuck Harris
Bob Albert via groups.io wrote:
>? I don't see a whole lot of encouragement
here regarding the possibility of getting my
3455A working.? Maybe I will just put it aside
and wait for a scrap unit to swap out boards or
parts.? My 3456A still works so I am grateful
for that.
> Bob
>? ? On Wednesday, April 1, 2020, 06:40:17 AM
PDT, Chuck Harris < cfharris@...>
wrote:
>
>? The 3455A was a dinosaur when it was first
released.
>
> The 3455A and 3456A were sold side by side,
with the
> 3455A being about twice as expensive as the
much more
> capable 3456A.? The 3455A's interesting
feature was
> it was supposed to be cheap to calibrate.?
All you
> had to do to do a complete calibration was
to unplug
> the module in the back panel, and replace
it with a
> recently calibrated module.? The idea was
you could
> order up a calibration by mail, and return
your old
> calibration module as a core... never
having to send
> the full meter in for calibration.
>
> Interesting idea, but surely not worth
twice the price
> of a 3456A.
>
> I say it was a dinosaur, because it used
all small
> scale logic rather than large highly
capable hybrid
> circuitry, like the 3456A.? It should be
much easier
> to repair, but it should also be much less
reliable.
>
> And, that is my experience with the two
units.
>
> -Chuck Harris
>
> paul swedberg wrote:
>> With respect to the 3455 and 3456 the
difference is night and day. I have
>> numbers of 3456s and was given a
seriously troubled 3455. Quite a
>> interesting meter.
>> Replaced at least 11 actually
malfunctioning chips on the logic board. The
>> logics simple so easy to verify the
chips were bad. No shot gunning
>> involved. Never really did get it
working but ended up with a different
>> 3455 that needed help and the AC board
and it is in fine shape. With
>> respect to the old 3455 it has both a
logic and voltmeter issue. The 3455
>> uses an unusual processor and teh DVM
uses the same proc and a masked eprom
>> that is known to go bad. It acts as a
state machine.
>> The 3456s are losing there eproms and
copies can be found on the net. Thank
>> heavens.
>> Regards
>> Paul
>> WB8TSL
>>
>>
>>
>>
>
>
>
>
>
>
>
>
|
Re: Maximum depth of HP rack mountable gear?
Or a concrete slab floor on California clay like I have in my garage lab.? Fortunate not to be worried about the weight.? Just the space it takes up and what the wife thinks about more "machines"!
Jim Ford?
Sent from my Verizon, Samsung Galaxy smartphone
toggle quoted message
Show quoted text
-------- Original message -------- From: "Jared Cabot via groups.io" <jaredcabot@...> Date: 4/1/20 4:04 PM (GMT-08:00) Subject: Re: [HP-Agilent-Keysight-equipment] Maximum depth of HP rack mountable gear?
A good point, especially for wooden floors. No worries on my part though as I have a steel framed house with very strong floors. Being in Japan, the building is rated for the big earthquakes we get here so a rack is easy work for the structure. :) On Thu, Apr 2, 2020 at 05:52 AM, Matt Corbett wrote:
You guys already know this, but just in case: an important consideration in rack mounting your gear is the weight that the floor can support.? With some of the taller racks filled with equipment, you will have hundreds of kgs. of weight.? As an extreme (and fabricated) example, a common 7 foot (44U) rack could physically accommodate eleven HP8664A signal generators.? At 35kg each, these generators would weigh a total of 385 kg. (850 lbs.) -- and this doesn't including the weight of the rack itself, and any mounting accessories.? Be safe, everyone!
|
Re: HP 3325A/B Option 002 HV output kit - who wants one?
Confirmed for a kit, whenever...
_Dave KC6UPS
|