Re: E4407B strange peak shapes at low RBW
¿ªÔÆÌåÓý
Keysight will no longer sell the licenses for the N7800A unless you are a large company and want to do your own calibrations but the catch is that you HAVE to use the equipment that's written into the software.? It is a required list of equipment, not a suggested
list which leaves all but a very rare few who are able to do that. ? The cost of the required equipment, just the minimum runs about $100K "used" and that will allow you to do only a limited calibration on most pieces of equipment. ? Everything these days
is in software and very little can be done manually anymore due to the use of A/D's and D/A's and signal processing.? Those allow for great specs but at the expense of allowing for manual calibrations.
You pretty?much have to make the decision of whether or not you need what the later generation equipment can give you or do you want the ability to be able to do your own calibrations.? Of course cost is a huge consideration also.
I pretty much designed my lab around the ability to do the repairs and adjustments myself and have been happy with that decision.? There's a LOT of good equipment out there.
Richard Parrish
From: [email protected] <[email protected]> on behalf of Lothar baier via groups.io <Lothar@...>
Sent: Tuesday, May 31, 2022 9:02 PM To: [email protected] <[email protected]> Subject: Re: [HP-Agilent-Keysight-equipment] E4407B strange peak shapes at low RBW ?
HP used discrete filters for the IF filters , those filters are constructed using multiple poles that are coupled and each pole is adjusted individually to form a specific filter shape and bandwidth as a result if one pole is off than you get a funny shape like the one you are seeing ! On the older analyzers (855x, 8566/68, 70K, 8590¡..? the adjustments were done manually , you needed to build of some shorting clips or cables and follow the instructions on the manual and usually this worked quite well . However on the later model analyzers HP switched from manual adjustments and performance tests and providing HP basic test/alignment software free of charge to using a integrated test environment (N7800A )? with individual test modules for each analyzer family (N7811A for the ESA Series )? , this software requires a paid license however !? I am not aware of any alternate ways to do the adjustments however richard is the authority on this and he might know ? From: [email protected] <[email protected]>
On Behalf Of hardyhansendk via groups.io ? Hi Are you using an external test signal?-it seems so. First i would make a self alignment with the inbuildt? 50 Mhz cal signal. Hardy ? Fra:
[email protected] [mailto:[email protected]]
P? vegne af
Richard Parrish ? You will need to go in and adjust your Xtal and LC filters.? Not hard, just a pain. Richard From:
[email protected] <[email protected]> on behalf of Kalle Kempe via groups.io <kalle.kempe@...> ? Hello Everyone, ?
|
||
Re: Keithley 236 firmware version A10
¿ªÔÆÌåÓýHi, I finally found a link on EEVblog and was able to find and install the new firmware successfully. Thanks! ? ? De?: [email protected] <[email protected]> De la part de Rico Sonderegger ? Hi Yves, I own a 237 (high voltage version). These devices share the same firmware. Only a resistor on the mainboard decides whether the 236 or the 237 version. ? I have the A10 version and can send it to you if you want. On the mainboard there is a second device which is also marked with a firmware version (236 801 A10). In an emergency, I also have this data and can send it to you. Many greetings Rico |
||
hp 8407A Service Cable
All,
Would anyone have a service cable for the 8407A Network Analyzer for the 8412A, 8413A ,and 8414A/B plugins? By looking at the number of pins in the Amphenol connector, it appears to be the same as the 11592-60018 for the 141T Spectrum Analyzer, but shorter. Cheers, Christopher |
||
Re: Keithley 236 firmware version A10
Hi Yves,
I own a 237 (high voltage version). These devices share the same firmware. Only a resistor on the mainboard decides whether the 236 or the 237 version.
?
I have the A10 version and can send it to you if you want.
On the mainboard there is a second device which is also marked with a firmware version (236 801 A10).
But I don't know if you also need it. Is an EEPROM ev. are there only the calibration data stored. In an emergency, I also have this data and can send it to you.
Many greetings Rico
|
||
Re: HP 8510B Service manual CLIP
Hi folks,
I have a broken 8510B display section that I'm trying to fix and I have been chasing a schematic for a long time. Has anyone spotted the mentioned B-supplement? I know those display sections are said to be dirt cheap, but not so on this side of the gread pond (Germany). Wayne, would you please share your reverse engineering results with me? I'd highly appreciate it. This could help a lot with my repair efforts. Thanks and best regards, Tom |
||
Re: E4407B strange peak shapes at low RBW
Lothar baier
¿ªÔÆÌåÓýHP used discrete filters for the IF filters , those filters are constructed using multiple poles that are coupled and each pole is adjusted individually to form a specific filter shape and bandwidth as a result if one pole is off than you get a funny shape like the one you are seeing ! On the older analyzers (855x, 8566/68, 70K, 8590¡..? the adjustments were done manually , you needed to build of some shorting clips or cables and follow the instructions on the manual and usually this worked quite well . However on the later model analyzers HP switched from manual adjustments and performance tests and providing HP basic test/alignment software free of charge to using a integrated test environment (N7800A )? with individual test modules for each analyzer family (N7811A for the ESA Series )? , this software requires a paid license however !? I am not aware of any alternate ways to do the adjustments however richard is the authority on this and he might know ? From: [email protected] <[email protected]>
On Behalf Of hardyhansendk via groups.io
Sent: Tuesday, May 31, 2022 6:59 PM To: [email protected] Subject: Re: [HP-Agilent-Keysight-equipment] E4407B strange peak shapes at low RBW ? Hi Are you using an external test signal?-it seems so. First i would make a self alignment with the inbuildt? 50 Mhz cal signal. Hardy ? Fra:
[email protected] [mailto:[email protected]]
P? vegne af
Richard Parrish ? You will need to go in and adjust your Xtal and LC filters.? Not hard, just a pain. Richard From:
[email protected] <[email protected]> on behalf of Kalle Kempe via groups.io <kalle.kempe@...> ? Hello Everyone, ?
|
||
Re: E4407B strange peak shapes at low RBW
¿ªÔÆÌåÓýHi Are you using an external test signal?-it seems so. First i would make a self alignment with the inbuildt? 50 Mhz cal signal. Hardy ? Fra: [email protected]
[mailto:[email protected]] P? vegne af Richard Parrish ? You will need to go in and adjust your Xtal and LC filters.? Not hard, just a pain. Richard From: [email protected]
<[email protected]> on behalf of Kalle Kempe via
groups.io <kalle.kempe@...>
Sent: Tuesday, May 31, 2022 11:30 AM To: [email protected] <[email protected]> Subject: [HP-Agilent-Keysight-equipment] E4407B strange peak shapes at low RBW ? Hello Everyone, |
||
Keithley 236 firmware version A10
¿ªÔÆÌåÓýHi, ? All the instruments in my lab are mainly HP/Agilent/Keysight, and two from tektronix, but now I have two Keithley sources, a 236 and a 238 functional received very recently. I was able to find the A10 firmware version for the 238, but I'm still looking for the A10 firmware for the 236, I know it exists but can't find it...does anyone have any information where to find this firmware. ? -Yves |
||
Re: HP8920A display onto computer monitor
¿ªÔÆÌåÓýThe SimmConn Labs LCD replacement display does a nice job, although may be more than some want to spend. As I recall, it can bring out a VGA port as an option. ? Also, I have a CRT display removed from my 8920A which is in very good condition, as I only removed it to get the LCD; the display was working fine when removed about a year ago. Can be had for whatever it costs to ship in the US. ? Tom, N8ZM ? From: [email protected] <[email protected]> On Behalf Of RFI-EMI-GUY
Sent: Tuesday, May 31, 2022 3:14 PM To: [email protected] Subject: Re: [HP-Agilent-Keysight-equipment] HP8920A display onto computer monitor ? That youtube video is essential to anyone wanting to use the Sony.? I thought the resolution looked pretty good.
|
||
Re: HP8920A display onto computer monitor
There is more info on the 8920 group /g/HP8924-Family-Spectrum-Analyzers/topics |
||
Re: HP8920A display onto computer monitor
There is no why spend so much money on that Enviado do meu Tele-Movel Em ter, 31 de mai de 2022 16:14, RFI-EMI-GUY <rhyolite@...> escreveu: That youtube video is essential to anyone wanting to use the Sony.? I thought the resolution looked pretty good. |
||
Re: HP8920A display onto computer monitor
That youtube video is essential to anyone wanting to use the Sony.? I thought the resolution looked pretty good.
toggle quoted message
Show quoted text
There are a few different models of the DSC-1024. I wonder which would be the best for use with the HP98XX series? On Tue, May 31, 2022 at 01:35 PM, W6KX wrote: Using the DSC-1024HD yields just O.K. image sharpness.? Lots of hardware but the below youtube method works for viewing and acquiring. |
||